510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Continuflo Solution Set/Secondary Medication Set

K961225510(k) TraditionalCleared Jun 21, 199684-day reviewClass II21 CFR 880.5440Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Baxter Healthcare Corp · FPA Set, Administration, Intravascular

Cleared on Jun 21, 1996 after an 84-day review. The FDA summary names its predicate without a number. 4 later devices descend from it. It has been recalled 7 times.

Decided
Jun 21, 1996
Review
84 days
Class
Class II
Product code
FPA
Recalls
7
Predicates
0

4 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2014: 1 descendant cleared2013: 1 descendant cleared2023: 1 descendant cleared2015: 1 descendant clearedReference device: K860272 · Solution Administration Sets · Travenol Laboratories, S.A. · 1986 · recalled1995200020052010201520202025
  1. This deviceK961225 Continuflo Solution Set/Secondary Medication SetBaxter Healthcare Corp · 1996
  2. Descendant · cites it directly · 1 of its ownK142011 Solution Set/Extension SetBaxter Healthcare Corporation · 2014
  3. Descendant · cites it directly · 1 of its ownK130245 Large Bore Stopcock with Rotating Male Luer Lock/Large Bore Stopcock…Baxter Healthcare Corp · 2013 · Recalled
  4. Descendant · 2 generations on · 0 of its ownK223175 lntravascular Administration Sets with Stopcock and ManifoldBaxter Healthcare Corporation · 2023
  5. Descendant · 2 generations on · 0 of its ownK150860 Nitroglycerin Sets with DUO-VENT SpikeBaxter Healthcare Corporation · 2015

Most influential descendants

Showing 4 of 4 · See all
  • K142011Solution Set/Extension SetBaxter Healthcare Corporation · 2014 · cites it directly1descendant
  • K130245Large Bore Stopcock with Rotating Male Luer Lock/Large Bore Stopcock with Extension Set…Baxter Healthcare Corp · 2013 · cites it directly1descendant
  • K223175lntravascular Administration Sets with Stopcock and ManifoldBaxter Healthcare Corporation · 2023 · 2 generations on0descendants
  • K150860Nitroglycerin Sets with DUO-VENT SpikeBaxter Healthcare Corporation · 2015 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K961225Continuflo Solution Set/Secondary Medication SetBaxter Healthcare Corp1996—
+1K142011Solution Set/Extension SetBaxter Healthcare Corporation2014—
+1K130245Large Bore Stopcock with Rotating Male Luer Lock/Large Bore Stopcock with…Baxter Healthcare Corp20131
+2K223175lntravascular Administration Sets with Stopcock and ManifoldBaxter Healthcare Corporation2023—
+2K150860Nitroglycerin Sets with DUO-VENT SpikeBaxter Healthcare Corporation2015—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201410
201510
202310
FIG. 1 — Lineage of K961225, 1996–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FPA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FPA device

  1. #1 K243841 Sparta Infusion Set for Insulinscore 13.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Current Baxter Solution Administration Sets
Named without a submission number
—
†K860272Solution Administration SetsTravenol Laboratories, S.A.19868

What descends from it? (4)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K142011Solution Set/Extension SetBaxter Healthcare Corporation20141
K130245Large Bore Stopcock with Rotating Male Luer Lock/Large Bore Stopcock with Extension Set…Baxter Healthcare Corp20131

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 12, 2019Class IIZ-0794-2020Device DesignContinu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.
Dec 12, 2019Class IIZ-0793-2020Device DesignContinu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.
Dec 6, 2011Not classifiedZ-0475-2012Employee errorDuring production of the IV sets, solvent was inadvertently applied to the locking collar of the male luer connector, which may have damaged the component, preventing proper engagement of the locking collar to secure a connection.
Jul 21, 2011Not classifiedZ-2920-2011Employee errorThe Solution Sets are being recalled because the position of the slide clamp and roller clamp may have been reversed during assembly. The reversed orientation of the slide clamp and roller clamp will cause the set to be incorrectly loaded…
Dec 15, 2005Not classifiedZ-0653-06OtherThe extension sets contain microbore female luers that may crack during use, resulting in leakage of medication or incomplete delivery of medication.
Feb 28, 2005Not classifiedZ-0845-05OtherPossible occlusion of the Clearlink injection site valve due to inadequate silicone lubrication of the slit in the upper seal of the valve.
Jul 16, 2003Not classifiedZ-1152-03OtherThere are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K142011 (Baxter Healthcare Corporation, 2014) has gone furthest, with 1 descendant of its own.

How does its category compare?

All FPA devices

FPA Set, Administration, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FPA

11in 2025

2007: 162008: 182009: 102010: 232011: 122012: 212013: 132014: 222015: 172016: 122017: 172018: 42019: 72020: 102021: 132022: 142023: 142024: 72025: 112026 (partial year): 5
2007high 23 (2010) · low 4 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FPA

225in 2025

2007: 75 days median2008: 89 days median2009: 166 days median2010: 120 days median2011: 87 days median2012: 96 days median2013: 90 days median2014: 89 days median2015: 123 days median2016: 195 days median2017: 240 days median2018: 177 days median2019: 161 days median2020: 327 days median2021: 186 days median2022: 213 days median2023: 340 days median2024: 244 days median2025: 225 days median2026 (partial year): 266 days median
2007high 340 (2023) · low 75 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FPA

98in 2025

2007: 172008: 152009: 1662010: 582011: 282012: 16, 4 Class I2013: 642014: 242015: 62016: 102017: 132018: 42019: 132020: 1072021: 12022: 9, 5 Class I2023: 12, 4 Class I2024: 13, 1 Class I2025: 98, 1 Class I2026 (partial year): 10, 2 Class I
2007high 166 (2009) · low 1 (2021)2026

Class I (17 since 2007) Class II/III not classified

Adverse event reports per year · FPA

34,506in 2025

2007: 1,4992008: 1,9412009: 2,5122010: 2,5372011: 4,4072012: 4,9302013: 4,6512014: 4,6952015: 6,6252016: 8,4702017: 6,4622018: 6,5742019: 11,2272020: 9,5632021: 10,4142022: 13,3932023: 10,9162024: 47,2102025: 34,506
2007high 47,210 (2024) · low 1,499 (2007)2025

MAUDE reports naming product code FPA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071675171,499
20081889151,941
2009101661662,512
201023120582,537
20111287284,407
2012219616 (4)4,930
20131390644,651
20142289244,695
20151712366,625
201612195108,470
201717240136,462
2018417746,574
201971611311,227
2020103271079,563
202113186110,414
2022142139 (5)13,393
20231434012 (4)10,916
2024724413 (1)47,210
20251122598 (1)34,506
2026526610 (2)—
Cite this page

510(k) Report. “K961225: Continuflo Solution Set/Secondary Medication Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K961225

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026