# K911400: Walkaway TM 40

- Pathway: 510(k) (Traditional)
- Applicant: Baxter Healthcare Corp
- Decided: Nov 19, 1991 after 235 days
- Product code: LRG (Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems)
- Device class: 2
- Regulation: 21 CFR 866.1640
- Recalls of this device: 11 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA has not posted a summary or statement for it, so its predicates are not known.

## Recalls of this device
- Oct 21, 2019 · Class II · Z-1084-2022: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Oct 21, 2019 · Class II · Z-1085-2022: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Oct 21, 2019 · Class II · Z-1088-2022: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Oct 21, 2019 · Class II · Z-1087-2022: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Oct 21, 2019 · Class II · Z-1083-2022: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Oct 21, 2019 · Class II · Z-1086-2022: WalkAway Instruments may have been assembled with the incorrect AC power entry connector inlet with is rated at 1 ampere current rating rather than the correct current rating of 10 ampere
- Sep 4, 2019 · Class II · Z-1370-2021: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
- Sep 4, 2019 · Class II · Z-1371-2021: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
- Sep 4, 2019 · Class II · Z-1372-2021: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
- Sep 4, 2019 · Class II · Z-1373-2021: Instruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
- Jun 2, 2014 · Class II · Z-1990-2014: Springs contained in the access door hinge assembly on the Walk Away Plus instruments are becoming detached causing premature failure of the hinge assembly and a potential safety issue for the custome

Source page: https://510kreport.com/device/K911400