510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Cefpodoxime 10 MCG, Sensi Disc

K924797510(k) TraditionalCleared Nov 2, 199239-day reviewClass II21 CFR 866.1620Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bd Becton Dickinson Vacutainer Systems Preanalytic · JTN Susceptibility Test Discs, Antimicrobial

Cleared on Nov 2, 1992 after a 39-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. It has been recalled once.

Decided
Nov 2, 1992
Review
39 days
Class
Class II
Product code
JTN
Recalls
1
Predicates
0

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

FIG. 1 — Lineage of K924797Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JTN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JTN device

  1. #1 K230651 BD BBL™ Sensi-Disc™ Lefamulin 20µg (LMU-20)score 5.5 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 8, 2024Class IIZ-1021-2024Process controlThere is a possibility of reproducibility, accuracy, and/or QC failures in antibiotic susceptibility testing (AST) for H. influenzae. This may cause product discard, delayed results, or additional adverse diagnostic outcomes, such as a…

No ancestor or descendant has a Class I recall.

How does its category compare?

All JTN devices

JTN Susceptibility Test Discs, Antimicrobial: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JTN

7in 2025

2007: 12008: 42010: 32011: 12015: 32017: 32018: 32019: 32020: 22021: 22022: 22023: 42024: 22025: 72026 (partial year): 4
2007high 7 (2025) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JTN

87in 2025

2008: 44 days median2010: 15 days median2015: 56 days median2017: 33 days median2018: 27 days median2019: 23 days median2023: 67 days median2025: 87 days median2026 (partial year): 56 days median
2007high 87 (2025) · low 15 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JTN

31in 2024

2007: 12012: 32013: 52015: 12024: 312026 (partial year): 1
2007high 31 (2024) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JTN

7in 2025

2007: 22021: 52022: 52023: 42024: 52025: 7
2007high 7 (2025) · low 0 (2008)2025

MAUDE reports naming product code JTN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—12
200844400
20090—00
201031500
20111—00
20120—30
20130—50
20140—00
201535610
20160—00
201733300
201832700
201932300
20202—00
20212—05
20222—05
202346704
20242—315
202578707
20264561—
Cite this page

510(k) Report. “K924797: Cefpodoxime 10 MCG, Sensi Disc.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K924797

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026