Visart

K961092510(k) TraditionalCleared Jun 14, 199687-day reviewClass II21 CFR 892.1000Recalls 5

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshiba America Medical Systems, In.C · LNH System, Nuclear Magnetic Resonance Imaging

Cleared on Jun 14, 1996 as substantially equivalent to K903768 (Toshiba America Mri, Inc., 1990) after an 87-day review. Its family tree reaches back 1 generation to 1990. 6 later devices descend from it. It has been recalled 5 times.

Decided
Jun 14, 1996
Review
87 days
Class
Class II
Product code
LNH
Recalls
5
Predicates
1

Its longest line of descent runs back 1 generation to K903768 (1990). 6 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2000: 2 descendants cleared1997: 1 descendant cleared1999: 2 descendants cleared2002: 1 descendant clearedReference device: K933018 · MRT-50GP · Toshiba America Mri, Inc. · 1994 · recalledK903768 · Magnetic Resonance Device MRT 150A · Toshiba America Mri, Inc. · 19901990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 1 descendant K903768 · Magnetic Resonance Device MRT 150A · Toshiba America Mri, Inc. · 19900
  1. Root · 1 generation backK903768 Magnetic Resonance Device MRT 150AToshiba America Mri, Inc. · 1990
  2. This deviceK961092 VisartToshiba America Medical Systems, In.C · 1996
  3. Descendant · cites it directly · 5 of its ownK965068 Magnetic Resonanc Dvce Visart V 3.5 SoftwareToshiba America Medical Systems, In.C · 1997
  4. Descendant · 2 generations on · 2 of its ownK990620 ExcelartToshiba America Mri, Inc. · 1999 · Recalled
  5. Descendant · 2 generations on · 2 of its ownK983110 Flexart,Flexart/Hyper, Visart, Visart/HyperToshiba America Mri, Inc. · 1999
  6. Descendant · cites it directly · 1 of its ownK993803 Excelart with PianissimoToshiba America Mri, Inc. · 2000 · Recalled
  7. Descendant · 3 generations on · 1 of its ownK993574 Opart/Prodiga, Opart/Protenza, Opart/ParagonToshiba America Mri, Inc. · 2000

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

1997: 11999: 22000: 22002: 1
1997high 2 (1999) · low 0 (1998)2026

Most influential descendants

Showing 5 of 6 · See all
  • K965068Magnetic Resonanc Dvce Visart V 3.5 SoftwareToshiba America Medical Systems, In.C · 1997 · cites it directly5descendants
  • K990620ExcelartToshiba America Mri, Inc. · 1999 · 2 generations on2descendants
  • K983110Flexart,Flexart/Hyper, Visart, Visart/HyperToshiba America Mri, Inc. · 1999 · 2 generations on2descendants
  • K993803Excelart with PianissimoToshiba America Mri, Inc. · 2000 · cites it directly1descendant
  • K993574Opart/Prodiga, Opart/Protenza, Opart/ParagonToshiba America Mri, Inc. · 2000 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K903768Magnetic Resonance Device MRT 150AToshiba America Mri, Inc.1990—
0 (this device)K961092VisartToshiba America Medical Systems, In.C1996—
+1K965068Magnetic Resonanc Dvce Visart V 3.5 SoftwareToshiba America Medical Systems, In.C1997—
+2K990620ExcelartToshiba America Mri, Inc.19995
+2K983110Flexart,Flexart/Hyper, Visart, Visart/HyperToshiba America Mri, Inc.1999—
+1K993803Excelart with PianissimoToshiba America Mri, Inc.20003
+3K993574Opart/Prodiga, Opart/Protenza, Opart/ParagonToshiba America Mri, Inc.2000—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199710
199920
200020
200210
FIG. 1 — Lineage of K961092, 1990–2002Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LNH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LNH device

  1. #1 K213693 MAGNETOM Vida with syngo MR XA50Ascore 23.0 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K903768Magnetic Resonance Device MRT 150A
Summary page 2
…ists of hardware and software upgrades that improve the performance of the MRT-150A, without introducing new questions of safety or efficacy. The MRT-150A was cleared by K903768A. The improved computer system architecture and user interface were cleared with the Flexart™ via K933018. The increased gradient…
Toshiba America Mri, Inc.1990—
†K933018MRT-50GPToshiba America Mri, Inc.19942

What descends from it? (6)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K965068Magnetic Resonanc Dvce Visart V 3.5 SoftwareToshiba America Medical Systems, In.C19975
K993803Excelart with PianissimoToshiba America Mri, Inc.20001

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 25, 2005Not classifiedZ-0986-05OtherPatient information A is displayed with images for patient B.
Oct 6, 2004Not classifiedZ-0270-05OtherSoftware defect causes images to be rotated 180 degrees.
Jun 7, 2004Not classifiedZ-1045-04OtherWrong positioning may cause patient burns.
May 6, 2004Not classifiedZ-0937-04OtherSoftware anomaly.
Jan 27, 2004Not classifiedZ-0546-04OtherSoftware anomally causes image slices to be in incorrect order.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K965068 (Toshiba America Medical Systems, In.C, 1997) has gone furthest, with 5 descendants of its own.

How does its category compare?

All LNH devices

LNH System, Nuclear Magnetic Resonance Imaging: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LNH

26in 2025

2007: 122008: 192009: 172010: 122011: 162012: 222013: 192014: 172015: 192016: 202017: 182018: 292019: 262020: 322021: 272022: 302023: 302024: 262025: 262026 (partial year): 24
2007high 32 (2020) · low 12 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LNH

122in 2025

2007: 35 days median2008: 53 days median2009: 70 days median2010: 94 days median2011: 81 days median2012: 89 days median2013: 108 days median2014: 108 days median2015: 140 days median2016: 108 days median2017: 98 days median2018: 128 days median2019: 62 days median2020: 121 days median2021: 87 days median2022: 74 days median2023: 124 days median2024: 104 days median2025: 122 days median2026 (partial year): 126 days median
2007high 140 (2015) · low 35 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LNH

8in 2025

2007: 252008: 122009: 72010: 172011: 112012: 52013: 22014: 132015: 12, 2 Class I2016: 12018: 712019: 52020: 42021: 412022: 132023: 10, 1 Class I2024: 42025: 8, 5 Class I2026 (partial year): 43
2007high 71 (2018) · low 0 (2017)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · LNH

1,263in 2025

2007: 1142008: 1672009: 1942010: 1652011: 1822012: 1702013: 1072014: 1192015: 1132016: 1352017: 1212018: 1582019: 1582020: 982021: 842022: 1382023: 1592024: 2812025: 1,263
2007high 1,263 (2025) · low 84 (2021)2025

MAUDE reports naming product code LNH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123525114
2008195312167
200917707194
2010129417165
2011168111182
201222895170
2013191082107
20141710813119
20151914012 (2)113
2016201081135
201718980121
20182912871158
201926625158
202032121498
202127874184
2022307413138
20233012410 (1)159
2024261044281
2025261228 (5)1,263
20262412643—

What is it indicated for?

Anatomical Region: Head, Body, Extremity, Spine, Neck, TMJ, and Heart

Nuclei excited: Hydrogen

Diagnostic Use: Imaging of the whole body (including the head, abdomen, heart, pelvis, spine, blood vessels, limbs and extremities), fluid visualization, 2D/3D Imaging, MR Angiography, MR. Fluoroscopy

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K961092: Visart.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K961092

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026