510(k) Report

510(k) · Cleared

Recalled · 2008

Simview (TM) 3000

K946053510(k) TraditionalCleared Mar 31, 1995109-day reviewClass II21 CFR 892.5840Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Solutions USA, Inc. · KPQ System, Simulation, Radiation Therapy

Cleared on Mar 31, 1995 after a 109-day review. FDA has not posted a summary or statement for it, so its predicates are not known. It has been recalled 5 times.

Decided
Mar 31, 1995
Review
109 days
Class
Class II
Product code
KPQ
Recalls
5
Predicates
0

FDA has not posted a summary or statement for it, so its predicates are not known.

FIG. 1 — Lineage of K946053Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPQ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 6, 2008Not classifiedZ-0646-2009Device DesignExcessive wear on drive belts used for Image Intensifier and/or X-ray head drive mechanisms may result in uncontrolled movement, which may cause injury.
Mar 10, 2008Not classifiedZ-2196-2008Software designThe firm sent an Update Instruction, "TH003/08/S" for Motion Enable Switch/holder assembly. The switch holder allows the motion enable switch to be activated if inserted partially; thereby enabling inadvertent motion and serious injury…
Feb 27, 2007Not classifiedZ-0281-2008Labeling designLoose Mounting Screws -- Cassette holder may become loose and result in the device dislodging and colliding with the patient.
Feb 27, 2007Not classifiedZ-0283-2008Labeling designLoose Mounting Screws -- Cassette holder may become loose and result in the device dislodging and colliding with the patient.
Feb 27, 2007Not classifiedZ-0284-2008Labeling designLoose Mounting Screws -- Cassette holder may become loose and result in the device dislodging and colliding with the patient.

No ancestor or descendant has a Class I recall.

How does its category compare?

All KPQ devices

KPQ System, Simulation, Radiation Therapy: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPQ

1in 2014

2007: 12008: 12009: 12010: 12013: 12014: 1
2007high 1 (2007) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPQ

Too few KPQ 510(k)s in any year since 2007 to give a median.

Recalls per year · KPQ

1in 2014

2007: 62008: 32009: 12012: 12013: 12014: 1
2007high 6 (2007) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPQ

1in 2020

2007: 172008: 122009: 72010: 102011: 122012: 112013: 12014: 32019: 12020: 1
2007high 17 (2007) · low 0 (2015)2025

MAUDE reports naming product code KPQ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—617
20081—312
20091—17
20101—010
20110—012
20120—111
20131—11
20141—13
20150—00
20160—00
20170—00
20180—00
20190—01
20200—01
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—
Cite this page

510(k) Report. “K946053: Simview (TM) 3000.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K946053

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Page last updated Nov 3, 2026