Abbott Cell-Dyn 4000 System

K961439510(k) TraditionalCleared Oct 18, 1996186-day reviewClass II21 CFR 864.5220Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Diagnostics · GKZ Counter, Differential Cell

Cleared on Oct 18, 1996 as substantially equivalent to K951496 (Abbott Laboratories, 1995) and 1 other predicate after a 186-day review. Its family tree reaches back 1 generation to 1987. 7 later devices descend from it. It has been recalled 7 times.

Decided
Oct 18, 1996
Review
186 days
Class
Class II
Product code
GKZ
Recalls
7
Predicates
2

Its longest line of descent runs back 1 generation to K872166 (1987). 2 ancestors in all. 7 later devices descend from it, 5 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor2003: 2 descendants cleared2001: 1 descendant cleared2002: 1 descendant cleared2005: 1 descendant cleared2006: 1 descendant cleared2015: 1 descendant clearedK872166 · Facscan Flow Cytometer - Reticcount Reti. Enu. Kit · Bd Becton Dickinson Vacutainer Systems Preanalytic · 19871990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 5 descendants2 generations forward: 2 descendants K872166 · Facscan Flow Cytometer - Reticcount Reti. Enu. Kit · Bd Becton Dickinson Vacutainer Systems Preanalytic · 19870
  1. Root · 1 generation backK872166 Facscan Flow Cytometer - Reticcount Reti. Enu. KitBd Becton Dickinson Vacutainer Systems Preanalytic · 1987
  2. This deviceK961439 Abbott Cell-Dyn 4000 SystemAbbott Diagnostics · 1996
  3. Descendant · cites it directly · 2 of its ownK022200 Pentra 120/Pentra 120 Retic Options SPS, Model Pentra 120Abx Diagnostics · 2002 · Recalled
  4. Descendant · cites it directly · 1 of its ownK050719 Abx Pentra DX 120: Option SPS EvolutionHoriba Abx · 2005 · Recalled
  5. Descendant · cites it directly · 1 of its ownK024002 Pentra 80Abx Diagnostics · 2003 · Recalled
  6. Descendant · cites it directly · 1 of its ownK012934 Cell-Dyn 3200 System with Absolute and Percent ReticulocyteAbbott Laboratories · 2001
  7. Descendant · 2 generations on · 0 of its ownK151133 Pentra XLRHoriba Abx Sas · 2015

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2001: 12002: 12003: 22005: 12006: 12015: 1
2001high 2 (2003) · low 0 (2004)2026

Most influential descendants

Showing 5 of 7 · See all
  • K022200Pentra 120/Pentra 120 Retic Options SPS, Model Pentra 120Abx Diagnostics · 2002 · cites it directly2descendants
  • K050719Abx Pentra DX 120: Option SPS EvolutionHoriba Abx · 2005 · cites it directly1descendant
  • K024002Pentra 80Abx Diagnostics · 2003 · cites it directly1descendant
  • K012934Cell-Dyn 3200 System with Absolute and Percent ReticulocyteAbbott Laboratories · 2001 · cites it directly1descendant
  • K151133Pentra XLRHoriba Abx Sas · 2015 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K872166Facscan Flow Cytometer - Reticcount Reti. Enu. KitBd Becton Dickinson Vacutainer Systems Preanalytic1987—
0 (this device)K961439Abbott Cell-Dyn 4000 SystemAbbott Diagnostics1996—
+1K022200Pentra 120/Pentra 120 Retic Options SPS, Model Pentra 120Abx Diagnostics20022
+1K050719Abx Pentra DX 120: Option SPS EvolutionHoriba Abx20051
+1K024002Pentra 80Abx Diagnostics20032
+1K012934Cell-Dyn 3200 System with Absolute and Percent ReticulocyteAbbott Laboratories2001—
+2K151133Pentra XLRHoriba Abx Sas2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200110
200210
200320
200510
200610
201510
FIG. 1 — Lineage of K961439, 1987–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKZ device

  1. #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K951496Cell-Dyn 3500 Multi-Parameter Automated Hematology Analyzer
Summary page 1
…ent determination. Substantial equivalence is demonstrated between the CELL-DYN® 4000 System and the Abbott CELL-DYN 3500 Multi-Parameter Automated Hematology Analyzer, #K951496/S001 for the hemagram and white cell (WBC) differential parameters. Substantial equivalence is also demonstrated between the CELL…
Abbott Laboratories19951
*K872166Facscan Flow Cytometer - Reticcount Reti. Enu. Kit
Summary page 1
…ntial equivalence is also demonstrated between the CELL-DYN® 4000 System and the Becton Dickinson FACScan™ Flow Cytometer ReticCOUNT™ Reticulocyte Enumeration Software, #K872166/A for reticulocytes. For enumeration of nucleated red blood cells (NRBCs) equivalence is demonstrated by comparison to the manua…
Bd Becton Dickinson Vacutainer Systems Preanalytic1987—

