510(k) Report

510(k) · Cleared

Cleared

Intelliflow Respiratory Analyzer

K970840510(k) TraditionalCleared Nov 26, 1997264-day reviewClass II21 CFR 868.1880

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Simtec, Inc. · BZC Calculator, Pulmonary Function Data

Cleared on Nov 26, 1997 after a 264-day review. FDA’s file for it is a 510(k) statement, which names no predicate.

Decided
Nov 26, 1997
Review
264 days
Class
Class II
Product code
BZC
Recalls
None
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K970840Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in BZC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new BZC device

  1. #1 K240293 STSscore 9.8 · 7 generations · no Class I recall
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All BZC devices

BZC Calculator, Pulmonary Function Data: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · BZC

1in 2024

2007: 12011: 12014: 22015: 12016: 12017: 12018: 12019: 22024: 1
2007high 2 (2014) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · BZC

Too few BZC 510(k)s in any year since 2007 to give a median.

Recalls per year · BZC

3in 2009

2009: 3
2007high 3 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · BZC

3in 2025

2007: 12008: 32016: 42017: 12018: 12019: 72020: 32021: 52022: 92023: 82024: 82025: 3
2007high 9 (2022) · low 0 (2009)2025

MAUDE reports naming product code BZC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—01
20080—03
20090—30
20100—00
20111—00
20120—00
20130—00
20142—00
20151—00
20161—04
20171—01
20181—01
20192—07
20200—03
20210—05
20220—09
20230—08
20241—08
20250—03
20260—0—

What is it indicated for?

The SIMTEC Intelliflow Respiratory Analyzer is intended to acquire and interpret pulmonary function data and for CO_ analysis. The Intelliflow unit was tested at LDS Hospital to verify that it conforms to the 24 wave forms recommended by the American Thoracic Society Standard of Spirometry, Am. Rev. Respir. Dis. 1987;136, 1285-1298. The Intelliflow does not meet the ATS guidelines for accuracy. The following five wave forms have not been qualified for…

Read the full statement (131 words)

The SIMTEC Intelliflow Respiratory Analyzer is intended to acquire and interpret pulmonary function data and for CO_ analysis. The Intelliflow unit was tested at LDS Hospital to verify that it conforms to the 24 wave forms recommended by the American Thoracic Society Standard of Spirometry, Am. Rev. Respir. Dis. 1987;136, 1285-1298. The Intelliflow does not meet the ATS guidelines for accuracy. The following five wave forms have not been qualified for accuracy according to the LFS simulator: #3FVC, #15MMEF, #21FVC, #23FVC, #24FVC. A copy of the actual results is enclosed with the operator's manual. Please contact SIMTEC for an explanation of the differences in the LDS simulator results and the method used by SIMTEC to verify the accuracy of all 24 wave forms. This device is not intended for patient monitoring.

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K970840: Intelliflow Respiratory Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K970840

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026