510(k) Report

510(k) · Cleared

Cleared

SmartSpectra Vital Signs Monitor 1.0

K254169510(k) TraditionalCleared Jun 18, 2026177-day reviewClass II21 CFR 870.2785

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Presage Security, Inc. d/b/a Presage Technologies, Inc. · QME Software For Optical Camera-Based Measurement Of Pulse Rate, Heart Rate, Breathing Rate, And/Or Respiratory Rate

Cleared on Jun 18, 2026 as substantially equivalent to K211906 (Oxehealth Limited, 2021) after a 177-day review. Its family tree reaches back 2 generations to 2021.

Decided
Jun 18, 2026
Review
177 days
Class
Class II
Product code
QME
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to DEN200019 (2021). No later device cites it as a predicate yet.

  1. Root · 2 generations backDEN200019 Oxehealth Vital SignsOxehealth Limited · 2021
  2. 1 generation backK211906 Vital SignsOxehealth Limited · 2021
  3. This deviceK254169 SmartSpectra Vital Signs Monitor 1.0Presage Security, Inc. d/b/a Presage Technologies, Inc. · 2026
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2DEN200019Oxehealth Vital SignsOxehealth Limited2021—
−1K211906Vital SignsOxehealth Limited2021—
0 (this device)K254169SmartSpectra Vital Signs Monitor 1.0Presage Security, Inc. d/b/a Presage Technologies, Inc.2026—
FIG. 1 — Lineage of K254169, 2021–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in QME? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new QME device

  1. #1 K211906 Vital Signsscore 28.7 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K211906Vital Signs
Summary page 5
…-Based Measurement Of Pulse Rate, Heart Rate, Breathing Rate, And/Or Respiratory Rate Regulation Number: 870.2785 Product Code(s): QME Legally Marketed Predicate Device K211906 Oxehealth Vital Signs Device Description Summary The SmartSpectra device is a software development kit (“SDK”) which uses s…
Oxehealth Limited2021—

Also cleared against K211906: 8 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 2 companies; a new generation was cleared every 4.9 years on median.
  • None of the 2 ancestors has a recall.
  • 8 other devices also cite its predicate K211906.

How does its category compare?

All QME devices

QME Software For Optical Camera-Based Measurement Of Pulse Rate, Heart Rate, Breathing Rate, And/Or Respiratory Rate: FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · QME

4so far in 2026, the most since 2007

2021: 22022: 12023: 12024: 32025: 22026 (partial year): 4
2007high 4 (2026) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · QME

266in 2024

2024: 266 days median2026 (partial year): 190 days median
2007high 266 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · QME

No recall names QME since 2007.

Adverse event reports per year · QME

No MAUDE report names QME since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20130—00
20140—00
20150—00
20160—00
20170—00
20180—00
20190—00
20200—00
20212—00
20221—00
20231—00
2024326600
20252—00
202641900—

What is it indicated for?

The Presage SmartSpectra Vital Signs Monitor is intended for non-invasive spot measurement of pulse rate and respiratory rate in healthy individuals during routine interactions with healthcare professionals such as during routine checkups. It is software assessing video footage from a fixed-installation solution (smartphone mounted in a static-mount) in admissions and examination areas of hospitals, clinics, general care, or other secured environments…

Read the full statement (153 words)

The Presage SmartSpectra Vital Signs Monitor is intended for non-invasive spot measurement of pulse rate and respiratory rate in healthy individuals during routine interactions with healthcare professionals such as during routine checkups. It is software assessing video footage from a fixed-installation solution (smartphone mounted in a static-mount) in admissions and examination areas of hospitals, clinics, general care, or other secured environments, where a single human face can be maintained in view, with professional healthcare oversight. The SmartSpectra system is intended for use by trained healthcare professionals and should not be used by untrained users. The SmartSpectra Vital Signs Monitor is intended for use on humans 18 to 70 years old who do not require critical care or continuous vital signs monitoring. The device is not intended to be the sole method of checking the physical health of a subject or for use on patients that have been admitted to hospitals or in ICUs.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K254169: SmartSpectra Vital Signs Monitor 1.0.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K254169

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded Oct 30, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026