Accu-Chek Advantage Blood Glucose System

K930979510(k) TraditionalCleared Sep 8, 1993195-day reviewClass II21 CFR 862.1345Recalls 10

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boehringer Mannheim Corp. · CGA Glucose Oxidase, Glucose

Cleared on Sep 8, 1993 after a 195-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 3 of the 36 devices in its family, counting itself, have a Class I recall. 35 later devices descend from it. It has been recalled 10 times.

Decided
Sep 8, 1993
Review
195 days
Class
Class II
Product code
CGA
Recalls
10
Predicates
0

35 later devices descend from it, 2 citing it directly. 3 of the 36 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1997: 1 descendant cleared2000: 1 descendant cleared2002: 4 descendants cleared2001: 2 descendants cleared2006: 2 descendants cleared2003: 1 descendant cleared2010: 3 descendants cleared, 2 with a Class I recall22005: 2 descendants cleared2007: 3 descendants cleared2008: 2 descendants cleared2012: 4 descendants cleared, 1 with a Class I recall12017: 1 descendant cleared2009: 1 descendant cleared2013: 3 descendants cleared2015: 3 descendants cleared2011: 1 descendant cleared2016: 1 descendant cleared1990200020102020
  1. This deviceK930979 Accu-Chek Advantage Blood Glucose SystemBoehringer Mannheim Corp. · 1993
  2. Descendant · cites it directly · 33 of its ownK992684 Freestyle Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2000 · Recalled
  3. Descendant · 2 generations on · 23 of its ownK012014 Modification to Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2001
  4. Descendant · 2 generations on · 22 of its ownK020866 Freestyle Tracker Diabetes Management SystemAbbott Diabetes Care, Inc. · 2002
  5. Descendant · 2 generations on · 18 of its ownK031260 Freestyle Blood Glucose Monitoring System, Freestyle Tracker…Abbott Diabetes Care, Inc. · 2003
  6. Descendant · 2 generations on · 12 of its ownK050500 Freestyle 600 Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2005

Showing 5 of 35 descendants. See all descendants

Descendants cleared per year

1997: 12000: 12001: 22002: 42003: 12005: 22006: 22007: 32008: 22009: 12010: 32011: 12012: 42013: 32015: 32016: 12017: 1
1997high 4 (2002) · low 0 (1998)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 35 · See all
  • K992684Freestyle Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2000 · cites it directly33descendants
  • K012014Modification to Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2001 · 2 generations on23descendants
  • K020866Freestyle Tracker Diabetes Management SystemAbbott Diabetes Care, Inc. · 2002 · 2 generations on22descendants
  • K031260Freestyle Blood Glucose Monitoring System, Freestyle Tracker Diabetes Management System…Abbott Diabetes Care, Inc. · 2003 · 2 generations on18descendants
  • K050500Freestyle 600 Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2005 · 2 generations on12descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K930979Accu-Chek Advantage Blood Glucose SystemBoehringer Mannheim Corp.1993—
+1K992684Freestyle Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.20002
+2K012014Modification to Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.2001—
+2K020866Freestyle Tracker Diabetes Management SystemAbbott Diabetes Care, Inc.2002—
+2K031260Freestyle Blood Glucose Monitoring System, Freestyle Tracker Diabetes…Abbott Diabetes Care, Inc.2003—
+2K050500Freestyle 600 Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.2005—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199710
200010
200120
200240
200310
200520
200620
200730
200820
200910
201032
201110
201241
201330
201530
201610
201710
FIG. 1 — Lineage of K930979, 1993–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CGA device

  1. #1 K252488 ABL90 FLEX PLUS Systemscore 12.8 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (35)

See all 35

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K992684Freestyle Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc.200033
K973662Accu-Chek (R)Complete System/Meter & Accu-Chek Advantage Test Strips/Accu-Chek Advantage…Boehringer Mannheim Corp.19970

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 22, 2006Not classifiedZ-1357-06OtherThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that the test strip may be damaged or test was not performed…
Jun 22, 2006Not classifiedZ-1358-06OtherThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error.
Jun 22, 2006Not classifiedZ-1369-06OtherThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Jun 22, 2006Not classifiedZ-1362-06OtherThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Jun 22, 2006Not classifiedZ-1354-06OtherThe meter gives a strip error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error and the manual only instructs the…
Jun 22, 2006Not classifiedZ-1360-06OtherThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.
Jun 22, 2006Not classifiedZ-1359-06OtherThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error.
Jun 22, 2006Not classifiedZ-1355-06OtherThe meter gives a strip error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error and the manual only instructs the…
Jun 22, 2006Not classifiedZ-1356-06OtherThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a strip error.
Jun 22, 2006Not classifiedZ-1361-06OtherThe meter gives an error message that can "actually" mean either a problem with the strip or a blood glucose too low to measure, but the meter error message only reports that there is a bad strip.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K092638Freestyle Blood Glucose Test Strips, Models 100-Count: 12401-01, 12101-21, 12101-22, 12101-23Abbott Diabetes Care, Inc. · 2010 · 2 generations on · ◆ in the 2010 descendants bar of FIG. 12 Class I recalls
  • K092602FREESTYLE Lite Blood Glucose Monitoring System and Freestyle Freedom Lite Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2010 · 4 generations on · ◆ in the 2010 descendants bar of FIG. 12 Class I recalls
  • K120568Freestyle Insulinx Blood Glucose Monitoring SystemAbbott Diabetes Care, Inc. · 2012 · 4 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K992684 (Abbott Diabetes Care, Inc., 2000) has gone furthest, with 33 descendants of its own.

How does its category compare?

All CGA devices

CGA Glucose Oxidase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CGA

2in 2024

2007: 72008: 62009: 102010: 82011: 92012: 102013: 32014: 52015: 42016: 42017: 22018: 22019: 22020: 22021: 12022: 12023: 42024: 22026 (partial year): 1
2007high 10 (2009) · low 0 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CGA

249in 2023

2007: 122 days median2008: 56 days median2009: 155 days median2010: 122 days median2011: 217 days median2012: 123 days median2013: 441 days median2014: 265 days median2015: 150 days median2016: 157 days median2023: 249 days median
2007high 441 (2013) · low 56 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CGA

3in 2023

2007: 22008: 22009: 42010: 62011: 32013: 32014: 22015: 6, 2 Class I2016: 22019: 12023: 3
2007high 6 (2010) · low 0 (2012)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · CGA

59in 2025

2007: 292008: 372009: 752010: 1722011: 4232012: 2332013: 1722014: 1222015: 1552016: 362017: 922018: 692019: 1162020: 682021: 482022: 2722023: 1,1412024: 2662025: 59
2007high 1,141 (2023) · low 29 (2007)2025

MAUDE reports naming product code CGA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20077122229
2008656237
200910155475
201081226172
201192173423
2012101230233
201334413172
201452652122
201541506 (2)155
20164157236
20172—092
20182—069
20192—1116
20202—068
20211—048
20221—0272
2023424931,141
20242—0266
20250—059
20261—0—
Cite this page

510(k) Report. “K930979: Accu-Chek Advantage Blood Glucose System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K930979

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026