510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Kwik Stik (TM) Lyfo (TM) Disk Microorganisms

K861022510(k) TraditionalCleared Apr 4, 198617-day reviewClass I21 CFR 866.2480Recalls 40

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Micro-Bio-Logics · JTR Kit, Quality Control For Culture Media

Cleared on Apr 4, 1986 after a 17-day review. FDA has not posted a summary or statement for it, so its predicates are not known. It has been recalled 40 times.

Decided
Apr 4, 1986
Review
17 days
Class
Class I
Product code
JTR
Recalls
40
Predicates
0

FDA has not posted a summary or statement for it, so its predicates are not known.

FIG. 1 — Lineage of K861022Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JTR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 2, 2025Class IIZ-0984-2026Under Investigation by firmAffected units may fail to recover the target microorganism Campylobacter jejuni subsp. jejuni derived from ATCC 29428.
Dec 2, 2025Class IIZ-0985-2026Under Investigation by firmAffected units may fail to recover the target microorganism Campylobacter jejuni subsp. jejuni derived from ATCC 29428.
Dec 2, 2025Class IIZ-0986-2026Under Investigation by firmAffected units may fail to recover the target microorganism Campylobacter jejuni subsp. jejuni derived from ATCC 29428.
Jun 27, 2025Class IIZ-2194-2025Under Investigation by firmKWIK-STIK REF numbers: 0894P, 0894K and LYFO DISK" 0894L may result in failure to recover the target microorganism. All lots of REF 894-76 are impacted.
Jun 17, 2025Class IIZ-2124-2025Under Investigation by firmAffected products contain Staphylococcus epidermidis instead of the expected Brevundimonas diminuta. This may cause the user's quality control to fail and delayed diagnosis for the patient.
May 27, 2025Class IIZ-1957-2025Mixed-up of materials/componentsAffected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was packaged 01065K swabs; 01065K was packaged with 0621K swabs. This may lead to delayed diagnosis.
Mar 6, 2024Class IIZ-1483-2024Use errorThe positive control material (Rhizopus stolonifer) within the KWIK-STIK assembly was contaminated with another fungal organism (Aspergillus brasiliensis). As a result, users may observe A. brasiliensis growth when using the control…
Oct 18, 2023Class IIZ-0307-2024Under Investigation by firmA complaint investigation showed that the bulk pellets (lot 794-142) were contaminated with Aspergillus flavus.
Oct 18, 2023Class IIZ-0308-2024Under Investigation by firmA complaint investigation showed that the bulk pellets (lot 794-142) were contaminated with Aspergillus flavus.
Aug 14, 2023Class IIIZ-2576-2023Packaging process controlOne lot of KWIK STIK ANA (3 strains) QC Set is missing one of the strains, 6018P Parabacteroides distasonis.
May 6, 2022Class IIIZ-1125-2022Labeling mix-upsLabeling error.
Feb 24, 2021Class IIIZ-0543-2022Labeling mix-upsPotential that products packaged for lot 726-57-10 were packaged in pouches labeled as 323-108-7.
Feb 24, 2021Class IIIZ-0544-2022Labeling mix-upsPotential of products packaged for lot 323-108-7 were packaged in pouches labeled as 726-57-10.
Dec 16, 2020Class IIIZ-0546-2022Employee errorThe expiration date on the label of one component (negative control 0243U) in the set lot is incorrectly labeled. The KWIK-STIK negative control 0243U was labeled with an expiration date of 7/31/2020 when it should have been 7/31/2022.
Dec 16, 2020Class IIZ-0511-2022Material/Component ContaminationContamination with S. epidermidis
Oct 8, 2020Class IIZ-0302-2021Under Investigation by firmKWIK-STIK 2 Pack Klebsiella pneumoniae derived from NCTC 13443 (Catalog# 01145P) should contain the NDM-1 positive strain, but was confirmed to be NDM-1 negative.
Oct 8, 2020Class IIZ-0303-2021Under Investigation by firmKWIK-STIK 6 Pack Klebsiella pneumoniae derived from NCTC 13443 (Catalog# 01145K) should contain the NDM-1 positive strain, but was confirmed to be NDM-1 negative.
Oct 8, 2020Class IIZ-0304-2021Under Investigation by firmQC Sets and Panels: KWIK-STIK 2 Pack (Catalog# 5226P) should contain the NDM-1 positive strain, but was confirmed to be NDM-1 negative.
Oct 8, 2020Class IIZ-0402-2021Under Investigation by firmStrain should be VIM-1 positive but is confirmed to be VIM-1 negative.
Oct 8, 2020Class IIIZ-0581-2022Employee errorThe incorrect lot number was printed on the foil pouch that the KWIK-STIK" is placed in. All foil pouches for lot 318-234-4 have the lot number misprinted as 218-234-4. The correct lot number was printed on the KWIK-STIK" canister label…
