Zoll Medical Corp · MKJ Automated External Defibrillators (Non-Wearable)
Cleared on Mar 16, 1998 after a 273-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled once.
- Decided
- Mar 16, 1998
- Review
- 273 days
- Class
- Class III
- Product code
- MKJ
- Recalls
- 1
- Predicates
- 0
FDA’s file for it is a 510(k) statement, which names no predicate.
Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MKJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
- Candidate predicates for a device like this, rankedin the report
- Recall history across each candidate’s familyin the report
What was it cleared against?
FDA’s file for it is a 510(k) statement, which names no predicate.
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 23, 2003 | Not classified | Z-1000-03 | Other | Defibrillator may fail to charge or hold its charge above energies of 75 joules |
No ancestor or descendant has a Class I recall.
How does its category compare?
All MKJ devicesMKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MKJ
48in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MKJ
156in 2015
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MKJ
1in 2025
Class I (23 since 2007) Class II/III not classified
Adverse event reports per year · MKJ
13,315in 2025
MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 69 | 12 | 3,958 |
| 2008 | 13 | 141 | 24 | 4,752 |
| 2009 | 8 | 165 | 37 | 6,782 |
| 2010 | 4 | 250 | 35 | 7,399 |
| 2011 | 11 | 36 | 7 | 8,170 |
| 2012 | 10 | 197 | 5 (1) | 9,732 |
| 2013 | 7 | 191 | 4 | 14,955 |
| 2014 | 11 | 196 | 24 (15) | 15,399 |
| 2015 | 6 | 156 | 1 | 16,129 |
| 2016 | 0 | — | 2 (1) | 17,062 |
| 2017 | 4 | — | 6 (4) | 16,684 |
| 2018 | 20 | — | 3 | 16,591 |
| 2019 | 25 | — | 6 (2) | 16,861 |
| 2020 | 52 | — | 2 | 16,197 |
| 2021 | 41 | — | 4 | 16,300 |
| 2022 | 19 | — | 3 | 14,867 |
| 2023 | 24 | — | 11 | 13,766 |
| 2024 | 27 | — | 5 | 15,109 |
| 2025 | 48 | — | 1 | 13,315 |
| 2026 | 31 | — | 3 | — |
What is it indicated for?
Defibrillator Function Indications for Use — Manual Operation Use of the ZOLL REVIVER in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: » Unconsciousness * Absence of breathing + Absence of pulse. The ZOLL REVIVER contains a standard DC defibrillator capable of delivering up to 360 joules of energy. In manual mode, it may also be used for synchronized cardioversion…
Read the full statement (541 words)
Defibrillator Function
Indications for Use — Manual Operation
Use of the ZOLL REVIVER in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
» Unconsciousness
* Absence of breathing
+ Absence of pulse.
The ZOLL REVIVER contains a standard DC defibrillator capable of delivering up to 360 joules of energy. In manual mode, it may also be used for synchronized cardioversion by using the R-wave of the patient’s ECG as a timing reference.
This product is to be used only by qualified medical personnel for the purposes of converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
Indications for Use — Advisory Function
The advisory function should only be used to confirm ventricular fibrillation and ventricular tachycardia (greater than 150 beats per minute) in patients meeting the following clinical criteria:
*- Unconsciousness
*+ Absence of breathing
< Absence of pulse.
Indications for Use — Semiautomatic Operation (if equipped)
The ZOLL REVIVER is designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the operator of the device controls the delivery of the shock to the patient. It is specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated in a medically-approved patient care protocol.
The ZOLL REVIVER must be prescribed for use by a physician or medical advisor of an emergency response team.
Use of the ZOLL REVIVER in the semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
* Unconsciousness
« Absence of breathing
« Absence of pulse.
Pacemaker Function (if equipped)
Indications for Use — Pacemaker
This product may be used for cardiac pacing in conscious or unconscious patients for up to a few hours duration as an alternative to endocardial stimulation. The purposes of pacing include:
1 Resuscitation from standstill or bradycardia of any etiology:
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidine, digitalis, B-blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). it has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.
2 As a standby when standstill or bradycardia might be expected:
Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker imptants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous therapy which avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
3 Suppression of tachycardia:
An increase in heart rate from external pacing often suppresses ventricular ectopic activity and may prevent tachycardia.
Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K972241 · printed Nov 6, 2026
Cite this page
510(k) Report. “K972241: Zoll Reviver.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K972241
Report an extraction error