K955848510(k) TraditionalCleared Mar 20, 199684-day reviewClass II21 CFR 888.3040Recalls 23

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Orthofix, Inc. · JDW Pin, Fixation, Threaded

Cleared on Mar 20, 1996 as substantially equivalent to K831576 (Electro-Biology, Inc., 1983) after an 84-day review. Its family tree reaches back 1 generation to 1983. 95 later devices descend from it. It has been recalled 23 times.

Decided
Mar 20, 1996
Review
84 days
Class
Class II
Product code
JDW
Recalls
23
Predicates
1

Its longest line of descent runs back 1 generation to K831576 (1983). 95 later devices descend from it, 14 citing it directly.

Descendants, by year cleared 2014: 10 descendants cleared2012: 3 descendants cleared2007: 1 descendant cleared2005: 4 descendants cleared2013: 3 descendants cleared2003: 3 descendants cleared2020: 8 descendants cleared2019: 5 descendants cleared2025: 8 descendants cleared2011: 2 descendants cleared2001: 1 descendant cleared2017: 8 descendants cleared2015: 6 descendants cleared2016: 3 descendants cleared2010: 2 descendants cleared2021: 5 descendants cleared2024: 5 descendants cleared2018: 4 descendants cleared2026: 8 descendants cleared2023: 5 descendants cleared2022: 1 descendant clearedK831576 · Orthofix · Electro-Biology, Inc. · 198319801990200020102020
Descendants, by generation 1 generation forward: 14 descendants2 generations forward: 21 descendants3 generations forward: 21 descendants4 generations forward: 17 descendants5 generations forward: 12 descendants6 generations forward: 10 descendants K831576 · Orthofix · Electro-Biology, Inc. · 19830+5
  1. Root · 1 generation backK831576 OrthofixElectro-Biology, Inc. · 1983
  2. This deviceK955848 Orthofix ModulsystemOrthofix, Inc. · 1996
  3. Descendant · cites it directly · 35 of its ownK010322 Iskd (Intramedullary Skeletal Kinetic Distractor)Orthodyne, Inc. · 2001 · Recalled
  4. Descendant · cites it directly · 32 of its ownK141078 Orthofix TL-Hex True Lok Hexalobe System (TL-Hex)Orthofix Srl · 2014
  5. Descendant · 2 generations on · 30 of its ownK031219 Modification to Iskd (Intramedullary Skeletal Kinetic Distractor)Orthofix, Inc. · 2003 · Recalled
  6. Descendant · 2 generations on · 29 of its ownK101997 Ellipse Intramedullary Limb Lengthening SystemEllipse Technologies, Inc. · 2011 · Recalled
  7. Descendant · 2 generations on · 26 of its ownK143125 Orthofix True Lok Hexapod System VI.3Orthofix Srl · 2014

Showing 5 of 95 descendants. See all descendants

Descendants cleared per year

2001: 12003: 32005: 42007: 12010: 22011: 22012: 32013: 32014: 102015: 62016: 32017: 82018: 42019: 52020: 82021: 52022: 12023: 52024: 52025: 82026: 8
2001high 10 (2014) · low 0 (2002)2026

Most influential descendants

Showing 5 of 95 · See all
  • K010322Iskd (Intramedullary Skeletal Kinetic Distractor)Orthodyne, Inc. · 2001 · cites it directly35descendants
  • K141078Orthofix TL-Hex True Lok Hexalobe System (TL-Hex)Orthofix Srl · 2014 · cites it directly32descendants
  • K031219Modification to Iskd (Intramedullary Skeletal Kinetic Distractor)Orthofix, Inc. · 2003 · 2 generations on30descendants
  • K101997Ellipse Intramedullary Limb Lengthening SystemEllipse Technologies, Inc. · 2011 · 2 generations on29descendants
  • K143125Orthofix True Lok Hexapod System VI.3Orthofix Srl · 2014 · 2 generations on26descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K831576OrthofixElectro-Biology, Inc.1983—
0 (this device)K955848Orthofix ModulsystemOrthofix, Inc.1996—
+1K010322Iskd (Intramedullary Skeletal Kinetic Distractor)Orthodyne, Inc.20011
+1K141078Orthofix TL-Hex True Lok Hexalobe System (TL-Hex)Orthofix Srl2014—
+2K031219Modification to Iskd (Intramedullary Skeletal Kinetic Distractor)Orthofix, Inc.20031
+2K101997Ellipse Intramedullary Limb Lengthening SystemEllipse Technologies, Inc.20112
+2K143125Orthofix True Lok Hexapod System VI.3Orthofix Srl2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200110
200330
200540
200710
201020
201120
201230
201330
2014100
201560
201630
201780
201840
201950
202080
202150
202210
202350
202450
202580
202680
FIG. 1 — Lineage of K955848, 1983–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JDW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JDW device

