510(k) Report

510(k) · Cleared

Cleared

Carescape Canvas 1000; Carescape Canvas Smart Display

K260343510(k) TraditionalCleared Aug 28, 2026207-day reviewClass II21 CFR 870.1025

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

GE Healthcare Finland Oy · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Aug 28, 2026 as substantially equivalent to K223531 (GE Healthcare Finland Oy, 2023) after a 207-day review. Its family tree reaches back 9 generations to 2000.

Decided
Aug 28, 2026
Review
207 days
Class
Class II
Product code
MHX
Recalls
None
Predicates
1

Its longest line of descent runs back 9 generations to K000815 (2000). 16 ancestors in all, the oldest K900598 (1990). No later device cites it as a predicate yet.

  1. Root · 9 generations backK000815 AS/3 Anesthesia Montior with S-ANE99(a)/L-ARK99(a)Datex-Ohmeda · 2000
  2. 5 generations between
  3. 3 generations backK191323 Carescape B850GE Healthcare Finland Oy · 2020 · Recalled
  4. 2 generations backK213336 Carescape B850, E-musbGE Healthcare Finland Oy · 2022 · Recalled
  5. 1 generation backK223531 CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, CARESCAPE…GE Healthcare Finland Oy · 2023 · Recalled
  6. This deviceK260343 Carescape Canvas 1000; Carescape Canvas Smart DisplayGE Healthcare Finland Oy · 2026
Show all 9 generations
  1. Root · 9 generations backK000815 AS/3 Anesthesia Montior with S-ANE99(a)/L-ARK99(a)Datex-Ohmeda · 2000
  2. 8 generations backK021279 Datex-Ohmeda S/5 Anesthesia Monitor with L-ANE02 and L-ANE02A SoftwareDatex-Ohmeda · 2002
  3. 7 generations backK030812 Datex-Ohmeda S/5 Anesthesia Monitor with L-AN03 and L-ANE03A SoftwareDatex-Ohmeda · 2003
  4. 6 generations backK051400 Datex-Ohmeda S/5 Anesthesia Monitor with L-ANEO5 and L-ANEO5A…GE Healthcare · 2006
  5. 5 generations backK092027 Carescape Monitor B850Ge Medical Systems Information Technologies · 2009 · Recalled
  6. 4 generations backK131414 Carescape Monitor B850GE Healthcare Finland Oy · 2013 · Recalled
  7. 3 generations backK191323 Carescape B850GE Healthcare Finland Oy · 2020 · Recalled
  8. 2 generations backK213336 Carescape B850, E-musbGE Healthcare Finland Oy · 2022 · Recalled
  9. 1 generation backK223531 CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, CARESCAPE…GE Healthcare Finland Oy · 2023 · Recalled
  10. This deviceK260343 Carescape Canvas 1000; Carescape Canvas Smart DisplayGE Healthcare Finland Oy · 2026
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K000815AS/3 Anesthesia Montior with S-ANE99(a)/L-ARK99(a)Datex-Ohmeda2000—
−8K021279Datex-Ohmeda S/5 Anesthesia Monitor with L-ANE02 and L-ANE02A SoftwareDatex-Ohmeda2002—
−7K030812Datex-Ohmeda S/5 Anesthesia Monitor with L-AN03 and L-ANE03A SoftwareDatex-Ohmeda2003—
−6K051400Datex-Ohmeda S/5 Anesthesia Monitor with L-ANEO5 and L-ANEO5A Software, Using…GE Healthcare2006—
−5K092027Carescape Monitor B850Ge Medical Systems Information Technologies20099
−4K131414Carescape Monitor B850GE Healthcare Finland Oy20133
−3K191323Carescape B850GE Healthcare Finland Oy20201
−2K213336Carescape B850, E-musbGE Healthcare Finland Oy20221
−1K223531CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, CARESCAPE Canvas D19…GE Healthcare Finland Oy20233
0 (this device)K260343Carescape Canvas 1000; Carescape Canvas Smart DisplayGE Healthcare Finland Oy2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199010
199310
200010
200110
200210
200410
200710
FIG. 1 — Lineage of K260343, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K223531CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, CARESCAPE Canvas D19, F2 Frame…
Summary page 8
…CCL, GWJ, GWQ, KOI, NHO, NHP, NHQ, OLT, OLW, OMC, ORT Predicate Device(s) Primary predicate: The predicate for this submission is K223531 (807.92(a)(3)): CARESCAPE Canvas 1000 with CARESCAPE software version 3.3, and…
GE Healthcare Finland Oy20233
†K211561INVOS PM7100 Patient Monitor, INVOS Pediatric rSO2 Sensor, INVOS Infant Regional…Covidien, LLC20211

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 4 companies; a new generation was cleared every 2.7 years on median.
  • 9 of the 16 ancestors have a recall of any class.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

Carescape Canvas 1000 is a multi-parameter patient monitor intended for use in multiple areas within a professional healthcare facility. Carescape Canvas 1000 is intended for use on adult, pediatric, and neonatal patients one patient at a time. Carescape Canvas 1000 is indicated for monitoring of: • hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse…

Read the full statement (514 words)

Carescape Canvas 1000 is a multi-parameter patient monitor intended for use in multiple areas within a professional healthcare facility.

Carescape Canvas 1000 is intended for use on adult, pediatric, and neonatal patients one patient at a time. Carescape Canvas 1000 is indicated for monitoring of:

• hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, regional oxygen saturation, total hemoglobin concentration, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation),

• respiratory (impedance respiration, airway gases (CO2, 02, N20O, and anesthetic agents), spirometry, gas exchange), and

• neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).

Carescape Canvas 1000 is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, irregular, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, tachycardia, trigeminy, ventricular bradycardia, ventricular fibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. Carescape Canvas 1000 also shows alarms from other ECG sources.

Carescape Canvas 1000 also provides other alarms, trends, snapshots and events, and calculations and can be connected to displays, printers and recording devices.

Carescape Canvas 1000 can interface to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.

Carescape Canvas 1000 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. Carescape Canvas 1000 is not intended for use in a controlled MRI environment.

Carescape Canvas Smart Display is a multi-parameter patient monitor intended for use in multiple areas within a professional healthcare facility.

Carescape Canvas Smart Display is intended for use on adult, pediatric, and neonatal patients one patient at a time.

Carescape Canvas Smart Display is indicated for monitoring of:

• hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, regional oxygen saturation, total hemoglobin concentration, cardiac output (thermodilution), and temperature, and

• respiratory (impedance respiration, airway gases (CO2)

Carescape Canvas Smart Display is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, irregular, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, tachycardia, trigeminy, ventricular bradycardia, ventricular tibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. Carescape Canvas Smart Display also shows alarms from other ECG sources.

Carescape Canvas Smart Display also provides other alarms, trends, snapshots and events.

Carescape Canvas Smart Display can use Carescape ONE or Carescape Patient Data Module (PDM) as patient data acquisition devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.

Carescape Canvas Smart Display is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility. Carescape Canvas Smart Display is not intended for use in a controlled MRI environment.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K260343: Carescape Canvas 1000; Carescape Canvas Smart Display.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K260343

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.