510(k) Report

510(k) · Cleared

Recalled · 2010

Varian C-Series Clinacs (Clinac 2300C/D)

K913119510(k) TraditionalCleared Nov 7, 1991115-day reviewClass II21 CFR 892.5050Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Varian Assoc., Inc. · IYE Accelerator, Linear, Medical

Cleared on Nov 7, 1991 after a 115-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 89 later devices descend from it. It has been recalled 6 times.

Decided
Nov 7, 1991
Review
115 days
Class
Class II
Product code
IYE
Recalls
6
Predicates
0

89 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2002: 1 descendant cleared2003: 1 descendant cleared2007: 2 descendants cleared2006: 2 descendants cleared2009: 6 descendants cleared2008: 1 descendant cleared2012: 3 descendants cleared2010: 3 descendants cleared2011: 4 descendants cleared2013: 5 descendants cleared2024: 3 descendants cleared2014: 4 descendants cleared2017: 7 descendants cleared2018: 7 descendants cleared2016: 3 descendants cleared2019: 5 descendants cleared2015: 3 descendants cleared2022: 4 descendants cleared2021: 7 descendants cleared2026: 5 descendants cleared2023: 9 descendants cleared2025: 2 descendants cleared2020: 2 descendants cleared1990200020102020
  1. This deviceK913119 Varian C-Series Clinacs (Clinac 2300C/D)Varian Assoc., Inc. · 1991
  2. Descendant · cites it directly · 87 of its ownK033343 Trilogy Radiotherapy Delivery SystemVarian Medical Systems, Inc. · 2003 · Recalled
  3. Descendant · 2 generations on · 86 of its ownK061140 Trilogy Radiotherapy Delivery SystemVarian Medical Systems · 2006
  4. Descendant · 2 generations on · 78 of its ownK070094 Trilogy Radiotherapy Delivery System and Trilogy TX Delivery SystemVarian Medical Systems, Inc. · 2007 · Recalled
  5. Descendant · 3 generations on · 77 of its ownK072916 Trilogy System with RapidarcVarian Medical Systems, Inc. · 2007 · Recalled
  6. Descendant · 4 generations on · 61 of its ownK092871 Trilogy MXVarian Medical Systems, Inc. · 2009 · Recalled

Showing 5 of 89 descendants. See all descendants

Descendants cleared per year

2002: 12003: 12006: 22007: 22008: 12009: 62010: 32011: 42012: 32013: 52014: 42015: 32016: 32017: 72018: 72019: 52020: 22021: 72022: 42023: 92024: 32025: 22026: 5
2002high 9 (2023) · low 0 (2004)2026

Most influential descendants

Showing 5 of 89 · See all
  • K033343Trilogy Radiotherapy Delivery SystemVarian Medical Systems, Inc. · 2003 · cites it directly87descendants
  • K061140Trilogy Radiotherapy Delivery SystemVarian Medical Systems · 2006 · 2 generations on86descendants
  • K070094Trilogy Radiotherapy Delivery System and Trilogy TX Delivery SystemVarian Medical Systems, Inc. · 2007 · 2 generations on78descendants
  • K072916Trilogy System with RapidarcVarian Medical Systems, Inc. · 2007 · 3 generations on77descendants
  • K092871Trilogy MXVarian Medical Systems, Inc. · 2009 · 4 generations on61descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K913119Varian C-Series Clinacs (Clinac 2300C/D)Varian Assoc., Inc.1991—
+1K033343Trilogy Radiotherapy Delivery SystemVarian Medical Systems, Inc.20033
+2K061140Trilogy Radiotherapy Delivery SystemVarian Medical Systems2006—
+2K070094Trilogy Radiotherapy Delivery System and Trilogy TX Delivery SystemVarian Medical Systems, Inc.20075
+3K072916Trilogy System with RapidarcVarian Medical Systems, Inc.20072
+4K092871Trilogy MXVarian Medical Systems, Inc.20095

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200210
200310
200620
200720
200810
200960
201030
201140
201230
201350
201440
201530
201630
201770
201870
201950
202020
202170
202240
202390
202430
202520
202650
FIG. 1 — Lineage of K913119, 1991–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (89)

See all 89

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K033343Trilogy Radiotherapy Delivery SystemVarian Medical Systems, Inc.200387
K023636Blocking TraySoutheastern Radiation Products, Inc.20020

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 23, 2010Not classifiedZ-1249-2011Device DesignPosition errors may result in misdelivery of radiation.
Jun 10, 2010Not classifiedZ-0886-2011OtherThe Auto goto or Auto setup functions in C-Series version 7 software ignore the couch angle and exceed motion zone (unless a tolerance is defined by the user); it has the potential for collision with the patient on the couch.
Oct 8, 2009Not classifiedZ-0910-2011Nonconforming Material/ComponentProduct may deliver radiation treatment to areas larger than intended,to healthy tissue.
May 2, 2007Not classifiedZ-1074-2007Process controlThe wedge tray may slip out of the Clinac interface mount during gantry rotation.
Aug 31, 2006Not classifiedZ-0194-2007Process controlThe chain holding the device gantry in position may break.
Mar 30, 2005Not classifiedZ-1530-05Device DesignEmergency Off Switch located on the device may not function correctly resulting in the device remaining powered on after the switch is activated.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K033343 (Varian Medical Systems, Inc., 2003) has gone furthest, with 87 descendants of its own.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—
Cite this page

510(k) Report. “K913119: Varian C-Series Clinacs (Clinac 2300C/D).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K913119

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026