510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Access Theophylline Immunoassay Reagents

K922823510(k) TraditionalCleared Aug 5, 199255-day reviewClass II21 CFR 862.1295Recalls 19

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bio-Rad Laboratories, Inc. · CGN Acid, Folic, Radioimmunoassay

Cleared on Aug 5, 1992 after a 55-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 16 later devices descend from it. It has been recalled 19 times.

Decided
Aug 5, 1992
Review
55 days
Class
Class II
Product code
CGN
Recalls
19
Predicates
0

16 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2003: 1 descendant cleared2013: 2 descendants cleared2019: 2 descendants cleared2009: 1 descendant cleared2018: 2 descendants cleared2025: 3 descendants cleared2022: 1 descendant cleared2011: 1 descendant cleared2023: 2 descendants cleared2014: 1 descendant cleared1990200020102020
  1. This deviceK922823 Access Theophylline Immunoassay ReagentsBio-Rad Laboratories, Inc. · 1992
  2. Descendant · cites it directly · 8 of its ownK121214 Access Accutni+3 Reagent and Access ACCUTN1+3 Calibrators for Use on…Beckman Coulter, Inc. · 2013 · Recalled
  3. Descendant · cites it directly · 6 of its ownK023764 Unicel Dxi 800 Access Immunoassay System, Model 973100Beckman Coulter, Inc. · 2003 · Recalled
  4. Descendant · 2 generations on · 4 of its ownK172787 Access hsTnlBeckman Coulter, Inc. · 2018 · Recalled
  5. Descendant · 2 generations on · 4 of its ownK090588 Power Processor Sample Processing System with Generic Connection…Beckman Coulter, Inc. · 2009
  6. Descendant · 3 generations on · 3 of its ownK110413 Power Processor Sample Processing System with Generic ConnectionsBeckman Coulter, Inc. · 2011 · Recalled

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2003: 12009: 12011: 12013: 22014: 12018: 22019: 22022: 12023: 22025: 3
2003high 3 (2025) · low 0 (2004)2026

Most influential descendants

Showing 5 of 16 · See all
  • K121214Access Accutni+3 Reagent and Access ACCUTN1+3 Calibrators for Use on the Access 2…Beckman Coulter, Inc. · 2013 · cites it directly8descendants
  • K023764Unicel Dxi 800 Access Immunoassay System, Model 973100Beckman Coulter, Inc. · 2003 · cites it directly6descendants
  • K172787Access hsTnlBeckman Coulter, Inc. · 2018 · 2 generations on4descendants
  • K090588Power Processor Sample Processing System with Generic Connection Module, Model: 4210Beckman Coulter, Inc. · 2009 · 2 generations on4descendants
  • K110413Power Processor Sample Processing System with Generic ConnectionsBeckman Coulter, Inc. · 2011 · 3 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K922823Access Theophylline Immunoassay ReagentsBio-Rad Laboratories, Inc.1992—
+1K121214Access Accutni+3 Reagent and Access ACCUTN1+3 Calibrators for Use on the…Beckman Coulter, Inc.20135
+1K023764Unicel Dxi 800 Access Immunoassay System, Model 973100Beckman Coulter, Inc.200331
+2K172787Access hsTnlBeckman Coulter, Inc.20181
+2K090588Power Processor Sample Processing System with Generic Connection Module…Beckman Coulter, Inc.2009—
+3K110413Power Processor Sample Processing System with Generic ConnectionsBeckman Coulter, Inc.20112

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200910
201110
201320
201410
201820
201920
202210
202320
202530
FIG. 1 — Lineage of K922823, 1992–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CGN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGN device

  1. #1 K233060 Elecsys Folate IIIscore 9.9 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (16)

See all 16

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K121214Access Accutni+3 Reagent and Access ACCUTN1+3 Calibrators for Use on the Access 2…Beckman Coulter, Inc.20138
K023764Unicel Dxi 800 Access Immunoassay System, Model 973100Beckman Coulter, Inc.20036

