Shiley Single Cannula Tracheostomy Tube

K810106510(k) TraditionalCleared Feb 2, 198117-day reviewClass II21 CFR 868.5800Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Shiley, Inc. · BTO Tube, Tracheostomy (W/Wo Connector)

Cleared on Feb 2, 1981 after a 17-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 3 of the 14 devices in its family, counting itself, have a Class I recall. 13 later devices descend from it. It has been recalled 6 times.

Decided
Feb 2, 1981
Review
17 days
Class
Class II
Product code
BTO
Recalls
6
Predicates
0

13 later devices descend from it, 5 citing it directly. 3 of the 14 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1997: 2 descendants cleared1996: 2 descendants cleared, 1 with a Class I recall12003: 1 descendant cleared2012: 1 descendant cleared2013: 1 descendant cleared2015: 2 descendants cleared, 2 with a Class I recall22007: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant cleared2020: 1 descendant cleared19801990200020102020
  1. This deviceK810106 Shiley Single Cannula Tracheostomy TubeShiley, Inc. · 1981
  2. Descendant · cites it directly · 7 of its ownK962173 Shikey Tracheostomy Tubes and AccessoriesMallinckrodt Medical · 1996 · Recalled (Class I)
  3. Descendant · cites it directly · 5 of its ownK963732 Percutaneous Dual Cannula Tracheostomy Tube with Low Pressure/Low…Mallinckrodt Medical · 1997 · Recalled
  4. Descendant · 2 generations on · 3 of its ownK142296 Shiley Adult Flexible Tracheostomy Tube Cuffless, Disposable Inner…Covidien · 2015 · Recalled (Class I)
  5. Descendant · cites it directly · 3 of its ownK961449 Tracheostomy Tubes Tracoe-FlexTracoe Medical GmbH · 1996
  6. Descendant · 2 generations on · 2 of its ownK120079 Primed Tracheostomy Tubes (Multiple)Primed Halberstadt Medizintechnik GmbH · 2012

Showing 5 of 13 descendants. See all descendants

Descendants cleared per year

1996: 21997: 22003: 12007: 12012: 12013: 12015: 22020: 12024: 12025: 1
1996high 2 (1996) · low 0 (1998)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 13 · See all
  • K962173Shikey Tracheostomy Tubes and AccessoriesMallinckrodt Medical · 1996 · cites it directly7descendants
  • K963732Percutaneous Dual Cannula Tracheostomy Tube with Low Pressure/Low Profile Cuff and…Mallinckrodt Medical · 1997 · cites it directly5descendants
  • K142296Shiley Adult Flexible Tracheostomy Tube Cuffless, Disposable Inner Cannula; Shiley Adult…Covidien · 2015 · 2 generations on3descendants
  • K961449Tracheostomy Tubes Tracoe-FlexTracoe Medical GmbH · 1996 · cites it directly3descendants
  • K120079Primed Tracheostomy Tubes (Multiple)Primed Halberstadt Medizintechnik GmbH · 2012 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K810106Shiley Single Cannula Tracheostomy TubeShiley, Inc.1981—
+1K962173Shikey Tracheostomy Tubes and AccessoriesMallinckrodt Medical1996Class I
+1K963732Percutaneous Dual Cannula Tracheostomy Tube with Low Pressure/Low Profile Cuff…Mallinckrodt Medical19972
+2K142296Shiley Adult Flexible Tracheostomy Tube Cuffless, Disposable Inner Cannula…Covidien2015Class I
+1K961449Tracheostomy Tubes Tracoe-FlexTracoe Medical GmbH1996—
+2K120079Primed Tracheostomy Tubes (Multiple)Primed Halberstadt Medizintechnik GmbH2012—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199621
199720
200310
200710
201210
201310
201522
202010
202410
202510
FIG. 1 — Lineage of K810106, 1981–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BTO device

  1. #1 K242921 5300600-5301000 Rota-Trach Disposable Standard…score 13.7 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (13)

See all 13

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K962173Shikey Tracheostomy Tubes and Accessories Class IMallinckrodt Medical19967
K963732Percutaneous Dual Cannula Tracheostomy Tube with Low Pressure/Low Profile Cuff and…Mallinckrodt Medical19975
K961449Tracheostomy Tubes Tracoe-FlexTracoe Medical GmbH19963
K030787Shiley Flextra Tracheostomy TubeNellcor Puritan Bennett, Inc.20030
K971267Shiley Extended Length, Disposable Cannula, Tracheostomy Tube: Cuffed, Distal Extension…Mallinckrodt Medical19970

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 13, 2010Not classifiedZ-1484-2010Process designInflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
Apr 13, 2010Not classifiedZ-1483-2010Process designInflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
Apr 13, 2010Not classifiedZ-1485-2010Process designInflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
Apr 13, 2010Not classifiedZ-1482-2010Process designInflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
Apr 13, 2010Not classifiedZ-1486-2010Process designInflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.
Apr 13, 2010Not classifiedZ-1487-2010Process designInflatable cuff Tracheostomy tubes may leak air, resulting in inadequate ventilation.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K962173Shikey Tracheostomy Tubes and AccessoriesMallinckrodt Medical · 1996 · Cites it directly · ◆ in the 1996 descendants bar of FIG. 11 Class I recall
  • K142296Shiley Adult Flexible Tracheostomy Tube Cuffless, Disposable Inner Cannula; Shiley Adult Flexible…Covidien · 2015 · 2 generations on · ◆ in the 2015 descendants bar of FIG. 11 Class I recall
  • K150844Shiley Adult Flexible Tracheostomy Tube Cuffless, Reusable Inner Cannula, Shiley Adult Flexible Tracheostomy…Covidien · 2015 · 2 generations on · ◆ in the 2015 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K962173 (Mallinckrodt Medical, 1996) has gone furthest, with 7 descendants of its own.

How does its category compare?

All BTO devices

BTO Tube, Tracheostomy (W/Wo Connector): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BTO

1in 2025

2007: 12008: 12009: 12010: 32011: 12012: 12018: 12020: 12024: 12025: 12026 (partial year): 1
2007high 3 (2010) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BTO

98in 2010

2010: 98 days median
2007high 98 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BTO

1in 2023

2009: 22010: 62011: 192012: 12015: 12023: 1, 1 Class I2026 (partial year): 1
2007high 19 (2011) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · BTO

233in 2025

2007: 2262008: 2552009: 7952010: 1302011: 542012: 822013: 782014: 1732015: 1722016: 2122017: 1262018: 1622019: 872020: 1132021: 1982022: 7702023: 4022024: 4182025: 233
2007high 795 (2009) · low 54 (2011)2025

MAUDE reports naming product code BTO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—0226
20081—0255
20091—2795
20103986130
20111—1954
20121—182
20130—078
20140—0173
20150—1172
20160—0212
20170—0126
20181—0162
20190—087
20201—0113
20210—0198
20220—0770
20230—1 (1)402
20241—0418
20251—0233
20261—1—
Cite this page

510(k) Report. “K810106: Shiley Single Cannula Tracheostomy Tube.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K810106

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026