510(k) Report

510(k) · Cleared

Cleared

AMI™ Plasma Cartridge, AMI™

K260613510(k) TraditionalCleared Sep 1, 2026188-day reviewClass I21 CFR 862.1065

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sequitur Health Corp. · JID Photometric Method, Ammonia

Cleared on Sep 1, 2026 as substantially equivalent to K820723 (Eastman Kodak Company, 1982) and 1 other predicate after a 188-day review. Its family tree reaches back 1 generation to 1981.

Decided
Sep 1, 2026
Review
188 days
Class
Class I
Product code
JID
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K811919 (1981). 2 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1982: 1 other ancestorK811919 · Kodak Ektachem 400 Analyzer · Eastman Kodak Company · 198119801990200020102020
Other ancestors, by generation 1 generation back: 1 other ancestor K811919 · Kodak Ektachem 400 Analyzer · Eastman Kodak Company · 19810
  1. Root · 1 generation backK811919 Kodak Ektachem 400 AnalyzerEastman Kodak Company · 1981
  2. This deviceK260613 AMI™ Plasma Cartridge, AMI™Sequitur Health Corp. · 2026
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K811919Kodak Ektachem 400 AnalyzerEastman Kodak Company1981—
0 (this device)K260613AMI™ Plasma Cartridge, AMI™Sequitur Health Corp.2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198210
FIG. 1 — Lineage of K260613, 1981–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JID? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K820723Kodak Ektachem Clinical Chem. NH3
Summary page 5
…1065 Product Code(s) JJE, JID I Legally Marketed Predicate Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K820723 KODAK EKTACHEM CLINICAL CHEMISTRY SLIDES (NH3) JID K811919 KODAK EKTACHEM 400 ANALYZER JJE Device Description Summary 21 CFR 807.…
Eastman Kodak Company19821
*K811919Kodak Ektachem 400 Analyzer
Summary page 5
…Devices 21 CFR 807.92(a)(3) Predicate # Predicate Trade Name (Primary Predicate is listed first) Product Code K820723 KODAK EKTACHEM CLINICAL CHEMISTRY SLIDES (NH3) JID K811919 KODAK EKTACHEM 400 ANALYZER JJE Device Description Summary 21 CFR 807.92(a)(4) The AMI™ completes an ammonia measurement through…
Eastman Kodak Company1981—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.

How does its category compare?

All JID devices

JID Photometric Method, Ammonia: FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · JID

1so far in 2026, the most since 2007

2026 (partial year): 1
2007high 1 (2026) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JID

Too few JID 510(k)s in any year since 2007 to give a median.

Recalls per year · JID

1in 2013

2013: 1
2007high 1 (2013) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JID

6in 2025

2008: 42009: 62010: 42011: 152012: 32013: 182014: 92015: 402016: 102017: 82018: 62019: 132020: 262021: 192022: 122023: 262024: 52025: 6
2007high 40 (2015) · low 0 (2007)2025

MAUDE reports naming product code JID, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—04
20090—06
20100—04
20110—015
20120—03
20130—118
20140—09
20150—040
20160—010
20170—08
20180—06
20190—013
20200—026
20210—019
20220—012
20230—026
20240—05
20250—06
20261—0—

What is it indicated for?

The AMI™ Plasma Cartridge is a photometric in vitro test for the quantitative determination of ammonia in human plasma on the AMI™ analyzer. Ammonia measurements are used in the diagnosis and treatment of severe liver disorders, such as cirrhosis, hepatitis, and Reye's syndrome.

The AMI™ is an analyzer for in vitro diagnostic use in clinical laboratories, utilizing photometric technology for analyte detection.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K260613: AMI™ Plasma Cartridge, AMI™.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K260613

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded Oct 30, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026