What descends from it? (7)

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K022200Pentra 120/Pentra 120 Retic Options SPS, Model Pentra 120Abx Diagnostics20022
K050719Abx Pentra DX 120: Option SPS EvolutionHoriba Abx20051
K024002Pentra 80Abx Diagnostics20031
K012934Cell-Dyn 3200 System with Absolute and Percent ReticulocyteAbbott Laboratories20011
K032524Advia 60 Hematology AnalyzerBayer Healthcare, LLC20030

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 29, 2009Not classifiedZ-1623-2011OtherA component used in assembly is out of specification, resulting in failure to retract, eject specimens, exposure below the safety zone, piercing of specimen tubes, or error message to display.
Dec 19, 2007Not classifiedZ-1103-2008Material/Component ContaminationMicrobial Contamination: The WBC (White Blood Cell) Reagent Part A was confirmed positive for Pseudomonas contamination.
Dec 22, 2006Not classifiedZ-0565-2007OtherResults for tests of RBC, MCV, RDW, MPV and PLT may be falsely elevated or falsely decreased when the Cell Dyn Diluent/Sheath reagent is used with Cell-Dyn 4000, Ruby and Sapphire systems.
Apr 17, 2006Not classifiedZ-0708-2007OtherHigh platelet background count when using the CELL-DYN 4000 System and/or CELL-DYN Sapphire System; with patient results unacceptable -out-of-range.
May 12, 2004Not classifiedZ-0639-2007OtherIncorrect Test Results-test results may show a higher than expected platelet background count when used on the CELL-DYN 4000 System and report patient results that are unacceptable (out-of-range).
Feb 24, 2004Not classifiedZ-0640-2007OtherIncorrect Test Results-product may show a higher than expected platelet background count when used on the CELL-DYN 4000 System and report patient results that are unacceptable (out-of-range).
Oct 16, 2002Not classifiedZ-0648-03OtherProduct elicits fluorescent signal interference due to contamination.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 2 ancestors has a recall of any class.
  • Of its descendants, K022200 (Abx Diagnostics, 2002) has gone furthest, with 2 descendants of its own.

How does its category compare?

All GKZ devices

GKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GKZ

4in 2025

2007: 62008: 62009: 62010: 42011: 32012: 32013: 12014: 32015: 32016: 32017: 42018: 32019: 52020: 12022: 32023: 42024: 32025: 42026 (partial year): 4
2007high 6 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKZ

108in 2025

2007: 117 days median2008: 99 days median2009: 147 days median2010: 290 days median2011: 315 days median2012: 408 days median2014: 165 days median2015: 273 days median2016: 198 days median2017: 280 days median2018: 270 days median2019: 210 days median2022: 638 days median2023: 329 days median2024: 157 days median2025: 108 days median2026 (partial year): 254 days median
2007high 638 (2022) · low 99 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GKZ

1in 2025

2007: 372008: 52009: 132010: 272011: 282012: 72013: 62015: 32016: 32017: 32018: 2, 2 Class I2019: 22020: 12022: 42023: 12025: 1
2007high 37 (2007) · low 0 (2014)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · GKZ

77in 2025

2007: 2682008: 9252009: 5592010: 8962011: 2,9052012: 2,9592013: 3,0702014: 3,1122015: 1,9722016: 3,3382017: 2292018: 4972019: 9082020: 3422021: 3912022: 1982023: 1552024: 1152025: 77
2007high 3,338 (2016) · low 77 (2025)2025

MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611737268
20086995925
2009614713559
2010429027896
20113315282,905
2012340872,959
20131—63,070
2014316503,112
2015327331,972
2016319833,338
201742803229
201832702 (2)497
201952102908
20201—1342
20210—0391
202236384198
202343291155
202431570115
20254108177
202642540—

What is it indicated for?

The CELL-DYN® 4000 System is a fully automated hematology analyzer intended for in-vitro diagnostic use in the clinical hematology laboratory of a hospital, medical clinic, or reference laboratory.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K961439: Abbott Cell-Dyn 4000 System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K961439

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026