510(k) Report

510(k) · Cleared

Cleared

Asahi AM-R Series Dialyzers

K970650510(k) TraditionalCleared Sep 30, 1997222-day reviewClass II21 CFR 876.5820

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Asahi Medical Co., Ltd. · MSE Hemodialyzer, Re-Use, Low Flux

Cleared on Sep 30, 1997 as substantially equivalent to K892374 (Asahi Medical Co., Ltd., 1989) and 1 other predicate after a 222-day review. Its family tree reaches back 1 generation to 1989. 1 later device descends from it.

Decided
Sep 30, 1997
Review
222 days
Class
Class II
Product code
MSE
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K892374 (1989). 2 ancestors in all. 1 later device descends from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1989: 1 other ancestor1999: 1 descendant clearedK892374 · New AM-Series Dialyzers · Asahi Medical Co., Ltd. · 19891990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant K892374 · New AM-Series Dialyzers · Asahi Medical Co., Ltd. · 19890
  1. Root · 1 generation backK892374 New AM-Series DialyzersAsahi Medical Co., Ltd. · 1989
  2. This deviceK970650 Asahi AM-R Series DialyzersAsahi Medical Co., Ltd. · 1997
  3. Descendant · cites it directly · 0 of its ownK991512 Asahi AM-R-Series Dialyzers (Modified), Model AM-NR-XAsahi Medical Co., Ltd. · 1999 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K991512Asahi AM-R-Series Dialyzers (Modified), Model AM-NR-XAsahi Medical Co., Ltd. · 1999 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K892374New AM-Series DialyzersAsahi Medical Co., Ltd.1989—
0 (this device)K970650Asahi AM-R Series DialyzersAsahi Medical Co., Ltd.1997—
+1K991512Asahi AM-R-Series Dialyzers (Modified), Model AM-NR-XAsahi Medical Co., Ltd.19991

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
FIG. 1 — Lineage of K970650, 1989–1999Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MSE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K892374New AM-Series Dialyzers
Summary page 3
…s of fabrication, and manufacturing of the Asahi AM-R Series Dialyzers remain unchanged from the Asahi AM-Series Dialyzers cleared by FDA for marketing in the U.S. under K892374 and K892375. Therefore, from the perspective of technological characteristics, the AM-R Series Dialyzers under this 510(k) are id…
Asahi Medical Co., Ltd.1989—
*K892375Modified AM-Series Dialyzers
Summary page 3
…tion, and manufacturing of the Asahi AM-R Series Dialyzers remain unchanged from the Asahi AM-Series Dialyzers cleared by FDA for marketing in the U.S. under K892374 and K892375. Therefore, from the perspective of technological characteristics, the AM-R Series Dialyzers under this 510(k) are identical to t…
Asahi Medical Co., Ltd.1989—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K991512Asahi AM-R-Series Dialyzers (Modified), Model AM-NR-XAsahi Medical Co., Ltd.19990

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K991512 (Asahi Medical Co., Ltd., 1999) has gone furthest, with 0 descendants of its own.

How does its category compare?

All MSE devices

MSE Hemodialyzer, Re-Use, Low Flux: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MSE

FDA has decided no MSE device since 2007.

Median 510(k) review, days · MSE

Too few MSE 510(k)s in any year since 2007 to give a median.

Recalls per year · MSE

No recall names MSE since 2007.

Adverse event reports per year · MSE

5in 2014

2010: 62011: 52014: 5
2007high 6 (2010) · low 0 (2007)2025

MAUDE reports naming product code MSE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—06
20110—05
20120—00
20130—00
20140—05
20150—00
20160—00
20170—00
20180—00
20190—00
20200—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

Asahi AM-R Series Dialyzers are intended for use for hemodialysis treatment of patients who have chronic renal failure or acute renal failure. Asahi AM-R Series Dialyzers have been tested in vitro and in confirmatory clinical studies under reprocessing and reuse conditions for up to 15 reuse cycles. Based on the results from these evaluations, Asahi AM-R Series Dialyzers may be reprocessed for reuse on the same patient. If reprocessing and reuse is…

Read the full statement (293 words)

Asahi AM-R Series Dialyzers are intended for use for hemodialysis treatment of patients who have chronic renal failure or acute renal failure.

Asahi AM-R Series Dialyzers have been tested in vitro and in confirmatory clinical studies under reprocessing and reuse conditions for up to 15 reuse cycles. Based on the results from these evaluations, Asahi AM-R Series Dialyzers may be reprocessed for reuse on the same patient. If reprocessing and reuse is practiced, it is recommended that the reuse be done under the conditions as existed in the in vitro and confirmatory clinical studies as recommended immediately below. It is noted that the Asahi AM-R Series Dialyzers have not been tested for reuse when reprocessed with agents and/or processes other than these, and the performance of the dialyzers under other conditions are not known and cannot be recommended. Accordingly:

(1) The reprocessed dialyzer may be used only if the residual Total Cell Volume (TCV) is at least 80% of the original TCV and if such dialyzer otherwise meets the acceptance criteria of the instructions for use and the instructions of the reprocessing system utilized. Furthermore, the policies, instructions and criteria of the institution for reuse (e.g., concerning dialyzer performance, residual blood, and/or dialyzer leakage or damage) should be followed.

(2) The reprocessing agent may be either (1) 4% formaldehyde (also known as formalin) in conjunction with the Seratronics Reprocessing Systems for Dialyzer Reprocessing and Preparation (DRS4™ and DPS4™) manufactured by Seratronics, Inc., or (2) Renalin® in conjunction with the Renatron® Dialyzer Reprocessing System, manufactured by Renal Systems, Inc.

(3) The instructions provided by the manufacturer of the chosen reprocessing agent must be followed in reprocessing the dialyzer.

(4) The reprocessed dialyzer may be used only on dialysis systems equipped with volumetric ultrafiltration controllers.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K970650: Asahi AM-R Series Dialyzers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K970650

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026