# K905235: VDR Sinusoidal Percusionator Family/Relate Access

- Pathway: 510(k) (Traditional)
- Applicant: Percussionaire Corp.
- Decided: May 3, 1991 after 177 days
- Product code: CBK (Ventilator, Continuous, Facility Use)
- Device class: 2
- Regulation: 21 CFR 868.5895
- Recalls of this device: 3 (2 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
1 of 1 device in its family (counting itself) has a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA has not posted a summary or statement for it, so its predicates are not known.

## Recalls of this device
- Dec 23, 2024 · Class I · Z-0960-2025: Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.
- Mar 28, 2023 · Class I · Z-0693-2024: Breathing circuit reservoir bag may be assembled at the wrong attachment point, which will result in less than optimal therapy.
- Sep 14, 2020 · Class II · Z-0259-2024: Note this recall occurred in 2020 and 2021. Reports of screen freeze on visual displays of waveform analyzers that are paired with ventilators. Subsequently, waveform analyzers were serviced to instal