510(k) Report

510(k) · Cleared

Recalled · 2006

Stainless Steel Greenfield Vena Cava Filter with 12F Introducer System

K964284510(k) TraditionalCleared Jan 6, 199770-day reviewClass II21 CFR 870.3375Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corp · DTK Filter, Intravascular, Cardiovascular

Cleared on Jan 6, 1997 as substantially equivalent to K912035 (Boston Scientific Corp, 1995) after a 70-day review. Its family tree reaches back 1 generation to 1995. 7 later devices descend from it. It has been recalled 3 times.

Decided
Jan 6, 1997
Review
70 days
Class
Class II
Product code
DTK
Recalls
3
Predicates
1

Its longest line of descent runs back 1 generation to K912035 (1995). 7 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2009: 1 descendant cleared2013: 2 descendants cleared2012: 1 descendant cleared2024: 1 descendant cleared2015: 2 descendants clearedK912035 · Stainless Steel Green Vena Cava Filt W/12F Intr SY · Boston Scientific Corp · 19951995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 3 descendants4 generations forward: 1 descendant K912035 · Stainless Steel Green Vena Cava Filt W/12F Intr SY · Boston Scientific Corp · 19950
  1. Root · 1 generation backK912035 Stainless Steel Green Vena Cava Filt W/12F Intr SYBoston Scientific Corp · 1995
  2. This deviceK964284 Stainless Steel Greenfield Vena Cava Filter with 12F Introducer SystemBoston Scientific Corp · 1997
  3. Descendant · cites it directly · 6 of its ownK081410 Option Vena Cava FilterRex Medical · 2009
  4. Descendant · 2 generations on · 2 of its ownK133243 Option Elite Vena Cava Filter SystemRex Medical, L.P. · 2013 · Recalled
  5. Descendant · 2 generations on · 2 of its ownK120402 Crux Vena Cava Filter SystemCrux Biomedical · 2012
  6. Descendant · 3 generations on · 1 of its ownK143405 Option (TM) ELITE Vena Cava Filter 100cm SystemRex Medical, L.P. · 2015
  7. Descendant · 3 generations on · 1 of its ownK122585 Crux Vena Cava Filter SystemCrux Biomedical · 2013

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2009: 12012: 12013: 22015: 22024: 1
2009high 2 (2013) · low 0 (2010)2026

Most influential descendants

Showing 5 of 7 · See all
  • K081410Option Vena Cava FilterRex Medical · 2009 · cites it directly6descendants
  • K133243Option Elite Vena Cava Filter SystemRex Medical, L.P. · 2013 · 2 generations on2descendants
  • K120402Crux Vena Cava Filter SystemCrux Biomedical · 2012 · 2 generations on2descendants
  • K143405Option (TM) ELITE Vena Cava Filter 100cm SystemRex Medical, L.P. · 2015 · 3 generations on1descendant
  • K122585Crux Vena Cava Filter SystemCrux Biomedical · 2013 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K912035Stainless Steel Green Vena Cava Filt W/12F Intr SYBoston Scientific Corp1995—
0 (this device)K964284Stainless Steel Greenfield Vena Cava Filter with 12F Introducer SystemBoston Scientific Corp1997—
+1K081410Option Vena Cava FilterRex Medical2009—
+2K133243Option Elite Vena Cava Filter SystemRex Medical, L.P.20132
+2K120402Crux Vena Cava Filter SystemCrux Biomedical2012—
+3K143405Option (TM) ELITE Vena Cava Filter 100cm SystemRex Medical, L.P.2015—
+3K122585Crux Vena Cava Filter SystemCrux Biomedical2013—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201210
201320
201520
202410
FIG. 1 — Lineage of K964284, 1995–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTK device

  1. #1 K242612 Option ELITE Vena Cava Filter System…score 10.1 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K912035Stainless Steel Green Vena Cava Filt W/12F Intr SY
Our OCR of the scanned summary read K912038, FDA’s number for Dash Medical Gloves (Safehealth Labs). Corrected to the number one digit away whose FDA name carries the summary’s words for the device, “Stainless Steel Greenfield® Vena Cava Filter with 12F Introducer…”.
Boston Scientific Corp1995—

Also cleared against K912035: 2 other devices →

What descends from it? (7)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081410Option Vena Cava FilterRex Medical20096

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 6, 2006Not classifiedZ-0794-2007OtherOne lot of product, Stainless Steel Greenfield Vena Cava Filter, may have the incorrect expiration date of 2008-07 and the correct date is 2008-06.
Dec 2, 2005Not classifiedZ-0280-06OtherThere have been reports of detachment at the bond between the carrier capsule and the outer sheath of the Greenfield Vena Cava Filters with 12 Fr Femoral introducer Systems manufactured before Marcyh 10, 2004. If the capsule should detach…
Aug 5, 2005Not classifiedZ-0299-06OtherCertain units within eleven lots of the 12 Fr Greenfield Vena Cava Filter may lack a taper on the braided sheath of the delivery system. The lack of a taper could lead to vessel damage as a result of the sheath catching on the edge of the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K081410 (Rex Medical, 2009) has gone furthest, with 6 descendants of its own.
  • 2 other devices also cite its predicate K912035.

How does its category compare?

All DTK devices

DTK Filter, Intravascular, Cardiovascular: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DTK

5in 2024

2007: 32008: 62009: 32010: 32011: 22012: 52013: 32014: 22015: 22016: 32017: 52018: 22021: 22024: 5
2007high 6 (2008) · low 0 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTK

29in 2024

2007: 87 days median2008: 87 days median2009: 380 days median2010: 50 days median2012: 156 days median2013: 91 days median2016: 84 days median2017: 88 days median2024: 29 days median
2007high 380 (2009) · low 29 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTK

1in 2025

2007: 12013: 12015: 12016: 22019: 62022: 22025: 1
2007high 6 (2019) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DTK

432in 2025

2007: 3642008: 3452009: 4562010: 3692011: 4392012: 3512013: 4732014: 4512015: 8592016: 2,5742017: 4,3642018: 3,9382019: 6,4672020: 5,4052021: 5,3572022: 2,1912023: 5832024: 4062025: 432
2007high 6,467 (2019) · low 345 (2008)2025

MAUDE reports naming product code DTK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073871364
20086870345
200933800456
20103500369
20112—0439
201251560351
20133911473
20142—0451
20152—1859
201638422,574
201758804,364
20182—03,938
20190—66,467
20200—05,405
20212—05,357
20220—22,191
20230—0583
20245290406
20250—1432
20260—0—

What is it indicated for?

The Stainless Steel Greenfield® Vena Cava Filter with 12F Introducer System is indicated for the prevention of pulmonary embolism via placement in the vena cava in the following situations:

Venous thrombosis or pulmonary thromboembolism when anticoagulants are contraindicated or inadequate for management of venous thrombosis with significant risk of, or following, pulmonary thromboembolism.

Failure of anticoagulant therapy in thromboembolic disease.

Emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced.

Chronic, recurrent pulmonary embolism where anticoagulant therapy has failed or is contraindicated.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K964284: Stainless Steel Greenfield Vena Cava Filter with 12F Introducer System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K964284

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026