  1. #1 K260073 XT3 System; XT3 System, Mini; XT3 System…score 15.0 · 15 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K831576Orthofix
Summary page 2
…bed in this submission have all entered legal distribution in the U.S. as a result of the U.S. Food and Drug Administration’s (FDA’s) clearance of premarket notification K831576. Premarket notification [510(k)] K831576 was submitted to FDA on behalf of Orthofix by Electro-Biology Inc., (EBI), of Parsippany…
Electro-Biology, Inc.1983—

Also cleared against K831576: 5 other devices →

What descends from it? (95)

See all 95

14 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K010322Iskd (Intramedullary Skeletal Kinetic Distractor)Orthodyne, Inc.200135
K141078Orthofix TL-Hex True Lok Hexalobe System (TL-Hex)Orthofix Srl201432
K052196Vilex X-FixVilex, Inc.200513
K113770Orthofix Galaxy Fixation SystemOrthofix Srl20126
K132820Vilex Ultima External Fixation System with HA Coated Pins and WiresVilex, Inc.20145
K200518Pitkar External Fixation System - Rail MechanismS.H.Pitkar Orthotools Pvt., Ltd.20202
K241357XT3 SystemBiodynamik, Inc.20251
K071395Biomet Vision Footring SystemBiomet Manufacturing Corp20071
K051017R-X-Fix External FixatorR-X-Fix20051
K023714Reprocessed External Fixation DeviceAlliance Medical Corp.20031

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 27, 2018Class IIZ-0373-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0380-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0382-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0383-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0384-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0385-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0386-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0387-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0389-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0390-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0392-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0374-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0375-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0376-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0377-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0378-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0379-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0381-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0388-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0391-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Sep 27, 2018Class IIZ-0393-2020Device DesignAs a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Feb 22, 2010Not classifiedZ-1434-2010Nonconforming Material/ComponentDuring standard quality controls, firm identified a potential for breakage of the XCaliber clamp when locking the cam.
Feb 22, 2010Not classifiedZ-1435-2010Nonconforming Material/ComponentDuring standard quality controls, firm identified a potential for breakage of the XCaliber clamp when locking the cam.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K010322 (Orthodyne, Inc., 2001) has gone furthest, with 35 descendants of its own.
  • 5 other devices also cite its predicate K831576.

How does its category compare?

All JDW devices

JDW Pin, Fixation, Threaded: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JDW

2in 2025

2007: 22009: 12010: 22011: 12012: 12013: 22014: 22015: 12017: 12018: 12019: 12020: 22021: 22022: 22025: 22026 (partial year): 1
2007high 2 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JDW

Too few JDW 510(k)s in any year since 2007 to give a median.

Recalls per year · JDW

21in 2018

2009: 22010: 22012: 522017: 22018: 21
2007high 52 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JDW

29in 2025

2007: 252008: 262009: 292010: 292011: 552012: 752013: 692014: 272015: 702016: 462017: 432018: 942019: 5752020: 4022021: 1682022: 1212023: 722024: 652025: 29
2007high 575 (2019) · low 25 (2007)2025

MAUDE reports naming product code JDW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—025
20080—026
20091—229
20102—229
20111—055
20121—5275
20132—069
20142—027
20151—070
20160—046
20171—243
20181—2194
20191—0575
20202—0402
20212—0168
20222—0121
20230—072
20240—065
20252—029
20261—0—

What is it indicated for?

This product is a unilateral external fixation device, which is intended for use in the treatment of bone conditions including leg lengthening, osteotomies, arthrodesis, fracture fixation and other bone conditions amenable to treatment by use of the external fixation modality.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K955848: Orthofix Modulsystem.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K955848

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026