510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Ektachem Clinical Chemistry Slide C-Reactive Protein (CRP)

K953197510(k) TraditionalCleared Jul 31, 199521-day reviewClass II21 CFR 866.5270Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Johnson & Johnson Clinical Diagnostics, Inc. · DCK C-Reactive Protein, Antigen, Antiserum, And Control

Cleared on Jul 31, 1995 after a 21-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. 5 later devices descend from it. It has been recalled 3 times.

Decided
Jul 31, 1995
Review
21 days
Class
Class II
Product code
DCK
Recalls
3
Predicates
0

5 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2003: 1 descendant cleared2004: 2 descendants cleared2015: 2 descendants cleared1995200020052010201520202025
  1. This deviceK953197 Ektachem Clinical Chemistry Slide C-Reactive Protein (CRP)Johnson & Johnson Clinical Diagnostics, Inc. · 1995
  2. Descendant · cites it directly · 2 of its ownK041799 Vitros Chemistry Products Hscrp Reagent, Vitros Chemistry Products…Ortho-Clinical Diagnostics, Inc. · 2004 · Recalled
  3. Descendant · cites it directly · 1 of its ownK030626 Vitros Chemistry Products CRP Slide and Calibrator Kit 7Ortho-Clinical Diagnostics, Inc. · 2003 · Recalled
  4. Descendant · 2 generations on · 0 of its ownK152433 VITROS Chemistry Products CRP Slides; VITRO Chemistry Products…Ortho Clinical Diagnostics · 2015 · Recalled
  5. Descendant · 2 generations on · 0 of its ownK141689 FastPack High Sensitivity C-Reactive Protein Immunoassay, High…Qualigen, Inc. · 2015
  6. Descendant · 2 generations on · 0 of its ownK042838 Vitros Chemistry Products Calibrator Kit 23; Vitros Chemistry…Ortho-Clinical Diagnostics, Inc. · 2004

Most influential descendants

Showing 5 of 5 · See all
  • K041799Vitros Chemistry Products Hscrp Reagent, Vitros Chemistry Products Calibrator Kit 17…Ortho-Clinical Diagnostics, Inc. · 2004 · cites it directly2descendants
  • K030626Vitros Chemistry Products CRP Slide and Calibrator Kit 7Ortho-Clinical Diagnostics, Inc. · 2003 · cites it directly1descendant
  • K152433VITROS Chemistry Products CRP Slides; VITRO Chemistry Products Calibrator Kit 7Ortho Clinical Diagnostics · 2015 · 2 generations on0descendants
  • K141689FastPack High Sensitivity C-Reactive Protein Immunoassay, High Sensitivity C-Reactive…Qualigen, Inc. · 2015 · 2 generations on0descendants
  • K042838Vitros Chemistry Products Calibrator Kit 23; Vitros Chemistry Products Palb Performance…Ortho-Clinical Diagnostics, Inc. · 2004 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K953197Ektachem Clinical Chemistry Slide C-Reactive Protein (CRP)Johnson & Johnson Clinical Diagnostics, Inc.1995—
+1K041799Vitros Chemistry Products Hscrp Reagent, Vitros Chemistry Products Calibrator…Ortho-Clinical Diagnostics, Inc.20043
+1K030626Vitros Chemistry Products CRP Slide and Calibrator Kit 7Ortho-Clinical Diagnostics, Inc.20033
+2K152433VITROS Chemistry Products CRP Slides; VITRO Chemistry Products Calibrator Kit 7Ortho Clinical Diagnostics20153
+2K141689FastPack High Sensitivity C-Reactive Protein Immunoassay, High Sensitivity…Qualigen, Inc.2015—
+2K042838Vitros Chemistry Products Calibrator Kit 23; Vitros Chemistry Products Palb…Ortho-Clinical Diagnostics, Inc.2004—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200420
201520
FIG. 1 — Lineage of K953197, 1995–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DCK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DCK device

  1. #1 K242170 K-ASSAY CRP (Ver.2)score 9.2 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

What descends from it? (5)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041799Vitros Chemistry Products Hscrp Reagent, Vitros Chemistry Products Calibrator Kit 17…Ortho-Clinical Diagnostics, Inc.20042
K030626Vitros Chemistry Products CRP Slide and Calibrator Kit 7Ortho-Clinical Diagnostics, Inc.20031

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 14, 2025Class IIZ-1876-2025Process controlProduct was incorrectly assigned GEN 88 instead of GEN 83 on product carton, foil wrapper, and cartridge (cart) label.
Oct 11, 2024Class IIZ-0535-2025Under Investigation by firmMay Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample…
Oct 26, 2016Class IIIZ-0864-2017Process controlCRP samples diluted using VITROS Chemistry Products Specialty Diluent, VITROS Chemistry Products FS Diluent Pack 3, or a sample with a low concentration of CRP had positively biased results if the endogenous level of CRP in the diluent is…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K041799 (Ortho-Clinical Diagnostics, Inc., 2004) has gone furthest, with 2 descendants of its own.

How does its category compare?

All DCK devices

DCK C-Reactive Protein, Antigen, Antiserum, And Control: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DCK

1in 2025

2011: 12012: 12015: 22016: 12017: 12019: 22020: 12022: 12025: 1
2007high 2 (2015) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DCK

Too few DCK 510(k)s in any year since 2007 to give a median.

Recalls per year · DCK

1in 2024

2008: 12009: 12013: 12016: 12024: 1
2007high 1 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DCK

1in 2021

2008: 12011: 12012: 12014: 12021: 1
2007high 1 (2008) · low 0 (2007)2025

MAUDE reports naming product code DCK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—11
20090—10
20100—00
20111—01
20121—01
20130—10
20140—01
20152—00
20161—10
20171—00
20180—00
20192—00
20201—00
20210—01
20221—00
20230—00
20240—10
20251—00
20260—0—
Cite this page

510(k) Report. “K953197: Ektachem Clinical Chemistry Slide C-Reactive Protein (CRP).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K953197

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026