# K943780: Negative Control Strips for Urinalysis (Chex-Stix, Clinitek Auto 2000, Clinitex Atlas)

- Pathway: 510(k) (Traditional)
- Applicant: Heraeus Kulzer, Inc.
- Decided: Oct 31, 1994 after 89 days
- Product code: JJW (Urinalysis Controls (Assayed And Unassayed))
- Device class: 1
- Regulation: 21 CFR 862.1660
- Recalls of this device: 7 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA has not posted a summary or statement for it, so its predicates are not known.

## Recalls of this device
- Nov 21, 2012 · Class III · Z-0637-2013: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).
- Nov 21, 2012 · Class III · Z-0639-2013: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).
- Nov 21, 2012 · Class II · Z-0640-2013: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).
- Nov 21, 2012 · Class III · Z-0641-2013: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).
- Nov 21, 2012 · Class III · Z-0635-2013: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).
- Nov 21, 2012 · Class III · Z-0636-2013: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).
- Nov 21, 2012 · Class III · Z-0638-2013: When these lots of control material are reconstituted, the positive control solution leukocyte result may become negative if the solution is held for more than two hours above 25C (77F).