510(k) Report

510(k) · Cleared

Recalled · Class I · 2014

Isothermal Heated Ventilator & Anesthesia Circuit

K915226510(k) TraditionalCleared Jan 12, 1993419-day reviewClass II21 CFR 868.5270Recalls 4 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corp · BZE Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer

Cleared on Jan 12, 1993 after a 419-day review. FDA has not posted a summary or statement for it, so its predicates are not known. It has a Class I recall. It has been recalled 4 times, including 3 Class I recalls.

Decided
Jan 12, 1993
Review
419 days
Class
Class II
Product code
BZE
Recalls
4 (3 Class I)
Predicates
0

FDA has not posted a summary or statement for it, so its predicates are not known.

FIG. 1 — Lineage of K915226Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZE device

  1. #1 K252066 inspired™ Heated Breathing Circuit (510-008)…score 13.3 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 31, 2014Class IIZ-2689-2014No Marketing ApplicationThe CareFusion AirLife Heated Infant Breathing Circuit is being recalled due to a regulatory compliance risk involving a material change. The changes to the gas pathway may potentially impact the safety of the device and quality of the…
May 29, 2012Class IZ-1900-2012Equipment maintenanceThe Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.
May 29, 2012Class IZ-1901-2012Equipment maintenanceThe Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.
May 29, 2012Class IZ-1899-2012Equipment maintenanceThe Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.

No ancestor or descendant has a Class I recall.

How does its category compare?

All BZE devices

BZE Heater, Breathing System W/Wo Controller (Not Humidifier Or Nebulizer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BZE

1in 2025

2010: 22011: 12012: 22014: 12015: 12016: 52017: 12018: 12021: 12023: 12024: 22025: 12026 (partial year): 2
2007high 5 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZE

241in 2016

2016: 241 days median
2007high 241 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZE

No recall names BZE since 2007.

Adverse event reports per year · BZE

193in 2025

2007: 692008: 742009: 802010: 1262011: 1922012: 3342013: 3872014: 3382015: 1552016: 2142017: 2092018: 1032019: 1342020: 1232021: 1242022: 1522023: 1252024: 1722025: 193
2007high 387 (2013) · low 69 (2007)2025

MAUDE reports naming product code BZE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—069
20080—074
20090—080
20102—0126
20111—0192
20122—0334
20130—0387
20141—0338
20151—0155
201652410214
20171—0209
20181—0103
20190—0134
20200—0123
20211—0124
20220—0152
20231—0125
20242—0172
20251—0193
20262—0—
Cite this page

510(k) Report. “K915226: Isothermal Heated Ventilator & Anesthesia Circuit.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K915226

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026