510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Gentamicin Flex Reagent Cartridge

K962819510(k) TraditionalCleared Sep 13, 199656-day reviewClass II21 CFR 862.3450Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dade Intl., Inc. · LCD Enzyme Immunoassay, Gentamicin

Cleared on Sep 13, 1996 after a 56-day review. The FDA summary names its predicate without a number. 1 later device descends from it. It has been recalled 2 times.

Decided
Sep 13, 1996
Review
56 days
Class
Class II
Product code
LCD
Recalls
2
Predicates
0

1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2006: 1 descendant cleared1995200020052010201520202025
  1. This deviceK962819 Gentamicin Flex Reagent CartridgeDade Intl., Inc. · 1996
  2. Descendant · cites it directly · 0 of its ownK062024 Dimension Vista Digitoxin, Digoxin, Gentamicin…Dade Behring, Inc. · 2006 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K062024Dimension Vista Digitoxin, Digoxin, Gentamicin, N-Acetylprocainamide, Phenytoin…Dade Behring, Inc. · 2006 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K962819Gentamicin Flex Reagent CartridgeDade Intl., Inc.1996—
+1K062024Dimension Vista Digitoxin, Digoxin, Gentamicin, N-Acetylprocainamide…Dade Behring, Inc.20064

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
FIG. 1 — Lineage of K962819, 1996–2006Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LCD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LCD device

  1. #1 K243500 ARCHITECT iGentamicinscore 9.9 · 2 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Abbott AXSYM® Gentamicin assay
Named without a submission number
—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K062024Dimension Vista Digitoxin, Digoxin, Gentamicin, N-Acetylprocainamide, Phenytoin…Dade Behring, Inc.20060

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 19, 2018Class IIZ-2182-2018Under Investigation by firmThe affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2…
Apr 19, 2018Class IIZ-2181-2018Under Investigation by firmThe affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K062024 (Dade Behring, Inc., 2006) has gone furthest, with 0 descendants of its own.

How does its category compare?

All LCD devices

LCD Enzyme Immunoassay, Gentamicin: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LCD

1in 2025

2011: 12025: 1
2007high 1 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LCD

Too few LCD 510(k)s in any year since 2007 to give a median.

Recalls per year · LCD

1in 2010

2007: 12009: 42010: 1
2007high 4 (2009) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LCD

2in 2024

2009: 32010: 42011: 32012: 192013: 162014: 32015: 22016: 82017: 12018: 22019: 12020: 12024: 2
2007high 19 (2012) · low 0 (2007)2025

MAUDE reports naming product code LCD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20080—00
20090—43
20100—14
20111—03
20120—019
20130—016
20140—03
20150—02
20160—08
20170—01
20180—02
20190—01
20200—01
20210—00
20220—00
20230—00
20240—02
20251—00
20260—0—

What is it indicated for?

The GENT FLEX™ reagent cartridge used on the Dimension® clinical chemistry system is an in vitro diagnostic test intended to measure gentamicin, an aminoglycoside antibiotic drug, in human specimens. Measurements obtained by this assay are used in the diagnosis and treatment of gentamicin overdose and in monitoring levels of gentamicin to ensure appropriate therapy.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K962819: Gentamicin Flex Reagent Cartridge.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K962819

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026