510(k) · Cleared
Recalled · 2007Ecat Exact & Ecat Exact HR+
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Cti Pet Systems, Inc. · KPS System, Tomography, Computed, Emission
Cleared on Oct 15, 1996 after an 89-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. 6 later devices descend from it. It has been recalled 14 times.
- Decided
- Oct 15, 1996
- Review
- 89 days
- Class
- Class II
- Product code
- KPS
- Recalls
- 14
- Predicates
- 0
6 later devices descend from it, 5 citing it directly.
- This deviceK962797 Ecat Exact & Ecat Exact HR+Cti Pet Systems, Inc. · 1996
- Descendant · cites it directly · 2 of its ownK981027 E.Cam Lso 311 Pet/Spect SystemCti Pet Systems, Inc. · 1998
- Descendant · cites it directly · 1 of its ownK130269 Ecat Scintron PetMiegmbh · 2013
- Descendant · 2 generations on · 0 of its ownK201807 AncorisMiegmbh · 2020
- Descendant · cites it directly · 0 of its ownK040933 Ecat HRRTCti Pet Systems, Inc. · 2004
- Descendant · cites it directly · 0 of its ownK021633 Ecat Vision Pet Scanner, Models 3000 & 5000Cti Pet Systems, Inc. · 2002
Showing 5 of 6 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 6 · See all- K981027E.Cam Lso 311 Pet/Spect SystemCti Pet Systems, Inc. · 1998 · cites it directly2descendants
- K130269Ecat Scintron PetMiegmbh · 2013 · cites it directly1descendant
- K201807AncorisMiegmbh · 2020 · 2 generations on0descendants
- K040933Ecat HRRTCti Pet Systems, Inc. · 2004 · cites it directly0descendants
- K021633Ecat Vision Pet Scanner, Models 3000 & 5000Cti Pet Systems, Inc. · 2002 · cites it directly0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K962797 | Ecat Exact & Ecat Exact HR+ | Cti Pet Systems, Inc. | 1996 | — |
| +1 | K981027 | E.Cam Lso 311 Pet/Spect System | Cti Pet Systems, Inc. | 1998 | — |
| +1 | K130269 | Ecat Scintron Pet | Miegmbh | 2013 | — |
| +2 | K201807 | Ancoris | Miegmbh | 2020 | — |
| +1 | K040933 | Ecat HRRT | Cti Pet Systems, Inc. | 2004 | — |
| +1 | K021633 | Ecat Vision Pet Scanner, Models 3000 & 5000 | Cti Pet Systems, Inc. | 2002 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 2 | 0 |
| 2002 | 1 | 0 |
| 2004 | 1 | 0 |
| 2013 | 1 | 0 |
| 2020 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KPS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KPS device
- #1 K253564 uMI Panvivo (uMI Panvivo); uMI Panvivo (uMI…score 18.6 · 17 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.
What descends from it? (6)
5 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jul 23, 2007 | Not classified | Z-0271-2008 | Software design | Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions. |
| Jul 23, 2007 | Not classified | Z-0273-2008 | Software design | Incorrect Data Results - ECAT System's software versions 7.2.2 and 7.4 may calculate the decay correction for emission data incorrectly while performing whole-body studies with all transmissions followed by all emissions. |
| Jul 23, 2007 | Not classified | Z-0251-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0252-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0270-2008 | Software design | Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions. |
| Jul 23, 2007 | Not classified | Z-0246-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0247-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0249-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0248-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0250-2008 | Software design | Visual asymmetric appearance; An asymmetry can be introduced into the attenuation-corrected PET images when using iterative reconstruction in combination with a specific number of subsets and reconstruction matrix sizes. (This problem… |
| Jul 23, 2007 | Not classified | Z-0269-2008 | Software design | Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions. |
| Jul 23, 2007 | Not classified | Z-0272-2008 | Software design | Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions. |
| Apr 5, 2005 | Not classified | Z-0819-05 | Other | Failure to properly secure the Mobile Bed Supports during transit could cause the support screws on the Patient Handling System to become loose and possibly break due to increased vibration during transit. |
| Mar 4, 2003 | Not classified | Z-0674-03 | Other | Incorrect patient information is displayed in the reconstructed images of a whole body patient scan. |
The 5 latest of 14 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K981027 (Cti Pet Systems, Inc., 1998) has gone furthest, with 2 descendants of its own.
How does its category compare?
All KPS devicesKPS System, Tomography, Computed, Emission: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · KPS
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KPS
65in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KPS
16in 2025
Class I (17 since 2007) Class II/III not classified
Adverse event reports per year · KPS
36in 2025
MAUDE reports naming product code KPS, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 8 | 38 | 22 | 38 |
| 2008 | 17 | 25 | 16 | 11 |
| 2009 | 9 | 42 | 14 | 15 |
| 2010 | 3 | 91 | 4 | 44 |
| 2011 | 7 | 28 | 34 | 56 |
| 2012 | 6 | 94 | 13 | 34 |
| 2013 | 11 | 84 | 13 | 240 |
| 2014 | 11 | 92 | 18 | 33 |
| 2015 | 3 | 112 | 5 | 24 |
| 2016 | 11 | 65 | 16 | 10 |
| 2017 | 11 | 104 | 11 | 31 |
| 2018 | 13 | 59 | 12 (3) | 26 |
| 2019 | 3 | 60 | 5 (5) | 105 |
| 2020 | 12 | 72 | 8 | 22 |
| 2021 | 15 | 66 | 4 | 10 |
| 2022 | 5 | 60 | 8 (6) | 15 |
| 2023 | 12 | 63 | 6 (3) | 22 |
| 2024 | 11 | 59 | 15 | 23 |
| 2025 | 12 | 65 | 16 | 36 |
| 2026 | 6 | 67 | 6 | — |
https://510kreport.com/device/K962797 · printed Nov 5, 2026
Cite this page
510(k) Report. “K962797: Ecat Exact & Ecat Exact HR+.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K962797
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026