510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Mavig Angio Shield Models 6290 & 6272

K934206510(k) TraditionalCleared Oct 7, 199338-day reviewClass I21 CFR 892.6500Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

F. Walter Hanel GmbH · KPY Shield, Protective, Personnel

Cleared on Oct 7, 1993 after a 38-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. It has been recalled once.

Decided
Oct 7, 1993
Review
38 days
Class
Class I
Product code
KPY
Recalls
1
Predicates
0

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

FIG. 1 — Lineage of K934206Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 28, 2018Class IIZ-2962-2018Equipment maintenanceThe torque wrench used on the ceiling column assembly during a recent service visit was outside of the torque specification, potentially overtightening the bolts. The overtightening of bolts may compromise the structural integrity of the…

No ancestor or descendant has a Class I recall.

How does its category compare?

All KPY devices

KPY Shield, Protective, Personnel: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPY

1in 2011

2011: 1
2007high 1 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPY

Too few KPY 510(k)s in any year since 2007 to give a median.

Recalls per year · KPY

No recall names KPY since 2007.

Adverse event reports per year · KPY

1in 2023

2007: 42008: 32009: 32010: 42011: 12012: 22015: 12016: 12017: 12020: 22022: 12023: 1
2007high 4 (2007) · low 0 (2013)2025

MAUDE reports naming product code KPY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—04
20080—03
20090—03
20100—04
20111—01
20120—02
20130—00
20140—00
20150—01
20160—01
20170—01
20180—00
20190—00
20200—02
20210—00
20220—01
20230—01
20240—00
20250—00
20260—0—
Cite this page

510(k) Report. “K934206: Mavig Angio Shield Models 6290 & 6272.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K934206

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026