510(k) Report

510(k) · Cleared

Cleared

Gastrointestinal Videoscope Olympus GIF-2T1500

K260824510(k) TraditionalCleared Aug 27, 2026167-day reviewClass II21 CFR 876.1500

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Olympus Medical Systems Corporation · FDS Gastroscope And Accessories, Flexible/Rigid

Cleared on Aug 27, 2026 as substantially equivalent to K221551 (Fujifilm Corporaton, 2023) after a 167-day review. Its family tree reaches back 6 generations to 1995.

Decided
Aug 27, 2026
Review
167 days
Class
Class II
Product code
FDS
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K944759 (1995). 8 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 6 generations backK944759 200 Series Gastro-Intestinal Video SystemFujinon, Inc. · 1995 · Recalled
  2. 5 generations backK041903 Fujinon G5 Colonoscopes, Models EC-450HL5 and EC-250HL5Fujinon, Inc. · 2004 · Recalled
  3. 4 generations backK112391 Fujinon ColonoscopesFujinon, Inc. · 2012
  4. 3 generations backK142629 EG-530CT, EG-530D, EC-530DL, and ES-530WE EndoscopesFujifilm Medical System U.S.A., Inc. · 2015
  5. 2 generations backK210162 Endoscope Model EI-740D/SFujifilm Corporation · 2021
  6. 1 generation backK221551 FUJIFILM Endoscope Model EI-740D/SFujifilm Corporaton · 2023
  7. This deviceK260824 Gastrointestinal Videoscope Olympus GIF-2T1500Olympus Medical Systems Corporation · 2026
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K944759200 Series Gastro-Intestinal Video SystemFujinon, Inc.19955
−5K041903Fujinon G5 Colonoscopes, Models EC-450HL5 and EC-250HL5Fujinon, Inc.20041
−4K112391Fujinon ColonoscopesFujinon, Inc.2012—
−3K142629EG-530CT, EG-530D, EC-530DL, and ES-530WE EndoscopesFujifilm Medical System U.S.A., Inc.2015—
−2K210162Endoscope Model EI-740D/SFujifilm Corporation2021—
−1K221551FUJIFILM Endoscope Model EI-740D/SFujifilm Corporaton2023—
0 (this device)K260824Gastrointestinal Videoscope Olympus GIF-2T1500Olympus Medical Systems Corporation2026—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
200410
FIG. 1 — Lineage of K260824, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FDS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FDS device

  1. #1 K251708 EvoEndo Single-Use Endoscopy Systemscore 16.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K221551FUJIFILM Endoscope Model EI-740D/S
Summary page 5
…ssification Class II III. Predicate Device Substantial equivalence is claimed to the following device: • FUJIFILM Endoscope Model EI-740D/S, K221551, Fujifilm Corporation IV. Device Description The Gastrointestinal Videoscope Olympus GIF-2T1500 (GIF-2T1500) is a non-st…
Fujifilm Corporaton2023—
†K112680Evis Exera Iii Video SystemOlympus Medical Systems Corporation20126
†K232997EVIS Exera III Gastrointestinal Videoscope GIF-1TH190Olympus Medical Systems Corporation20242

Also cleared against K221551: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 31 years and passes through 5 companies; a new generation was cleared every 5.9 years on median.
  • 3 of the 8 ancestors have a recall of any class.
  • 1 other device also cites its predicate K221551.

How does its category compare?

All FDS devices

FDS Gastroscope And Accessories, Flexible/Rigid: FDA decisions, review times, recalls and adverse event reports over 20 years. 2026, this device’s year, is highlighted.

FDA decisions per year · FDS

8in 2025

2007: 22010: 12011: 12012: 32013: 22014: 22015: 22016: 12017: 12018: 62019: 52020: 22021: 52022: 22023: 32024: 52025: 82026 (partial year): 7
2007high 8 (2025) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FDS

211in 2025

2012: 181 days median2018: 167 days median2019: 175 days median2021: 78 days median2023: 133 days median2024: 220 days median2025: 211 days median2026 (partial year): 164 days median
2007high 220 (2024) · low 78 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FDS

2in 2024

2013: 12016: 12020: 22024: 2
2007high 2 (2020) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FDS

18,905in 2025

2007: 272008: 112009: 212010: 452011: 362012: 352013: 82014: 232015: 292016: 1082017: 1782018: 3042019: 1,3762020: 6012021: 1,7552022: 6,2162023: 9,8742024: 9,2582025: 18,905
2007high 18,905 (2025) · low 8 (2013)2025

MAUDE reports naming product code FDS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—027
20080—011
20090—021
20101—045
20111—036
20123181035
20132—18
20142—023
20152—029
20161—1108
20171—0178
201861670304
2019517501,376
20202—2601
202157801,755
20222—06,216
2023313309,874
2024522029,258
20258211018,905
202671640—

What is it indicated for?

The GASTROINTESTINAL VIDEOSCOPE OLYMPUS GIF-2T1500 is intended to be used with a video system center, documentation equipment, monitor, endoscopic therapy accessories (such as a biopsy forceps), and other ancillary equipment for endoscopy and endoscopic surgery. This instrument is indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K260824: Gastrointestinal Videoscope Olympus GIF-2T1500.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K260824

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded Oct 30, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026