# K935013: DSL Estrone (DSL 8700)

- Pathway: 510(k) (Traditional)
- Applicant: Diagnostic Systems Laboratories, Inc.
- Decided: Jun 16, 1994 after 239 days
- Product code: CGF (Radioimmunoassay, Estrone)
- Device class: 1
- Regulation: 21 CFR 862.1280
- Recalls of this device: 4 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

## Recalls of this device
- Dec 15, 2025 · Class II · Z-1369-2026: The affected Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability. Only in some (very rare) cases may the issue occur.
- Jun 12, 2025 · Class II · Z-2336-2025: Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability or rarely give falsely low patient results which might cause Permanent injury
- Jun 12, 2025 · Class II · Z-2152-2025: A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient
- Apr 22, 2010 · Not classified · Z-1685-2011: Beckman Coulter has confirmed that eight Estrone RIA kit (Ref DSL 8700) lots are inaccurately calibrated to the firm's internal reference calibrator. Lots 992637, 100115A, 100205A, 100205RA, and 10022

Source page: https://510kreport.com/device/K935013