510(k) · Cleared
Recalled · 2007Ecat PRT
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Cti Pet Systems, Inc. · KPS System, Tomography, Computed, Emission
Cleared on Mar 22, 1994 after a 48-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. It has been recalled 14 times.
- Decided
- Mar 22, 1994
- Review
- 48 days
- Class
- Class II
- Product code
- KPS
- Recalls
- 14
- Predicates
- 0
FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KPS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KPS device
- #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 23, 2007 | Not classified | Z-0271-2008 | Software design | Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions. |
| Jul 23, 2007 | Not classified | Z-0273-2008 | Software design | Incorrect Data Results - ECAT System's software versions 7.2.2 and 7.4 may calculate the decay correction for emission data incorrectly while performing whole-body studies with all transmissions followed by all emissions. |
| Jul 23, 2007 | Not classified | Z-0251-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0252-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0270-2008 | Software design | Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions. |
| Jul 23, 2007 | Not classified | Z-0246-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0247-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0249-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0248-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0250-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0269-2008 | Software design | Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions. |
| Jul 23, 2007 | Not classified | Z-0272-2008 | Software design | Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions. |
| Mar 28, 2006 | Not classified | Z-0712-06 | Other | There is a potential for inaccurate, elevated SUVs (standardized uptake values). On follow-up scans, the elevated SUV could be misinterpreted, and inappropriately impact a patient's treatment plan. |
| Mar 4, 2003 | Not classified | Z-0674-03 | Other | Incorrect patient information is displayed in the reconstructed images of a whole body patient scan. |
The 5 latest of 14 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
How does its category compare?
All KPS devicesKPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · KPS
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KPS
65in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KPS
16in 2025
Class I (17 since 2007) Class II/III not classified
Adverse event reports per year · KPS
36in 2025
MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 8 | 38 | 22 | 38 |
| 2008 | 17 | 25 | 16 | 11 |
| 2009 | 9 | 42 | 14 | 15 |
| 2010 | 3 | 91 | 4 | 44 |
| 2011 | 7 | 28 | 34 | 56 |
| 2012 | 6 | 94 | 13 | 34 |
| 2013 | 11 | 84 | 13 | 240 |
| 2014 | 11 | 92 | 18 | 33 |
| 2015 | 3 | 112 | 5 | 24 |
| 2016 | 11 | 65 | 16 | 10 |
| 2017 | 11 | 104 | 11 | 31 |
| 2018 | 13 | 59 | 12 (3) | 26 |
| 2019 | 3 | 60 | 5 (5) | 105 |
| 2020 | 12 | 72 | 8 | 22 |
| 2021 | 15 | 66 | 4 | 10 |
| 2022 | 5 | 60 | 8 (6) | 15 |
| 2023 | 12 | 63 | 6 (3) | 22 |
| 2024 | 11 | 59 | 15 | 23 |
| 2025 | 12 | 65 | 16 | 36 |
| 2026 | 6 | 67 | 6 | — |
https://510kreport.com/device/K940478 · printed Nov 6, 2026
Cite this page
510(k) Report. “K940478: Ecat PRT.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K940478
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Page last updated Nov 3, 2026