510(k) Report

510(k) · Cleared

Cleared

SJM/Cad Isoflow (TM) Centrifugal Pump 2100CP, Modif

K912604510(k) TraditionalCleared Oct 28, 1991150-day reviewClass II21 CFR 870.4360

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Aries Medical, Inc. · KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type

Cleared on Oct 28, 1991 after a 150-day review. FDA has not posted a summary or statement for it, so its predicates are not known.

Decided
Oct 28, 1991
Review
150 days
Class
Class II
Product code
KFM
Recalls
None
Predicates
0

FDA has not posted a summary or statement for it, so its predicates are not known.

FIG. 1 — Lineage of K912604Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KFM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KFM device

  1. #1 K250326 Quantum Perfusion Centrifugal Blood Pump…score 11.4 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All KFM devices

KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KFM

2in 2025

2008: 42009: 52010: 32011: 72013: 22017: 12019: 32020: 32021: 32023: 32025: 2
2007high 7 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KFM

30in 2023

2008: 194 days median2009: 184 days median2010: 109 days median2011: 16 days median2019: 147 days median2020: 30 days median2021: 534 days median2023: 30 days median
2007high 534 (2021) · low 16 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KFM

2in 2024

2008: 22009: 42010: 22011: 12012: 52018: 12019: 3, 1 Class I2024: 2
2007high 5 (2012) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · KFM

249in 2025

2007: 602008: 622009: 852010: 812011: 572012: 822013: 1042014: 3472015: 2992016: 1562017: 1212018: 1762019: 3112020: 4372021: 4462022: 2942023: 3462024: 2962025: 249
2007high 446 (2021) · low 57 (2011)2025

MAUDE reports naming product code KFM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—060
20084194262
20095184485
20103109281
2011716157
20120—582
20132—0104
20140—0347
20150—0299
20160—0156
20171—0121
20180—1176
201931473 (1)311
20203300437
202135340446
20220—0294
20233300346
20240—2296
20252—0249
20260—0—
Cite this page

510(k) Report. “K912604: SJM/Cad Isoflow (TM) Centrifugal Pump 2100CP, Modif.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K912604

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026