# K952875: Accu-Chek Alpha Monitor & Accu-Chek Glucose Test Strips

- Pathway: 510(k) (Traditional)
- Applicant: Boehringer Mannheim Corp.
- Decided: Nov 21, 1995 after 155 days
- Product code: CGA (Glucose Oxidase, Glucose)
- Device class: 2
- Regulation: 21 CFR 862.1345
- Recalls of this device: 2 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
No predicates in its family tree.
No later device cites it as a predicate yet.
0 of 1 device in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- None. FDA has not posted a summary or statement for it, so its predicates are not known.

## Recalls of this device
- May 18, 2005 · Not classified · Z-0978-05: The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards while the meter is waiting for the sample o
- May 18, 2005 · Not classified · Z-0979-05: The meter may self-start without a sample having been applied and report erroneous results without an error message if an undosed test strip is bent upwards while the meter is waiting for the sample o