# K940478: Ecat PRT

- Pathway: 510(k) (Traditional)
- Applicant: Cti Pet Systems, Inc.
- Decided: Mar 22, 1994 after 48 days
- Product code: KPS (System, Tomography, Computed, Emission)
- Device class: 2
- Regulation: 21 CFR 892.1200
- Recalls of this device: 14 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

## Recalls of this device
- Jul 23, 2007 · Not classified · Z-0271-2008: Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissi
- Jul 23, 2007 · Not classified · Z-0273-2008: Incorrect Data Results - ECAT System's software versions 7.2.2 and 7.4 may calculate the decay correction for emission data incorrectly while performing whole-body studies with all transmissions follo
- Jul 23, 2007 · Not classified · Z-0251-2008: Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstr
- Jul 23, 2007 · Not classified · Z-0252-2008: Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstr
- Jul 23, 2007 · Not classified · Z-0270-2008: Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissi
- Jul 23, 2007 · Not classified · Z-0246-2008: Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstr
- Jul 23, 2007 · Not classified · Z-0247-2008: Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstr
- Jul 23, 2007 · Not classified · Z-0249-2008: Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstr
- Jul 23, 2007 · Not classified · Z-0248-2008: Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstr
- Jul 23, 2007 · Not classified · Z-0250-2008: Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstr
- Jul 23, 2007 · Not classified · Z-0269-2008: Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissi
- Jul 23, 2007 · Not classified · Z-0272-2008: Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissi
- Mar 28, 2006 · Not classified · Z-0712-06: There is a potential for inaccurate, elevated SUVs (standardized uptake values). On follow-up scans, the elevated SUV could be misinterpreted, and inappropriately impact a patient's treatment plan.
- Mar 4, 2003 · Not classified · Z-0674-03: Incorrect patient information is displayed in the reconstructed images of a whole body patient scan.

Source page: https://510kreport.com/device/K940478