510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Synthes 2.7 MM Cannulated Screw and Threaded Washer

K951304510(k) TraditionalCleared Jun 16, 199586-day reviewClass II21 CFR 888.3040Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Synthes (Usa) · HWC Screw, Fixation, Bone

Cleared on Jun 16, 1995 after an 86-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 later devices descend from it. It has been recalled once.

Decided
Jun 16, 1995
Review
86 days
Class
Class II
Product code
HWC
Recalls
1
Predicates
0

2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2019: 1 descendant cleared2026: 1 descendant cleared1995200020052010201520202025
  1. This deviceK951304 Synthes 2.7 MM Cannulated Screw and Threaded WasherSynthes (Usa) · 1995
  2. Descendant · cites it directly · 1 of its ownK191118 Avanti Distal Radius and Forearm SystemAvanti Orthopaedics, Inc. · 2019
  3. Descendant · 2 generations on · 0 of its ownK261145 Avanti Distal Elbow ORIF SystemAvanti Orthopaedics, LLC · 2026

Most influential descendants

Showing 2 of 2 · See all
  • K191118Avanti Distal Radius and Forearm SystemAvanti Orthopaedics, Inc. · 2019 · cites it directly1descendant
  • K261145Avanti Distal Elbow ORIF SystemAvanti Orthopaedics, LLC · 2026 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K951304Synthes 2.7 MM Cannulated Screw and Threaded WasherSynthes (Usa)1995—
+1K191118Avanti Distal Radius and Forearm SystemAvanti Orthopaedics, Inc.2019—
+2K261145Avanti Distal Elbow ORIF SystemAvanti Orthopaedics, LLC2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
202610
FIG. 1 — Lineage of K951304, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HWC device

  1. #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K191118Avanti Distal Radius and Forearm SystemAvanti Orthopaedics, Inc.20191

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 30, 2014Class IIZ-2265-2014Employee errorA screw set was returned from the field for destruction after being exposed to flood water, which was erroneously reprocessed and returned to the field. Initial investigative testing indicated the components within the set may be cytotoxic.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K191118 (Avanti Orthopaedics, Inc., 2019) has gone furthest, with 1 descendant of its own.

How does its category compare?

All HWC devices

HWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · HWC

15in 2025

2007: 352008: 332009: 272010: 232011: 352012: 202013: 252014: 312015: 412016: 262017: 402018: 202019: 302020: 282021: 262022: 262023: 242024: 172025: 152026 (partial year): 21
2007high 41 (2015) · low 15 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HWC

111in 2025

2007: 77 days median2008: 76 days median2009: 113 days median2010: 110 days median2011: 88 days median2012: 99 days median2013: 90 days median2014: 88 days median2015: 85 days median2016: 131 days median2017: 127 days median2018: 106 days median2019: 108 days median2020: 75 days median2021: 84 days median2022: 167 days median2023: 104 days median2024: 90 days median2025: 111 days median2026 (partial year): 125 days median
2007high 167 (2022) · low 75 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HWC

1in 2025

2007: 252008: 332009: 292010: 72011: 122012: 302013: 142014: 22015: 52016: 42017: 42018: 12019: 52020: 42021: 22022: 132025: 12026 (partial year): 2
2007high 33 (2008) · low 0 (2023)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HWC

2,893in 2025

2007: 6692008: 6602009: 6512010: 7652011: 1,1042012: 3,3392013: 3,8502014: 2,5492015: 2,7112016: 2,2852017: 2,5612018: 3,6622019: 5,4632020: 4,8202021: 4,4462022: 3,9342023: 3,7182024: 3,5752025: 2,893
2007high 5,463 (2019) · low 651 (2009)2025

MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007357725669
2008337633660
20092711329651
2010231107765
20113588121,104
20122099303,339
20132590143,850
2014318822,549
2015418552,711
20162613142,285
20174012742,561
20182010613,662
20193010855,463
2020287544,820
2021268424,446
202226167133,934
20232410403,718
2024179003,575
20251511112,893
2026211252—
Cite this page

510(k) Report. “K951304: Synthes 2.7 MM Cannulated Screw and Threaded Washer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K951304

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026