510(k) Report

510(k) · Cleared

Cleared

Quartet Clinical System

K963761510(k) TraditionalCleared Apr 18, 1997212-day reviewClass II21 CFR 868.5905

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Respironics, Inc. · BZD Ventilator, Non-Continuous (Respirator)

Cleared on Apr 18, 1997 as substantially equivalent to K883825 (Respironics, Inc., 1988) and 2 other predicates after a 212-day review. Its family tree reaches back 1 generation to 1988.

Decided
Apr 18, 1997
Review
212 days
Class
Class II
Product code
BZD
Recalls
None
Predicates
3

Its longest line of descent runs back 1 generation to K883825 (1988). 3 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1994: 1 other ancestor1991: 1 other ancestorReference device: K953930 · Virtuoso Smart Cpap System · Respironics, Inc. · 1996K883825 · Modified Sanders BI-Papeasy · Respironics, Inc. · 19881986198819901992199419961998
Other ancestors, by generation 1 generation back: 2 other ancestors K883825 · Modified Sanders BI-Papeasy · Respironics, Inc. · 19880
  1. Root · 1 generation backK883825 Modified Sanders BI-PapeasyRespironics, Inc. · 1988
  2. This deviceK963761 Quartet Clinical SystemRespironics, Inc. · 1997
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K883825Modified Sanders BI-PapeasyRespironics, Inc.1988—
0 (this device)K963761Quartet Clinical SystemRespironics, Inc.1997—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199410
FIG. 1 — Lineage of K963761, 1988–1997Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BZD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZD device

  1. #1 K250624 myAirscore 15.7 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K883825Modified Sanders BI-Papeasy
Summary page 1
…tion Name Noncontinuous Ventilator Common/Usual Name CPAP System Proprietary Name Respironics Quartet Clinical System Predicate Device(s) Respironics BiPAP S-D System (K883825 & K905540), Innovative Medical Systems DPAP System (K933184). The technological characteristics of the Quartet Clinical System…
Respironics, Inc.1988—
*K905540Respironics Bipap Remote Control Unit
Summary page 1
…Noncontinuous Ventilator Common/Usual Name CPAP System Proprietary Name Respironics Quartet Clinical System Predicate Device(s) Respironics BiPAP S-D System (K883825 & K905540), Innovative Medical Systems DPAP System (K933184). The technological characteristics of the Quartet Clinical System are equiva…
Respironics, Inc.1991—
*K933184Positive Airway Pressure System
Summary page 1
…tem Proprietary Name Respironics Quartet Clinical System Predicate Device(s) Respironics BiPAP S-D System (K883825 & K905540), Innovative Medical Systems DPAP System (K933184). The technological characteristics of the Quartet Clinical System are equivalent to the predicate devices as listed above in te…
Innovative Medical Systems, Inc.1994—
†K953930Virtuoso Smart Cpap SystemRespironics, Inc.1996—

Also cleared against K883825: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 3 ancestors has a recall.
  • 4 other devices also cite its predicate K883825.

How does its category compare?

All BZD devices

BZD Ventilator, Non-Continuous (Respirator): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BZD

7in 2025

2007: 212008: 172009: 232010: 122011: 232012: 212013: 232014: 142015: 112016: 122017: 52018: 52019: 62020: 62021: 62022: 42023: 52024: 102025: 72026 (partial year): 10
2007high 23 (2009) · low 4 (2022)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZD

189in 2025

2007: 83 days median2008: 108 days median2009: 90 days median2010: 122 days median2011: 125 days median2012: 133 days median2013: 127 days median2014: 172 days median2015: 224 days median2016: 217 days median2017: 221 days median2018: 280 days median2019: 231 days median2020: 137 days median2021: 192 days median2022: 253 days median2023: 204 days median2024: 132 days median2025: 189 days median2026 (partial year): 214 days median
2007high 280 (2018) · low 83 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · BZD

8in 2023

2007: 292008: 22009: 12011: 12019: 4, 4 Class I2023: 8, 7 Class I2026 (partial year): 2, 2 Class I
2007high 29 (2007) · low 0 (2010)2026

Class I (13 since 2007) Class II/III not classified

Adverse event reports per year · BZD

60,279in 2025

2007: 652008: 672009: 1472010: 1432011: 1892012: 1392013: 822014: 862015: 1142016: 1182017: 1482018: 1012019: 1882020: 1262021: 6,0662022: 90,7682023: 27,9872024: 68,2622025: 60,279
2007high 90,768 (2022) · low 65 (2007)2025

MAUDE reports naming product code BZD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200721832965
200817108267
200923901147
2010121220143
2011231251189
2012211330139
201323127082
201414172086
2015112240114
2016122170118
201752210148
201852800101
201962314 (4)188
202061370126
2021619206,066
20224253090,768
202352048 (7)27,987
202410132068,262
20257189060,279
2026102142 (2)—

What is it indicated for?

The Quartet Clinical System delivers continuous positive airway pressure (CPAP) and bi-level positive airway pressure therapies for the treatment of adult OSA. The Quartet Clinical System is intended for use by a qualified clinician (any individual trained to perform sleep study diagnoses) in clinical (hospital or sleep laboratory) settings where adult patients suffering from OSA are diagnosed and treated. The device is used in conjunction with masks, headgear, and various combinations of patient circuit accessories.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K963761: Quartet Clinical System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K963761

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026