510(k) · Decided
Recalled · Class II · 2024Indeflator 20/30 Inflation Device and 20/30 Priority Pack (with the .096 or .115 Rotating Hemostatic Valve)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Advanced Cardiovascular Systems, Inc. · MAV Syringe, Balloon Inflation
Cleared on May 8, 1996 as substantially equivalent to K872981 (Advanced Cardiovascular Systems, Inc., 1987) and 1 other predicate after a 21-day review. Its family tree reaches back 1 generation to 1987. 9 later devices descend from it. It has been recalled 7 times.
- Decided
- May 8, 1996
- Review
- 21 days
- Class
- Class II
- Product code
- MAV
- Recalls
- 7
- Predicates
- 2
Its longest line of descent runs back 1 generation to K872981 (1987). 2 ancestors in all. 9 later devices descend from it, 2 citing it directly.
- Root · 1 generation backK872981 Acs Indeflator Plus (Modification #2)Advanced Cardiovascular Systems, Inc. · 1987
- This deviceK961471 Indeflator 20/30 Inflation Device and 20/30 Priority Pack (with the…Advanced Cardiovascular Systems, Inc. · 1996
- Descendant · cites it directly · 5 of its ownK042449 Dolphin Inflation DeviceSedat S.A. · 2005 · Recalled
- Descendant · 2 generations on · 3 of its ownK082755 Sedat FlamingoSedat · 2008 · Recalled
- Descendant · cites it directly · 3 of its ownK962495 Indeflator Plus 30/Plus 30 Priority PACK9WITH the.096 or.115…Advanced Cardiovascular Systems, Inc. · 1996 · Recalled
- Descendant · 3 generations on · 1 of its ownK152387 Fluent Inflation DeviceVascular Solutions, Inc. · 2015
- Descendant · 2 generations on · 1 of its ownK102648 Ant Inflation Device/Ant Inflation Device Compact PackShenzhen Ant Hi-Tech Industrial Co., Ltd. · 2011
Showing 5 of 9 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 9 · See all- K042449Dolphin Inflation DeviceSedat S.A. · 2005 · cites it directly5descendants
- K082755Sedat FlamingoSedat · 2008 · 2 generations on3descendants
- K962495Indeflator Plus 30/Plus 30 Priority PACK9WITH the.096 or.115 Rotating Hemostatic Valve)Advanced Cardiovascular Systems, Inc. · 1996 · cites it directly3descendants
- K152387Fluent Inflation DeviceVascular Solutions, Inc. · 2015 · 3 generations on1descendant
- K102648Ant Inflation Device/Ant Inflation Device Compact PackShenzhen Ant Hi-Tech Industrial Co., Ltd. · 2011 · 2 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K872981 | Acs Indeflator Plus (Modification #2) | Advanced Cardiovascular Systems, Inc. | 1987 | — |
| 0 (this device) | K961471 | Indeflator 20/30 Inflation Device and 20/30 Priority Pack (with the .096 or… | Advanced Cardiovascular Systems, Inc. | 1996 | — |
| +1 | K042449 | Dolphin Inflation Device | Sedat S.A. | 2005 | 4 |
| +2 | K082755 | Sedat Flamingo | Sedat | 2008 | 1 |
| +1 | K962495 | Indeflator Plus 30/Plus 30 Priority PACK9WITH the.096 or.115 Rotating… | Advanced Cardiovascular Systems, Inc. | 1996 | 3 |
| +3 | K152387 | Fluent Inflation Device | Vascular Solutions, Inc. | 2015 | — |
| +2 | K102648 | Ant Inflation Device/Ant Inflation Device Compact Pack | Shenzhen Ant Hi-Tech Industrial Co., Ltd. | 2011 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1989 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1996 | 1 | 0 |
| 2005 | 1 | 0 |
| 2008 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2013 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2016 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MAV? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MAV device
- #1 K260035 Digital Indeflator INFLANTEscore 12.6 · 3 generations · no Class I recall
