510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Disposable Endoscopic Retrieval Device

K962676510(k) TraditionalCleared Oct 4, 199687-day reviewClass II21 CFR 876.4680Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

United States Endoscopy Group, Inc. · FGO Dislodger, Stone, Flexible

Cleared on Oct 4, 1996 after an 87-day review. The FDA summary names no predicate by number. It has been recalled 2 times.

Decided
Oct 4, 1996
Review
87 days
Class
Class II
Product code
FGO
Recalls
2
Predicates
0

The FDA summary names no predicate by number.

FIG. 1 — Lineage of K962676Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FGO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 10, 2015Class IIZ-1453-2015Under Investigation by firmA wire component on the distal grasping assembly of the device became detached..
Nov 14, 2011Class IIZ-0006-2013Device DesignWhen the outer catheter is either coiled or in a contorted configuration outside of the endoscope and the patient, and excessive force is applied to the handle, the jaws may become inoperable and may become stuck in the open condition.

No ancestor or descendant has a Class I recall.

How does its category compare?

All FGO devices

FGO Dislodger, Stone, Flexible: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FGO

FDA has decided no FGO device since 2007.

Median 510(k) review, days · FGO

Too few FGO 510(k)s in any year since 2007 to give a median.

Recalls per year · FGO

1in 2011

2011: 1
2007high 1 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FGO

10in 2025

2007: 22008: 232009: 322010: 112011: 152012: 132013: 152014: 42015: 152016: 212017: 172018: 142019: 252020: 192021: 142022: 182023: 142024: 192025: 10
2007high 32 (2009) · low 2 (2007)2025

MAUDE reports naming product code FGO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—02
20080—023
20090—032
20100—011
20110—115
20120—013
20130—015
20140—04
20150—015
20160—021
20170—017
20180—014
20190—025
20200—019
20210—014
20220—018
20230—014
20240—019
20250—010
20260—0—
Cite this page

510(k) Report. “K962676: Disposable Endoscopic Retrieval Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K962676

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)