510(k) Report

510(k) · Cleared

Cleared

Vapocide

K970726510(k) TraditionalCleared May 18, 1998444-day reviewClass II21 CFR 880.6885

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cottrell, Ltd. · MED Sterilant, Medical Devices

Cleared on May 18, 1998 after a 444-day review. Our reading found no predicate number; beside “510(k)” the summary reads “2127197”, which our checks could not confirm as its predicate.

Decided
May 18, 1998
Review
444 days
Class
Class II
Product code
MED
Recalls
None
Predicates
0

Our reading found no predicate number; beside “510(k)” the summary reads “2127197”, which our checks could not confirm as its predicate.

FIG. 1 — Lineage of K970726Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MED? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MED device

  1. #1 K251666 SYSTEM 1E Liquid Chemical Sterilant Processing…score 13.8 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Vapo-Steril
Named without a submission number
—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All MED devices

MED Sterilant, Medical Devices: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MED

2in 2023

2007: 22008: 12010: 62012: 32013: 12014: 12016: 12017: 12018: 22019: 42021: 22022: 12023: 22026 (partial year): 2
2007high 6 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MED

84in 2019

2010: 281 days median2012: 111 days median2019: 84 days median
2007high 281 (2010) · low 84 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MED

2in 2011

2007: 32008: 22011: 2
2007high 3 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MED

9in 2025

2007: 342008: 792009: 1582010: 892011: 1062012: 3372013: 1402014: 1052015: 1102016: 1352017: 742018: 422019: 352020: 322021: 312022: 162023: 92024: 82025: 9
2007high 337 (2012) · low 8 (2024)2025

MAUDE reports naming product code MED, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—334
20081—279
20090—0158
20106281089
20110—2106
201231110337
20131—0140
20141—0105
20150—0110
20161—0135
20171—074
20182—042
2019484035
20200—032
20212—031
20221—016
20232—09
20240—08
20250—09
20262—0—

What is it indicated for?

Vapocide® is a sterilant solution intended to be used only in the Harvey Chemiclave Sterilizer to sterilize medical and dental surgical instruments and equipment. Vapocide is intended to be a direct substitute for Vapo-Steril. Vapocide is recommended for use in the following Chemiclave models: Model E, 4000, 5000, 5500, 6000, 7000, 8000, and EM series sterilizers. Vapocide is compatible with Chemiclave cycle parameters as follows: Model E, 4000, 5000…

Read the full statement (228 words)

Vapocide® is a sterilant solution intended to be used only in the Harvey Chemiclave Sterilizer to sterilize medical and dental surgical instruments and equipment. Vapocide is intended to be a direct substitute for Vapo-Steril. Vapocide is recommended for use in the following Chemiclave models: Model E, 4000, 5000, 5500, 6000, 7000, 8000, and EM series sterilizers.

Vapocide is compatible with Chemiclave cycle parameters as follows:

Model E, 4000, 5000, 5500, 6000, EM 5000, EM 5500, EM 6000

• Operating pressure: 20-40 psi (1.41-2.82 kg/cm²)

• Operating Temperature: 132±2°C (270±5°F)

• Maximum load: 3.3 pounds (1.5 kg) for models with 9 and 10 in. diameter chambers

Model 7000 and 8000

• Operating pressure: approximately 25 psi

• Operating Temperature: 132 + 6, -0°C

• Maximum load: 3.3 pounds (1.5 kg)

Vapocide is compatible with but not limited to the following instruments and material types:

• Surgical and cutting instruments, handpieces, and similar items

• Plastic materials recommended are: Polysulfone, Phenolic, Epoxy, Polyphenylene oxide, Polyester, Polypropylene

Note: Polycarbonate is not chemically compatible

Vapocide™ is not compatible with the following materials:

• Linen, textiles or fabrics

• Liquids and agars

• Items contained in tightly woven packs or wraps

• Plastics that cannot tolerate temperature in excess of 132°C (270°F)

• Nylon tubing & bags - inhibit penetration of chemical vapors

Vapocide is not recommended for use for “flash” sterilization cycles.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K970726: Vapocide.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K970726

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026