Unimed Coated Needle Electrode/Coated Blade Electrode/Coated Ball Electrode

K962935510(k) TraditionalCleared Sep 3, 199636-day reviewClass II21 CFR 878.4400Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Unimed Surgical Products, Inc. · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Sep 3, 1996 after a 36-day review. The FDA summary names no predicate by number. 30 later devices descend from it. It has been recalled 13 times.

Decided
Sep 3, 1996
Review
36 days
Class
Class II
Product code
GEI
Recalls
13
Predicates
0

30 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2013: 1 descendant cleared2008: 2 descendants cleared1997: 1 descendant cleared2002: 1 descendant cleared2018: 2 descendants cleared2012: 1 descendant cleared2014: 4 descendants cleared2010: 3 descendants cleared2016: 2 descendants cleared2015: 2 descendants cleared2017: 1 descendant cleared2020: 3 descendants cleared2019: 1 descendant cleared2022: 3 descendants cleared2026: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant cleared1995200020052010201520202025
  1. This deviceK962935 Unimed Coated Needle Electrode/Coated Blade Electrode/Coated Ball…Unimed Surgical Products, Inc. · 1996
  2. Descendant · cites it directly · 25 of its ownK073616 Fas-Clean Electrosurgical Coated Active BladeElectromedics Co., Ltd. · 2008
  3. Descendant · 2 generations on · 24 of its ownK082786 Pulsar Generator, Peak Plasmablade; 4.0, Needle, and Ext Tissue…Peak Surgical, Inc. · 2008 · Recalled
  4. Descendant · 3 generations on · 8 of its ownK093695 Peak Plasmablade 3.OS Tissue Dissection DevicePeak Surgical, Inc. · 2010
  5. Descendant · 3 generations on · 7 of its ownK142410 Elliquence ElectrodesElliquence, LLC · 2015
  6. Descendant · 3 generations on · 6 of its ownK160041 Cobbra RF Tissue DissectorElliquence, LLC · 2016

Showing 5 of 30 descendants. See all descendants

Descendants cleared per year

1997: 12002: 12008: 22010: 32012: 12013: 12014: 42015: 22016: 22017: 12018: 22019: 12020: 32022: 32024: 12025: 12026: 1
1997high 4 (2014) · low 0 (1998)2026

Most influential descendants

Showing 5 of 30 · See all
  • K073616Fas-Clean Electrosurgical Coated Active BladeElectromedics Co., Ltd. · 2008 · cites it directly25descendants
  • K082786Pulsar Generator, Peak Plasmablade; 4.0, Needle, and Ext Tissue Dissection DevicesPeak Surgical, Inc. · 2008 · 2 generations on24descendants
  • K093695Peak Plasmablade 3.OS Tissue Dissection DevicePeak Surgical, Inc. · 2010 · 3 generations on8descendants
  • K142410Elliquence ElectrodesElliquence, LLC · 2015 · 3 generations on7descendants
  • K160041Cobbra RF Tissue DissectorElliquence, LLC · 2016 · 3 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K962935Unimed Coated Needle Electrode/Coated Blade Electrode/Coated Ball ElectrodeUnimed Surgical Products, Inc.1996—
+1K073616Fas-Clean Electrosurgical Coated Active BladeElectromedics Co., Ltd.2008—
+2K082786Pulsar Generator, Peak Plasmablade; 4.0, Needle, and Ext Tissue Dissection…Peak Surgical, Inc.20081
+3K093695Peak Plasmablade 3.OS Tissue Dissection DevicePeak Surgical, Inc.2010—
+3K142410Elliquence ElectrodesElliquence, LLC2015—
+3K160041Cobbra RF Tissue DissectorElliquence, LLC2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199710
200210
200820
201030
201210
201310
201440
201520
201620
201710
201820
201910
202030
202230
202410
202510
202610
FIG. 1 — Lineage of K962935, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (30)

See all 30

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K073616Fas-Clean Electrosurgical Coated Active BladeElectromedics Co., Ltd.200825
K970066Unimed Coated Laparoscopic/Arthroscopy/Suction-Irrigation ElectrodeUnimed Surgical Products, Inc.19973
K132224Unimed Surgical Coated Electrosurgical ElectrodesUnimed Surgical Products, Inc.20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 15, 2013Class IIZ-1347-2013No Marketing ApplicationFirm is recalling due to concerns regarding changes that were made to the PTFE coating materials.
Jun 16, 2010Class IIZ-2292-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2286-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2287-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2296-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2294-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2288-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2295-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2290-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2293-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2289-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Jun 16, 2010Class IIZ-2291-2012Process controlOn 6/16/2010 Unimed Surgical Products, Inc., 10401 Belcher Road, Largo, FL 33777 initiated a recall of the Medline Industries, Inc. Electrosurgical electrodes used for cutting and coagulation of soft tissue due to concerns of sterility.
Aug 28, 2009Not classifiedZ-0069-2010Process controlUnimed Surgical Products, Inc., Largo, FL initiated a voluntary recall of the Extended PTFE Coated Blade electrosurgical electrodes, product number ES0014, due to concerns regarding product sterility for Lot #042209-03.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K073616 (Electromedics Co., Ltd., 2008) has gone furthest, with 25 descendants of its own.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The unimed Coated Electrodes are indicated for use in the cuffing and coagulation of soft tissue.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K962935: Unimed Coated Needle Electrode/Coated Blade Electrode/Coated Ball Electrode.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K962935

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026