510(k) Report

510(k) · Cleared

Recalled · Class II · 2016

Philips Computed Radiography

K964124510(k) TraditionalCleared Dec 18, 199664-day reviewClass II21 CFR 892.2050Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems, Inc. · LLZ System, Image Processing, Radiological

Cleared on Dec 18, 1996 as substantially equivalent to K915237 (Philips Medical Systems, Inc., 1992) and 1 other predicate after a 64-day review. Its family tree reaches back 1 generation to 1992. 18 later devices descend from it. It has been recalled 3 times.

Decided
Dec 18, 1996
Review
64 days
Class
Class II
Product code
LLZ
Recalls
3
Predicates
2

Its longest line of descent runs back 1 generation to K915237 (1992). 2 ancestors in all. 18 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor2010: 1 descendant cleared2007: 1 descendant cleared2014: 2 descendants cleared2015: 1 descendant cleared2021: 6 descendants cleared2016: 1 descendant cleared2019: 2 descendants cleared2017: 1 descendant cleared2020: 1 descendant cleared2023: 1 descendant cleared2024: 1 descendant clearedK915237 · Philips PCR-Ace System · Philips Medical Systems, Inc. · 19921990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 1 descendant3 generations forward: 4 descendants4 generations forward: 4 descendants5 generations forward: 4 descendants6 generations forward: 2 descendants7 generations forward: 1 descendant K915237 · Philips PCR-Ace System · Philips Medical Systems, Inc. · 19920+5
  1. Root · 1 generation backK915237 Philips PCR-Ace SystemPhilips Medical Systems, Inc. · 1992
  2. This deviceK964124 Philips Computed RadiographyPhilips Medical Systems, Inc. · 1996
  3. Descendant · cites it directly · 16 of its ownK063781 Philips XD-S Direct Radiography Workstation/PackagePhilips Medical Systems North America Co. · 2007 · Recalled
  4. Descendant · 2 generations on · 15 of its ownK140771 Philips Eleva WorkspotPhilips Medical Systems Dmc GmbH · 2014
  5. Descendant · 3 generations on · 6 of its ownK141736 Philips Prograde; Philips Eleva Workspot for Digital DiagnostPhilips Medical Systems Dmc GmbH · 2014 · Recalled
  6. Descendant · 3 generations on · 3 of its ownK203188 ADDH&abyz Co., Ltd. · 2021
  7. Descendant · 4 generations on · 3 of its ownK182973 DigitalDiagnost C90Philips Medical Systems Dmc GmbH · 2019

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2007: 12010: 12014: 22015: 12016: 12017: 12019: 22020: 12021: 62023: 12024: 1
2007high 6 (2021) · low 0 (2008)2026

Most influential descendants

Showing 5 of 18 · See all
  • K063781Philips XD-S Direct Radiography Workstation/PackagePhilips Medical Systems North America Co. · 2007 · cites it directly16descendants
  • K140771Philips Eleva WorkspotPhilips Medical Systems Dmc GmbH · 2014 · 2 generations on15descendants
  • K141736Philips Prograde; Philips Eleva Workspot for Digital DiagnostPhilips Medical Systems Dmc GmbH · 2014 · 3 generations on6descendants
  • K203188ADDH&abyz Co., Ltd. · 2021 · 3 generations on3descendants
  • K182973DigitalDiagnost C90Philips Medical Systems Dmc GmbH · 2019 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K915237Philips PCR-Ace SystemPhilips Medical Systems, Inc.1992—
0 (this device)K964124Philips Computed RadiographyPhilips Medical Systems, Inc.1996—
+1K063781Philips XD-S Direct Radiography Workstation/PackagePhilips Medical Systems North America Co.20071
+2K140771Philips Eleva WorkspotPhilips Medical Systems Dmc GmbH2014—
+3K141736Philips Prograde; Philips Eleva Workspot for Digital DiagnostPhilips Medical Systems Dmc GmbH20146
+3K203188ADDH&abyz Co., Ltd.2021—
+4K182973DigitalDiagnost C90Philips Medical Systems Dmc GmbH2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
201010
201420
201510
201610
201710
201920
202010
202160
202310
202410
FIG. 1 — Lineage of K964124, 1992–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K915237Philips PCR-Ace System
Summary page 2
…the technological characteristics of the PCR 5.2 system to those of the PCR ACe system. The PCR ACe system has been granted a 510(k) clearance. Refer to accession number K915237. Also, the image processing algorithm called Dynamic Range Reconstruction (DRR) used with PCR 5.2 is considered substantially equ…
Philips Medical Systems, Inc.1992—
*K951373Fuji Computed Radiography System FCR9000HQ, FCR9501HQ
Summary page 1
…number K896050A. The AC3 has been granted a 510(k) clearance. Refer to accession number K944046. The 9000 has been granted a 510(k) clearance. Refer to accession number K951373. Philips Medical Systems North America Company 710 Bridgeport Avenue Shelton, CT 06484-4708 Tel: (203) 926-7674 Fax: {203) 929-…
Fujifilm Medical System U.S.A., Inc.1995—

What descends from it? (18)

See all 18

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K063781Philips XD-S Direct Radiography Workstation/PackagePhilips Medical Systems North America Co.200716
K093355PCR Eleva 1.2Philips Medical Systems, Inc.20100

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 12, 2016Class IIZ-2681-2016Component design/selectionA defect component of the Compano reader power supply can cause the power supply to burn and some smoke to leave the housing.
Jul 27, 2011Not classifiedZ-3205-2011Software designThe mirror icon on the bottom of each image could be interpreted as a RIGHT marker for side indication if the image is mirrored again within a PACS system. This could potentially lead to a misdiagnosis
Dec 23, 2008Not classifiedZ-1019-2011Software designThere is a risk of selecting the wrong patient, when patient is selected in the patient list and the examination tab pressed with background query are running at the same time.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K063781 (Philips Medical Systems North America Co., 2007) has gone furthest, with 16 descendants of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

The PCR 5.2 System is a digital film processing system for reading and then digitizing X-ray images from reusable imaging plates which have been exposed in conventional radiographic examination devices. The digitized X-ray images can then be viewed, stored, post-processed and printed. The Philips PCR system can be used in all conventional RAD/RF examination situations, except for mammography. PCR is suitable for routine RAD exams as well as specialist areas, like intensive care units, trauma departments and pediatric departments.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K964124: Philips Computed Radiography.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K964124

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026