E-Z-EM Pulsed Infusion System

K905447510(k) TraditionalCleared Apr 1, 1991117-day reviewClass II21 CFR 870.1210Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

E-Z-Em, Inc. · KRA Catheter, Continuous Flush

Cleared on Apr 1, 1991 after a 117-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 4 of the 72 devices in its family, counting itself, have a Class I recall. 71 later devices descend from it. It has been recalled 4 times.

Decided
Apr 1, 1991
Review
117 days
Class
Class II
Product code
KRA
Recalls
4
Predicates
0

71 later devices descend from it, 1 citing it directly. 4 of the 72 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2004: 1 descendant cleared2007: 4 descendants cleared2005: 2 descendants cleared2006: 1 descendant cleared2013: 2 descendants cleared2009: 7 descendants cleared2010: 5 descendants cleared2012: 3 descendants cleared2011: 5 descendants cleared2014: 4 descendants cleared2015: 2 descendants cleared2018: 3 descendants cleared, 1 with a Class I recall12016: 1 descendant cleared2020: 2 descendants cleared2017: 1 descendant cleared2021: 6 descendants cleared, 1 with a Class I recall12023: 9 descendants cleared, 1 with a Class I recall12025: 5 descendants cleared, 1 with a Class I recall12026: 3 descendants cleared2022: 2 descendants cleared2024: 3 descendants cleared1990200020102020
  1. This deviceK905447 E-Z-EM Pulsed Infusion SystemE-Z-Em, Inc. · 1991
  2. Descendant · cites it directly · 70 of its ownK040013 Angiojet Xpeedior 120 Catheter and Angiojet Power Pulse Spray…Possis Medical, Inc. · 2004
  3. Descendant · 2 generations on · 68 of its ownK052256 Angiojet Xpeedior 120 CatheterPossis Medical, Inc. · 2005
  4. Descendant · 2 generations on · 66 of its ownK061951 Angiojet Xpeedior Rheolytic Thrombectomy Catheter, Model Xpeedior 120Possis Medical, Inc. · 2006
  5. Descendant · 2 generations on · 63 of its ownK071514 Angiojet Xpeedior Rheolytic Thrombectomy Catheter, Model 104307Possis Medical, Inc. · 2007
  6. Descendant · 3 generations on · 34 of its ownK071342 Angiojet Ultra Xpeedior Thrombectomy Set, Model 105040, 105042…Possis Medical, Inc. · 2007

Showing 5 of 71 descendants. See all descendants

Descendants cleared per year

2004: 12005: 22006: 12007: 42009: 72010: 52011: 52012: 32013: 22014: 42015: 22016: 12017: 12018: 32020: 22021: 62022: 22023: 92024: 32025: 52026: 3
2004high 9 (2023) · low 0 (2008)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 71 · See all
  • K040013Angiojet Xpeedior 120 Catheter and Angiojet Power Pulse Spray Ancillary KitPossis Medical, Inc. · 2004 · cites it directly70descendants
  • K052256Angiojet Xpeedior 120 CatheterPossis Medical, Inc. · 2005 · 2 generations on68descendants
  • K061951Angiojet Xpeedior Rheolytic Thrombectomy Catheter, Model Xpeedior 120Possis Medical, Inc. · 2006 · 2 generations on66descendants
  • K071514Angiojet Xpeedior Rheolytic Thrombectomy Catheter, Model 104307Possis Medical, Inc. · 2007 · 2 generations on63descendants
  • K071342Angiojet Ultra Xpeedior Thrombectomy Set, Model 105040, 105042, 106552, 105043, 105041…Possis Medical, Inc. · 2007 · 3 generations on34descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K905447E-Z-EM Pulsed Infusion SystemE-Z-Em, Inc.1991—
+1K040013Angiojet Xpeedior 120 Catheter and Angiojet Power Pulse Spray Ancillary KitPossis Medical, Inc.2004—
+2K052256Angiojet Xpeedior 120 CatheterPossis Medical, Inc.2005—
+2K061951Angiojet Xpeedior Rheolytic Thrombectomy Catheter, Model Xpeedior 120Possis Medical, Inc.2006—
+2K071514Angiojet Xpeedior Rheolytic Thrombectomy Catheter, Model 104307Possis Medical, Inc.2007—
+3K071342Angiojet Ultra Xpeedior Thrombectomy Set, Model 105040, 105042, 106552…Possis Medical, Inc.2007—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200520
200610
200740
200970
201050
201150
201230
201320
201440
201520
201610
201710
201831
202020
202161
202220
202391
202430
202551
202630
FIG. 1 — Lineage of K905447, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KRA device