- #2 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K872981 | Acs Indeflator Plus (Modification #2)Summary page 1…20/36 PRIORITY PACK™ Device Name: INDEFLATOR 20/30™ Inflation Device 20/30 PRIORITY PACK™ Accessory Kit Devices to which Equivalence is Claimed: ACS INDEFLATOR 20/20™ (K872981, October 20, 1987 and Baxter Inflation Pro™ II inflation device (K895836, December 20, 1989). ' Indication for Use: INDEFLATOR… | 1987 |
| * | K895836 | Inflation Pro IISummary page 1…30 PRIORITY PACK™ Accessory Kit Devices to which Equivalence is Claimed: ACS INDEFLATOR 20/20™ (K872981, October 20, 1987 and Baxter Inflation Pro™ II inflation device (K895836, December 20, 1989). ' Indication for Use: INDEFLATOR 20/30™ Infiation Device- The INDEFLATOR 20/30™ Inflation Device is recomm… | 1989 |
What descends from it? (9)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 6, 2024 | Class II | Z-0730-2025 | Component change control | Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism. |
| Nov 6, 2024 | Class II | Z-0731-2025 | Component change control | Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism. |
| Nov 6, 2024 | Class II | Z-0732-2025 | Component change control | Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism. |
| Nov 6, 2024 | Class II | Z-0733-2025 | Component change control | Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism. |
| Mar 11, 2022 | Class II | Z-0922-2022 | Component change control | Due to an increase in complaint trend for leaks and intermittent/loose connections. |
| Mar 11, 2022 | Class II | Z-0923-2022 | Component change control | Due to an increase in complaint trend for leaks and intermittent/loose connections. |
| Mar 11, 2022 | Class II | Z-0921-2022 | Component change control | Due to an increase in complaint trend for leaks and intermittent/loose connections. |
The 5 latest of 7 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All MAV devicesMAV Syringe, Balloon Inflation: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MAV
1in 2022
2026 to Oct 25, outlined.
Median 510(k) review, days · MAV
69in 2016
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MAV
4in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MAV
66in 2025
MAUDE reports naming product code MAV, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 37 |
| 2008 | 1 | — | 0 | 50 |
| 2009 | 0 | — | 0 | 99 |
| 2010 | 0 | — | 3 | 88 |
| 2011 | 1 | — | 0 | 94 |
| 2012 | 4 | 85 | 0 | 251 |
| 2013 | 1 | — | 0 | 245 |
| 2014 | 2 | — | 2 | 92 |
| 2015 | 3 | 116 | 0 | 121 |
| 2016 | 3 | 69 | 1 | 124 |
| 2017 | 0 | — | 0 | 246 |
| 2018 | 0 | — | 1 | 300 |
| 2019 | 0 | — | 0 | 199 |
| 2020 | 0 | — | 0 | 101 |
| 2021 | 0 | — | 0 | 186 |
| 2022 | 1 | — | 5 | 437 |
| 2023 | 0 | — | 0 | 220 |
| 2024 | 0 | — | 4 | 420 |
| 2025 | 0 | — | 0 | 66 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
INDEFLATOR 20/30™ Inflation Device- The INDEFLATOR 20/30™ Inflation Device is recommended for use during cardiovascular procedures in conjunction with interventional devices such as balloon catheters to create and monitor pressure in the balloon catheter.
20/30 PRIORITY PACK™ Accessory Kit- The Accessory Kit is recommended for use during cardiovascular procedures in conjunction with interventional and/or diagnostic devices (e.g., balloon dilatation catheters, artherectomy devices, stent delivery systems, intravascular ultrasound devices.)
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K961471 · printed Nov 4, 2026
Cite this page
510(k) Report. “K961471: Indeflator 20/30 Inflation Device and 20/30 Priority Pack (with the .096 or .115 Rotating Hemostatic Valve).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K961471
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026