510(k) Report

510(k) · Cleared

Recalled · Class I · 2023

5 FR and 6 FR Infiniti Angiographic Catheter

K970854510(k) TraditionalCleared Sep 30, 1997204-day reviewClass II21 CFR 870.1200Recalls 3 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cordis Corp. · DQO Catheter, Intravascular, Diagnostic

Cleared on Sep 30, 1997 as substantially equivalent to K960975 (Cordis Corp., 1996) and 1 other predicate after a 204-day review. Its family tree reaches back 2 generations to 1992. 1 of the 11 devices in its family, counting itself, has a Class I recall. 7 later devices descend from it. It has been recalled 3 times, including a Class I recall.

Decided
Sep 30, 1997
Review
204 days
Class
Class II
Product code
DQO
Recalls
3 (1 Class I)
Predicates
2

Its longest line of descent runs back 2 generations to K921310 (1992). 3 ancestors in all. 7 later devices descend from it, 4 citing it directly. 1 of the 11 devices in this family, counting this one, has a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1993: 1 other ancestor2010: 1 descendant cleared2023: 1 descendant cleared2005: 1 descendant cleared2011: 1 descendant cleared2008: 1 descendant cleared2013: 1 descendant cleared2009: 1 descendant clearedK921310 · Cordis 6 French Paragon Angiographic Catheter · Cordis Corp. · 1992K960975 · 1-4F Infiniti Angiographic Catheter · Cordis Corp. · 1996 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant K960975 · 1-4F Infiniti Angiographic Catheter · Cordis Corp. · 1996 · recalledK921310 · Cordis 6 French Paragon Angiographic Catheter · Cordis Corp. · 19920
  1. Root · 2 generations backK921310 Cordis 6 French Paragon Angiographic CatheterCordis Corp. · 1992
  2. 1 generation backK960975 1-4F Infiniti Angiographic CatheterCordis Corp. · 1996 · Recalled
  3. This deviceK970854 5 FR and 6 FR Infiniti Angiographic CatheterCordis Corp. · 1997
  4. Descendant · cites it directly · 2 of its ownK051561 Asahi Angiographic CatheterAsahi Intecc Co., Ltd. · 2005
  5. Descendant · cites it directly · 1 of its ownK090427 Ursa Angiographic CatheterCurative Medical Technology, Inc. · 2010
  6. Descendant · 2 generations on · 1 of its ownK082559 Spectranetics Cross-Pilot Turbo Elite Support CatheterSpectranetics Corp. · 2008
  7. Descendant · cites it directly · 0 of its ownK232573 INFINITI™ Ambi Angiographic CatheterCordis US Corp · 2023
  8. Descendant · 2 generations on · 0 of its ownK123818 Ursa Angiographic CatheterCurative Medical Technology, Inc. · 2013

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2005: 12008: 12009: 12010: 12011: 12013: 12023: 1
2005high 1 (2005) · low 0 (2006)2026

Most influential descendants

Showing 5 of 7 · See all
  • K051561Asahi Angiographic CatheterAsahi Intecc Co., Ltd. · 2005 · cites it directly2descendants
  • K090427Ursa Angiographic CatheterCurative Medical Technology, Inc. · 2010 · cites it directly1descendant
  • K082559Spectranetics Cross-Pilot Turbo Elite Support CatheterSpectranetics Corp. · 2008 · 2 generations on1descendant
  • K232573INFINITI™ Ambi Angiographic CatheterCordis US Corp · 2023 · cites it directly0descendants
  • K123818Ursa Angiographic CatheterCurative Medical Technology, Inc. · 2013 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K921310Cordis 6 French Paragon Angiographic CatheterCordis Corp.1992—
−1K9609751-4F Infiniti Angiographic CatheterCordis Corp.19961
0 (this device)K9708545 FR and 6 FR Infiniti Angiographic CatheterCordis Corp.1997—
+1K051561Asahi Angiographic CatheterAsahi Intecc Co., Ltd.2005—
+1K090427Ursa Angiographic CatheterCurative Medical Technology, Inc.2010—
+2K082559Spectranetics Cross-Pilot Turbo Elite Support CatheterSpectranetics Corp.2008—
+1K232573INFINITI™ Ambi Angiographic CatheterCordis US Corp2023—
+2K123818Ursa Angiographic CatheterCurative Medical Technology, Inc.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200510
200810
200910
201010
201110
201310
202310
FIG. 1 — Lineage of K970854, 1992–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQO device