  1. #1 K253509 TriSalus TriNav Advance Infusion Systemscore 14.4 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (71)

See all 71

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K040013Angiojet Xpeedior 120 Catheter and Angiojet Power Pulse Spray Ancillary KitPossis Medical, Inc.200470

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 6, 2014Class IIZ-2261-2014Component design/selectionAngioDynamics became aware that Uni*Fuse Infusion Systems with Cooper Wire have the potential to contain an incorrect sized guidewire (too long).
Aug 6, 2014Class IIZ-2260-2014Component design/selectionAngioDynamics became aware that Uni*Fuse Infusion Systems with Cooper Wire have the potential to contain an incorrect sized guidewire (too long).
Aug 6, 2014Class IIZ-2262-2014Component design/selectionAngioDynamics became aware that Uni*Fuse Infusion Systems with Cooper Wire have the potential to contain an incorrect sized guidewire (too long).
May 17, 2010Not classifiedZ-1707-2010Process controlThis lot was packaged with the wrong occluding ball guidewire. The PulseSpray Catheter is 90 cm in length and requires a 109 cm occluding ball guidewire, however this lot was packaged with a 60 cm occluding ball guidewire.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K172315Straub Endovascular SystemStraub Medical AG · 2018 · 7 generations on · ◆ in the 2018 descendants bar of FIG. 11 Class I recall
  • K211738Rotarex Atherectomy SystemC.R. Bard, Inc. · 2021 · 8 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall
  • K242757Rotarex Atherectomy SystemBard Peripheral Vascular, Inc. · 2025 · 9 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
  • K223210ClotTriever XL CatheterInari Medical, Inc. · 2023 · 11 generations on · ◆ in the 2023 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K040013 (Possis Medical, Inc., 2004) has gone furthest, with 70 descendants of its own.

How does its category compare?

All KRA devices

KRA Catheter, Continuous Flush: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KRA

1in 2025

2007: 52008: 72009: 52010: 22011: 22012: 32013: 32014: 42016: 82017: 42018: 42019: 42020: 62021: 62022: 62023: 72025: 12026 (partial year): 5
2007high 8 (2016) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KRA

63in 2023

2007: 44 days median2008: 144 days median2009: 69 days median2012: 29 days median2013: 126 days median2014: 61 days median2016: 122 days median2017: 166 days median2018: 178 days median2019: 133 days median2020: 29 days median2021: 108 days median2022: 228 days median2023: 63 days median2026 (partial year): 268 days median
2007high 268 (2026) · low 29 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KRA

41so far in 2026, the most since 2007

2007: 22008: 62009: 82010: 262011: 32012: 32013: 22014: 3, 2 Class I2015: 12016: 3, 1 Class I2019: 4, 4 Class I2020: 12023: 22026 (partial year): 41
2007high 41 (2026) · low 0 (2017)2026

Class I (7 since 2007) Class II/III not classified

Adverse event reports per year · KRA

980in 2025

2007: 1182008: 1172009: 1872010: 3552011: 2392012: 2172013: 2092014: 2762015: 3392016: 4182017: 5382018: 4342019: 4182020: 3812021: 4402022: 5432023: 8292024: 9422025: 980
2007high 980 (2025) · low 117 (2008)2025

MAUDE reports naming product code KRA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075442118
200871446117
20095698187
20102—26355
20112—3239
20123293217
201331262209
20144613 (2)276
20150—1339
201681223 (1)418
201741660538
201841780434
201941334 (4)418
20206291381
202161080440
202262280543
20237632829
20240—0942
20251—0980
2026526841—
Cite this page

510(k) Report. “K905447: E-Z-EM Pulsed Infusion System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K905447

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026