510(k) · Cleared
Recalled · Class II · 2022H/S Catheter Set (61-3005, 61-3007, 61-3605, 61-3607)
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ackrad Laboratories · LKF Cannula, Manipulator/Injector, Uterine
Cleared on Aug 8, 1996 as substantially equivalent to K953034 (Ackrad Laboratories, 1995) after a 94-day review. Its family tree reaches back 1 generation to 1995. 6 later devices descend from it. It has been recalled 2 times.
- Decided
- Aug 8, 1996
- Review
- 94 days
- Class
- Class II
- Product code
- LKF
- Recalls
- 2
- Predicates
- 1
Its longest line of descent runs back 1 generation to K953034 (1995). 6 later devices descend from it, 1 citing it directly.
- Root · 1 generation backK953034 H/S Catheter Set (61-3005, 61-3007, 61-3605, 61-3607)Ackrad Laboratories · 1995
- This deviceK961752 H/S Catheter Set (61-3005, 61-3007, 61-3605, 61-3607)Ackrad Laboratories · 1996
- Descendant · cites it directly · 5 of its ownK013972 H/S Elliptosphere Catheter Set, Model 61-4005Ackard Laboratories · 2001 · Recalled
- Descendant · 2 generations on · 2 of its ownK083594 EZ HSG Catheter.Foamseal CatheterObg Products · 2009
- Descendant · 3 generations on · 1 of its ownK180300 Margolin HSG Cannula, Goldstein Sonohysterography CatheterCook Incorporated · 2018
- Descendant · 4 generations on · 0 of its ownK222798 Rejoni Intrauterine CatheterRejoni, Inc. · 2022
- Descendant · 2 generations on · 0 of its ownK063611 Saline/ Contrast Media Infusion CatheterUtah Medical Products, Inc. · 2007
Showing 5 of 6 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 6 · See all- K013972H/S Elliptosphere Catheter Set, Model 61-4005Ackard Laboratories · 2001 · cites it directly5descendants
- K083594EZ HSG Catheter.Foamseal CatheterObg Products · 2009 · 2 generations on2descendants
- K180300Margolin HSG Cannula, Goldstein Sonohysterography CatheterCook Incorporated · 2018 · 3 generations on1descendant
- K222798Rejoni Intrauterine CatheterRejoni, Inc. · 2022 · 4 generations on0descendants
- K063611Saline/ Contrast Media Infusion CatheterUtah Medical Products, Inc. · 2007 · 2 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K953034 | H/S Catheter Set (61-3005, 61-3007, 61-3605, 61-3607) | Ackrad Laboratories | 1995 | — |
| 0 (this device) | K961752 | H/S Catheter Set (61-3005, 61-3007, 61-3605, 61-3607) | Ackrad Laboratories | 1996 | — |
| +1 | K013972 | H/S Elliptosphere Catheter Set, Model 61-4005 | Ackard Laboratories | 2001 | 2 |
| +2 | K083594 | EZ HSG Catheter.Foamseal Catheter | Obg Products | 2009 | — |
| +3 | K180300 | Margolin HSG Cannula, Goldstein Sonohysterography Catheter | Cook Incorporated | 2018 | — |
| +4 | K222798 | Rejoni Intrauterine Catheter | Rejoni, Inc. | 2022 | — |
| +2 | K063611 | Saline/ Contrast Media Infusion Catheter | Utah Medical Products, Inc. | 2007 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2001 | 1 | 0 |
| 2002 | 1 | 0 |
| 2007 | 1 | 0 |
| 2009 | 1 | 0 |
| 2018 | 1 | 0 |
| 2022 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LKF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LKF device
- #1 K241693 FemChec Controlled Saline-Air Device (FCD-250)score 11.6 · 3 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K953034 | H/S Catheter Set (61-3005, 61-3007, 61-3605, 61-3607)Summary page 1…Contact person: Bernard Ackerman, Ph.D. Date: May 2, 1996 2. Device name Modified Hysterosalpingography Set 3. Predicate Device Modified Hysterosalpingography Set - K953034 4. Description of the device (See Exhibit 2) The Modified Hysterosalpingography Set consists of three parts: a double-lumen, bal… | 1995 |
What descends from it? (6)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K013972 | H/S Elliptosphere Catheter Set, Model 61-4005 | 2001 | 5 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 25, 2022 | Class II | Z-1613-2022 | Environmental control | Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. |
| Mar 31, 2022 | Class II | Z-0502-2024 | Component design/selection | Incorrect bonding material used for assembly of the catheter, may fail, resulting in detachment of component parts, broken components, or leakage. This breakage may result in delay of procedure |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All LKF devicesLKF Cannula, Manipulator/Injector, Uterine: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LKF
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LKF
136in 2024
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LKF
3in 2022
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LKF
95in 2025
MAUDE reports naming product code LKF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 141 | 0 | 36 |
| 2008 | 0 | — | 2 | 59 |
| 2009 | 2 | — | 0 | 53 |
| 2010 | 6 | 178 | 0 | 46 |
| 2011 | 3 | 187 | 0 | 98 |
| 2012 | 0 | — | 2 | 104 |
| 2013 | 3 | 266 | 0 | 53 |
| 2014 | 3 | 169 | 0 | 45 |
| 2015 | 5 | 140 | 0 | 53 |
| 2016 | 1 | — | 0 | 62 |
| 2017 | 1 | — | 0 | 64 |
| 2018 | 4 | 267 | 0 | 23 |
| 2019 | 1 | — | 0 | 76 |
| 2020 | 0 | — | 0 | 119 |
| 2021 | 0 | — | 0 | 189 |
| 2022 | 4 | 79 | 3 | 100 |
| 2023 | 0 | — | 0 | 88 |
| 2024 | 3 | 136 | 0 | 125 |
| 2025 | 2 | — | 0 | 95 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Modified Hysterosalpingography Set is to be used for the injection of contrast material in the examination of the uterus and fallopian tubes.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K961752 · printed Nov 6, 2026
Cite this page
510(k) Report. “K961752: H/S Catheter Set (61-3005, 61-3007, 61-3605, 61-3607).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K961752
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026