  1. #1 K254278 Arterial Pressure Monitoring Set/Trayscore 14.4 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K9609751-4F Infiniti Angiographic Catheter
Summary page 1
…Proprietary Name: Cordis 5 F Infiniti® Angiographic Catheter Cordis 6 F Infiniti® Angiographic Catheter Name of Predicate Devices 4 F Infiniti® Angiographic Catheter, K960975, April 02, 1996 | 6 F Paragon (Infiniti® ) Angiographic Catheter, K921310/A, Sept. 17, 1992 Classification Class Il Performance…
Cordis Corp.19961
*K921310Cordis 6 French Paragon Angiographic Catheter
Summary page 1
…nfiniti® Angiographic Catheter Name of Predicate Devices 4 F Infiniti® Angiographic Catheter, K960975, April 02, 1996 | 6 F Paragon (Infiniti® ) Angiographic Catheter, K921310/A, Sept. 17, 1992 Classification Class Il Performance Standards Performance standards have not been established by the FDA und…
Cordis Corp.1992—

Also cleared against K960975: 4 other devices →

What descends from it? (7)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K051561Asahi Angiographic CatheterAsahi Intecc Co., Ltd.20052
K090427Ursa Angiographic CatheterCurative Medical Technology, Inc.20101
K232573INFINITI™ Ambi Angiographic CatheterCordis US Corp20230
K102633Bioteq Angiographic CatheterBioteque Corp.20110

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 26, 2023Class IZ-0286-2024Process controlNon-sterile product labeled as sterile was distributed.
Nov 7, 2022Class IIZ-0720-2023Under Investigation by firmThere is a potential for separation at the male connector.
Jun 26, 2006Not classifiedZ-1240-06OtherThe tip of the catheter may separate from the body during operation.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 1 company; a new generation was cleared every 3.5 years on median.
  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K051561 (Asahi Intecc Co., Ltd., 2005) has gone furthest, with 2 descendants of its own.
  • 4 other devices also cite its predicate K960975.

How does its category compare?

All DQO devices

DQO Catheter, Intravascular, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DQO

2in 2025

2007: 82008: 32009: 82010: 72011: 62012: 152013: 22014: 32015: 102016: 42017: 102018: 102019: 62020: 32021: 12022: 42023: 42024: 42025: 22026 (partial year): 3
2007high 15 (2012) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQO

148in 2024

2007: 106 days median2008: 80 days median2009: 58 days median2010: 137 days median2011: 74 days median2012: 94 days median2014: 109 days median2015: 209 days median2016: 89 days median2017: 176 days median2018: 234 days median2019: 113 days median2020: 30 days median2022: 181 days median2023: 73 days median2024: 148 days median2026 (partial year): 85 days median
2007high 234 (2018) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQO

1in 2023

2007: 142008: 42009: 72010: 112011: 72012: 22013: 12015: 1, 1 Class I2016: 2, 1 Class I2017: 12018: 22019: 712020: 4, 4 Class I2022: 12023: 1, 1 Class I2026 (partial year): 24, 4 Class I
2007high 71 (2019) · low 0 (2014)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · DQO

656in 2025

2007: 5642008: 7262009: 9332010: 5562011: 6252012: 5322013: 7532014: 6012015: 6282016: 6142017: 8732018: 7772019: 3862020: 3122021: 3382022: 3792023: 3912024: 5112025: 656
2007high 933 (2009) · low 312 (2020)2025

MAUDE reports naming product code DQO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007810614564
20083804726
20098587933
2010713711556
20116747625
201215942532
20132—1753
201431090601
2015102091 (1)628
20164892 (1)614
2017101761873
2018102342777
2019611371386
20203304 (4)312
20211—0338
202241811379
20234731 (1)391
202441480511
20252—0656
202638524 (4)—

What is it indicated for?

Cordis Angiographic Catheters are designed to deliver radiopaque contrast medium to

selected sites in the vascular system.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K970854: 5 FR and 6 FR Infiniti Angiographic Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K970854

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026