Oct 1, 2020Class IIZ-0582-2022Unknown/Undetermined by firmPotential contamination with S. epidermidis E. coli and S. warneri.
Oct 1, 2020Class IIZ-0583-2022Unknown/Undetermined by firmPotential contamination with Escherichia coli, Staphylococcus epidermidis and S. warneri.
Oct 1, 2020Class IIZ-0584-2022Unknown/Undetermined by firmPotential contamination with Escherichia coli Staphylococcus epidermidis and S. warneri.
Feb 18, 2020Class IIZ-0587-2022Process designTests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Feb 18, 2020Class IIZ-0589-2022Process designTests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Feb 18, 2020Class IIZ-0586-2022Process designTests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Feb 18, 2020Class IIZ-0588-2022Process designTests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Feb 18, 2020Class IIZ-0585-2022Process designTests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Feb 12, 2020Class IIZ-1691-2020Nonconforming Material/ComponentKWIKSTIK 2 Pack (REF 0778P) and QC Set and Panels: KWIKSTIK YST Comprehensive QC Set (REF 5195P) contain product labeled as Trichosporon dermatis (T. dermatis) which may contain Candida Utilis instead of T. dermatis.
Feb 12, 2020Class IIZ-1690-2020Nonconforming Material/ComponentKWIKSTIK 2 Pack (REF 0778P) and QC Set and Panels: KWIKSTIK YST Comprehensive QC Set (REF 5195P) contain product labeled as Trichosporon dermatis (T. dermatis) which may contain Candida Utilis instead of T. dermatis.
Mar 2, 2017Class IIIZ-0636-2022Labeling False and MisleadingPotential for product labeled as Actinomyces odontolyticus is actually Eggerthella lenta.
Mar 2, 2017Class IIIZ-0637-2022Labeling False and MisleadingPotential for product labeled as Actinomyces odontolyticus is actually Eggerthella lenta.
Dec 1, 2015Class IIZ-0704-2022Material/Component ContaminationThe Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in some cases. This…
Dec 1, 2015Class IIZ-0703-2022Material/Component ContaminationThe Lactobacillus acidophilus negative control (item 0243 single KWIK-STIK) contained in the 8164 Cepheid Xpert GBS QC Set and 8165 Cepheid Smart GBS QC Set was found to test positive for Group B Streptococcus (GBS) in some cases. This…
Jul 17, 2015Class IIIZ-0788-2022Labeling mix-upsLYFO Disk Catalog #0983L is to contain Candida tropicalis. However, lot number 983-40-3 under Catalog #0983L, labeled as Candida tropicalis, is actually Yersinia entercolitica. This labeling error may cause a user's quality control to…
Oct 6, 2014Class IIZ-0712-2022Nonconforming Material/ComponentThis strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
Oct 6, 2014Class IIZ-0713-2022Nonconforming Material/ComponentThis strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
Oct 6, 2014Class IIZ-0714-2022Nonconforming Material/ComponentThis strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.
Mar 25, 2013Class IIIZ-0641-2022Labeling mix-upsSome of the KWIK-STIK Catalog #01023P Campylobacter coli, Lot# 1023-05-7, contained an inner label stating the product is Catalog #0919K Haemophilus influenza. The organism has been confirmed to be C. coli, indicating this is a labeling…
Oct 12, 2007Not classifiedZ-0456-2008Employee errorMislabeled: DuoPacks of Proteus vulgaris contained Listeria monocytogenes instead of the labeled Proteus vulgaris.

No ancestor or descendant has a Class I recall.

How does its category compare?

All JTR devices

JTR Kit, Quality Control For Culture Media: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JTR

FDA has decided no JTR device since 2007.

Median 510(k) review, days · JTR

Too few JTR 510(k)s in any year since 2007 to give a median.

Recalls per year · JTR

4in 2025

2007: 42013: 12014: 32015: 32017: 22020: 152021: 22023: 22025: 4
2007high 15 (2020) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JTR

1in 2015

2015: 1
2007high 1 (2015) · low 0 (2007)2025

MAUDE reports naming product code JTR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—40
20080—00
20090—00
20100—00
20110—00
20120—00
20130—10
20140—30
20150—31
20160—00
20170—20
20180—00
20190—00
20200—150
20210—20
20220—00
20230—20
20240—00
20250—40
20260—0—
Cite this page

510(k) Report. “K861022: Kwik Stik (TM) Lyfo (TM) Disk Microorganisms.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K861022

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026