510(k) Report

510(k) · Cleared

Recalled · 2007

Axsym CA 125

K964020510(k) TraditionalCleared Nov 4, 1997393-day reviewClass II21 CFR 866.6010Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Laboratories · LTK Test, Epithelial Ovarian Tumor-Associated Antigen (Ca125)

Cleared on Nov 4, 1997 after a 393-day review. The FDA summary names its predicate without a number. 9 later devices descend from it. It has been recalled 4 times.

Decided
Nov 4, 1997
Review
393 days
Class
Class II
Product code
LTK
Recalls
4
Predicates
0

9 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2004: 1 descendant cleared2008: 1 descendant cleared2015: 1 descendant cleared2010: 2 descendants cleared2011: 1 descendant cleared2012: 1 descendant cleared2016: 1 descendant cleared2025: 1 descendant cleared1995200020052010201520202025
  1. This deviceK964020 Axsym CA 125Abbott Laboratories · 1997
  2. Descendant · cites it directly · 8 of its ownK042731 Architect CA 125 II AssayFujirebio Diagnostics,Inc. · 2004
  3. Descendant · 2 generations on · 7 of its ownK072939 HE4 Eia, Model: 404-10 USFujirebio Diagnostics,Inc. · 2008
  4. Descendant · 3 generations on · 4 of its ownK093957 Architect HE4, Architect HE4 Calibrators and Architect HE4 Controls…Fujirebio Diagnostics,Inc. · 2010
  5. Descendant · 4 generations on · 2 of its ownK112624 Elecsys HE4, Elecsys HE4 Calset, Elecsys Precicontrol HE4, Elecsys…Roche Diagnostics · 2012
  6. Descendant · 5 generations on · 1 of its ownK160915 Elecsys CYFRA 21-1, Elecsys CYFRA 21-1 CalSet, PreciControl Tumor…Roche Diagnostics · 2016

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2004: 12008: 12010: 22011: 12012: 12015: 12016: 12025: 1
2004high 2 (2010) · low 0 (2005)2026

Most influential descendants

Showing 5 of 9 · See all
  • K042731Architect CA 125 II AssayFujirebio Diagnostics,Inc. · 2004 · cites it directly8descendants
  • K072939HE4 Eia, Model: 404-10 USFujirebio Diagnostics,Inc. · 2008 · 2 generations on7descendants
  • K093957Architect HE4, Architect HE4 Calibrators and Architect HE4 Controls, Models 2, 2P54…Fujirebio Diagnostics,Inc. · 2010 · 3 generations on4descendants
  • K112624Elecsys HE4, Elecsys HE4 Calset, Elecsys Precicontrol HE4, Elecsys HE4 Calcheck 5Roche Diagnostics · 2012 · 4 generations on2descendants
  • K160915Elecsys CYFRA 21-1, Elecsys CYFRA 21-1 CalSet, PreciControl Tumor MarkerRoche Diagnostics · 2016 · 5 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K964020Axsym CA 125Abbott Laboratories1997—
+1K042731Architect CA 125 II AssayFujirebio Diagnostics,Inc.2004—
+2K072939HE4 Eia, Model: 404-10 USFujirebio Diagnostics,Inc.2008—
+3K093957Architect HE4, Architect HE4 Calibrators and Architect HE4 Controls, Models 2…Fujirebio Diagnostics,Inc.2010—
+4K112624Elecsys HE4, Elecsys HE4 Calset, Elecsys Precicontrol HE4, Elecsys HE4…Roche Diagnostics2012—
+5K160915Elecsys CYFRA 21-1, Elecsys CYFRA 21-1 CalSet, PreciControl Tumor MarkerRoche Diagnostics2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200810
201020
201110
201210
201510
201610
202510
FIG. 1 — Lineage of K964020, 1997–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LTK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LTK device

  1. #1 K240479 Access OV Monitorscore 7.3 · 1 generation · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—ABBOTT CA 125 II™ RIA assay
Named without a submission number
—

What descends from it? (9)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042731Architect CA 125 II AssayFujirebio Diagnostics,Inc.20048

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 23, 2007Not classifiedZ-1026-2007OtherFailure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent kit on the AxYM instrument.
Nov 25, 2003Not classifiedZ-0318-04OtherThe CA 125 assay product inserts list the incorrect storage instructions for the CA 125 Calibrators and Controls, and the product information letter (PI-P) shipped with the products to correct the storage conditions, 69-6349/R2, refers to…
Nov 25, 2003Not classifiedZ-0320-04OtherThe CA 125 assay product inserts list the incorrect storage instructions for the CA 125 Calibrators and Controls, and the product information letter (PI-P) shipped with the products to correct the storage conditions, 69-6349/R2, refers to…
Nov 25, 2003Not classifiedZ-0321-04OtherThe CA 125 assay product inserts list the incorrect storage instructions for the CA 125 Calibrators and Controls, and the product information letter (PI-P) shipped with the products to correct the storage conditions, 69-6349/R2, refers to…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K042731 (Fujirebio Diagnostics,Inc., 2004) has gone furthest, with 8 descendants of its own.

How does its category compare?

All LTK devices

LTK Test, Epithelial Ovarian Tumor-Associated Antigen (Ca125): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LTK

1in 2024

2007: 12009: 22011: 12015: 22020: 12022: 12023: 12024: 1
2007high 2 (2009) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LTK

Too few LTK 510(k)s in any year since 2007 to give a median.

Recalls per year · LTK

1in 2013

2007: 12010: 32013: 1
2007high 3 (2010) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LTK

4in 2025

2007: 42008: 92009: 312010: 122011: 52012: 332013: 142014: 302015: 112016: 42017: 152018: 82019: 152020: 52021: 22022: 22023: 62024: 62025: 4
2007high 33 (2012) · low 2 (2021)2025

MAUDE reports naming product code LTK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—14
20080—09
20092—031
20100—312
20111—05
20120—033
20130—114
20140—030
20152—011
20160—04
20170—015
20180—08
20190—015
20201—05
20210—02
20221—02
20231—06
20241—06
20250—04
20260—0—

What is it indicated for?

The AxSYM® CA 125™ assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative measurement of CA 125 assay values in human serum. The AxSYM® CA 125™ assay is to be used as an aid in monitoring response to therapy for patients with epithelial ovarian cancer. Serial testing for patient CA 125 assay values should be used in conjunction with other clinical methods used for monitoring ovarian cancer.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K964020: Axsym CA 125.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K964020

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026