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 6, 2017Class IIIZ-0308-2018Device DesignPotential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.
Nov 6, 2017Class IIIZ-0309-2018Device DesignPotential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.
Mar 7, 2017Class IIZ-1709-2017Software Design ChangeBeckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position…
Mar 7, 2017Class IIZ-1710-2017Software Design ChangeBeckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position…
Mar 7, 2017Class IIZ-1711-2017Software Design ChangeBeckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position…
May 17, 2016Class IIZ-1956-2016Under Investigation by firmDue to the lack of updated thermal components, these instruments may be more susceptible to the internal case temperature exceeding the 18 to 36 degrees Celsius operating range, causing SYS flags. The instrument will not report results…
Nov 26, 2013Class IIZ-0984-2014Nonconforming Material/ComponentThe Access 2 Immunoassay System Analyzer and the UniCel DxI 600 Access may have been manufactured with non-conforming bushings. Non-conforming bushings have a remote probability of generating incorrect results.
Sep 12, 2013Class IIZ-0291-2014Mixed-up of materials/componentsBeckman Coulter is recalling the Access 2 Immunoassay System because they determined that the fuses in the system may be of the wrong amperage.
Feb 12, 2013Class IIZ-1032-2013Device DesignBeckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley…
Nov 1, 2012Class IIZ-0612-2013Device DesignIt has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at…
Apr 15, 2011Not classifiedZ-3126-2011OtherThe recall was initiated because certain Access immunoassays can produce inaccurate results when performed within the published UniCel DxI Immunoassay Systems* room temperature operating specification of 18 degrees C to 32 degrees C (64.4…
Aug 13, 2010Not classifiedZ-0607-2011PendingThis recall was initiated because Beckman Coulter has received reports of erroneous test results being reported as a result of assay reagent packs being transferred between systems. Erroneous results will be generated as a result of…
Sep 11, 2008Not classifiedZ-1372-2012Software designThe recall was initiated because Beckman confirmed a customer report that a timing conflict can occur while aspirating samples for the Access Total BhCG assay.
Apr 3, 2008Not classifiedZ-2384-2008Nonconforming Material/ComponentThe field action was initiated after the firm confirmed reports of premature failure of the waste pump tubing in the Access and Access 2 Immunoassay Systems. Premature failure of the waste pump tubing may negatively affect precision.
Oct 17, 2007Not classifiedZ-1183-2008Process controlElectrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering.
Nov 8, 2004Not classifiedZ-0265-05OtherA rare condition of the Access 2 Immunoassay Systems software versions 2.0, 2.1 and 2.2 could present a risk of an erroneous confirmatory result under a specific set of circumstances. The Chlamydia Blocking results may be affected. The…
Jun 18, 2003Not classifiedZ-1148-03OtherDefective software media distributed with upgrade software may cause installation failure and systems lock.
Jan 30, 2003Not classifiedZ-0692-03OtherFalse results from secondary pour off tubes
Jan 27, 2003Not classifiedZ-0619-03OtherSpring failure, design defect in component.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K121214 (Beckman Coulter, Inc., 2013) has gone furthest, with 8 descendants of its own.

How does its category compare?

All CGN devices

CGN Acid, Folic, Radioimmunoassay: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CGN

1in 2024

2009: 22010: 12011: 12014: 12016: 12018: 12020: 12023: 12024: 1
2007high 2 (2009) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGN

Too few CGN 510(k)s in any year since 2007 to give a median.

Recalls per year · CGN

3in 2019

2007: 22008: 12009: 12010: 12012: 22013: 12018: 22019: 3
2007high 3 (2019) · low 0 (2011)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CGN

9in 2025

2007: 262009: 32011: 282012: 112013: 22014: 42015: 12016: 42017: 62018: 22019: 12020: 52021: 32022: 132023: 82024: 192025: 9
2007high 28 (2011) · low 0 (2008)2025

MAUDE reports naming product code CGN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—226
20080—10
20092—13
20101—10
20111—028
20120—211
20130—12
20141—04
20150—01
20161—04
20170—06
20181—22
20190—31
20201—05
20210—03
20220—013
20231—08
20241—019
20250—09
20260—0—
Cite this page

510(k) Report. “K922823: Access Theophylline Immunoassay Reagents.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K922823

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026