510(k) · Cleared
ClearedConsensus All Poly Tibia
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
U.S. Medical Products, Inc. · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Cleared on Apr 1, 1996 as substantially equivalent to K922048 (Howmedica Corp., 1992) and 2 other predicates after a 280-day review. Its family tree reaches back 1 generation to 1991.
- Decided
- Apr 1, 1996
- Review
- 280 days
- Class
- Class II
- Product code
- JWH
- Recalls
- None
- Predicates
- 3
Its longest line of descent runs back 1 generation to K910563 (1991). 3 ancestors in all. No later device cites it as a predicate yet.
- Root · 1 generation backK910563 P.F.C.(R) Modular Knee Syst, Stab All-Plast TibialJohnson & Johnson Orthopaedics, Inc. · 1991
- This deviceK952943 Consensus All Poly TibiaU.S. Medical Products, Inc. · 1996
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K910563 | P.F.C.(R) Modular Knee Syst, Stab All-Plast Tibial | Johnson & Johnson Orthopaedics, Inc. | 1991 | — |
| 0 (this device) | K952943 | Consensus All Poly Tibia | U.S. Medical Products, Inc. | 1996 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JWH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JWH device
- #1 K221479 Persona Personalized Knee Systemscore 22.4 · 11 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K922048 | Duracon All Plastic Tibial ComponentSummary page 1…lymer, under classification 21CFR888.3560. Substantial Equivalence: equivalent All poly tibia components are as follows: Howmedica Duracon All Plastic Tibial component K922048 SE Sep 8, 1992 Intermedics Orthopedics All-Poly Tibia K923443 SE 3-31-93 Johnson&Johnson PFC All Plastic Tibia K910563 SE 5-23-91… | 1992 |
| * | K923443 | Natural-Knee All-Poly TibiaSummary page 1…lence: equivalent All poly tibia components are as follows: Howmedica Duracon All Plastic Tibial component K922048 SE Sep 8, 1992 Intermedics Orthopedics All-Poly Tibia K923443 SE 3-31-93 Johnson&Johnson PFC All Plastic Tibia K910563 SE 5-23-91 Device Description: The Consensus® Knee All Poly Tibia compo… | 1993 |
| * | K910563 | P.F.C.(R) Modular Knee Syst, Stab All-Plast TibialSummary page 1…s: Howmedica Duracon All Plastic Tibial component K922048 SE Sep 8, 1992 Intermedics Orthopedics All-Poly Tibia K923443 SE 3-31-93 Johnson&Johnson PFC All Plastic Tibia K910563 SE 5-23-91 Device Description: The Consensus® Knee All Poly Tibia component is manufactured from ultra-high molecular weight pol… | 1991 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- None of the 3 ancestors has a recall.
How does its category compare?
All JWH devicesJWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JWH
20in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JWH
55in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JWH
2in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JWH
9,306in 2025
MAUDE reports naming product code JWH, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 18 | 74 | 14 | 1,305 |
| 2008 | 17 | 81 | 10 | 1,483 |
| 2009 | 11 | 168 | 109 | 1,870 |
| 2010 | 19 | 85 | 32 | 3,286 |
| 2011 | 25 | 87 | 27 | 5,133 |
| 2012 | 33 | 85 | 85 | 5,203 |
| 2013 | 31 | 128 | 31 | 6,135 |
| 2014 | 33 | 98 | 13 | 6,722 |
| 2015 | 27 | 87 | 11 | 6,881 |
| 2016 | 18 | 88 | 17 | 7,586 |
| 2017 | 25 | 118 | 61 | 12,263 |
| 2018 | 25 | 90 | 14 | 14,071 |
| 2019 | 28 | 133 | 5 | 14,987 |
| 2020 | 22 | 59 | 2 | 12,592 |
| 2021 | 24 | 61 | 8 | 9,408 |
| 2022 | 17 | 56 | 12 | 8,364 |
| 2023 | 20 | 57 | 67 | 8,947 |
| 2024 | 19 | 89 | 7 | 10,828 |
| 2025 | 20 | 55 | 2 | 9,306 |
| 2026 | 12 | 63 | 4 | — |
What is it indicated for?
The Consensus® All Poly Tibia is indicated for use in:
1. Primary intervention of rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, or degenerative arthritis.
2. Failed osteotomy or unicompartmental replacements
3. Replacement of unsatisfactory cemented or press-fit knee components when sufficient bone stock exists.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K952943 · printed Nov 4, 2026
Cite this page
510(k) Report. “K952943: Consensus All Poly Tibia.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K952943
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026