510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Cardiocall Cardiac Event Recorder

K972649510(k) TraditionalCleared Dec 4, 1997142-day reviewClass II21 CFR 870.2920Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Reynolds Medical, Ltd. · DXH Transmitters And Receivers, Electrocardiograph, Telephone

Cleared on Dec 4, 1997 as substantially equivalent to K923930 (Tz Medical, Inc., 1992) after a 142-day review. Its family tree reaches back 1 generation to 1992. 19 later devices descend from it. It has been recalled once.

Decided
Dec 4, 1997
Review
142 days
Class
Class II
Product code
DXH
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K923930 (1992). 19 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2007: 1 descendant cleared1999: 1 descendant cleared2009: 1 descendant cleared2008: 1 descendant cleared2012: 1 descendant cleared2013: 1 descendant cleared2015: 2 descendants cleared2018: 3 descendants cleared2016: 1 descendant cleared2022: 2 descendants cleared2020: 1 descendant cleared2023: 2 descendants cleared2021: 1 descendant cleared2025: 1 descendant clearedK923930 · Electrocardiograph Transmitter · Tz Medical, Inc. · 19921990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 3 descendants5 generations forward: 4 descendants6 generations forward: 3 descendants7 generations forward: 2 descendants8 generations forward: 1 descendant K923930 · Electrocardiograph Transmitter · Tz Medical, Inc. · 19920+5
  1. Root · 1 generation backK923930 Electrocardiograph TransmitterTz Medical, Inc. · 1992
  2. This deviceK972649 Cardiocall Cardiac Event RecorderReynolds Medical, Ltd. · 1997
  3. Descendant · cites it directly · 17 of its ownK063609 River- 1 Electrocardiograph (Ecg) Recorder and TransmitterShl Telemedicine International, Ltd. · 2007
  4. Descendant · 2 generations on · 11 of its ownK080047 Cardiosen'cShl Telemedicine International, Ltd. · 2008
  5. Descendant · 3 generations on · 10 of its ownK113514 SmartheartShl Telemedicine International, Ltd. · 2012
  6. Descendant · 4 generations on · 6 of its ownK142476 Master Caution Device MCDHealthwatch, Ltd. · 2015
  7. Descendant · 5 generations on · 5 of its ownK161431 SimplECGNanowear, Inc. · 2016

Showing 5 of 19 descendants. See all descendants

Descendants cleared per year

1999: 12007: 12008: 12009: 12012: 12013: 12015: 22016: 12018: 32020: 12021: 12022: 22023: 22025: 1
1999high 3 (2018) · low 0 (2000)2026

Most influential descendants

Showing 5 of 19 · See all
  • K063609River- 1 Electrocardiograph (Ecg) Recorder and TransmitterShl Telemedicine International, Ltd. · 2007 · cites it directly17descendants
  • K080047Cardiosen'cShl Telemedicine International, Ltd. · 2008 · 2 generations on11descendants
  • K113514SmartheartShl Telemedicine International, Ltd. · 2012 · 3 generations on10descendants
  • K142476Master Caution Device MCDHealthwatch, Ltd. · 2015 · 4 generations on6descendants
  • K161431SimplECGNanowear, Inc. · 2016 · 5 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K923930Electrocardiograph TransmitterTz Medical, Inc.1992—
0 (this device)K972649Cardiocall Cardiac Event RecorderReynolds Medical, Ltd.1997—
+1K063609River- 1 Electrocardiograph (Ecg) Recorder and TransmitterShl Telemedicine International, Ltd.2007—
+2K080047Cardiosen'cShl Telemedicine International, Ltd.2008—
+3K113514SmartheartShl Telemedicine International, Ltd.2012—
+4K142476Master Caution Device MCDHealthwatch, Ltd.2015—
+5K161431SimplECGNanowear, Inc.2016—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
200710
200810
200910
201210
201310
201520
201610
201830
202010
202110
202220
202320
202510
FIG. 1 — Lineage of K972649, 1992–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DXH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DXH device

  1. #1 K243438 MEMO Patch M (MPT-E08R-UNC01)score 12.9 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K923930Electrocardiograph Transmitter
Summary page 1
…tters and Receivers, Electrocardiograph, Telephone (74DXH) (3) Equivalent legally-marketed devices: 1. Ralin Heart-Aide Event recorder 2. Braemar ER700 Event Recorder, K923930 (4) Description The CardioCall Event Recorder offers the user the choice of a looping cardiac event recorder (records the elect…
Tz Medical, Inc.1992—
*—Ralin Heart-Aide Event recorder
Named without a submission number
—

Also cleared against K923930: 4 other devices →

What descends from it? (19)

See all 19

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K063609River- 1 Electrocardiograph (Ecg) Recorder and TransmitterShl Telemedicine International, Ltd.200717
K991211Eventstation Ecg Trans-Telephonic Receiving SystemReynolds Medical, Ltd.19990

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 3, 2015Class IIZ-0172-2016Vendor change controlWhen the battery is inserted into the CardioCall ECG Event Recorder, the device beeps twice and is followed by a continuous ringing sound that only stops when battery is removed. The device has a self-test failure and will not operate…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K063609 (Shl Telemedicine International, Ltd., 2007) has gone furthest, with 17 descendants of its own.
  • 4 other devices also cite its predicate K923930.

How does its category compare?

All DXH devices

DXH Transmitters And Receivers, Electrocardiograph, Telephone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DXH

4in 2025

2007: 32008: 72009: 22010: 22012: 12013: 32014: 62015: 42016: 12017: 52018: 42019: 62020: 52021: 32022: 42023: 12025: 42026 (partial year): 1
2007high 7 (2008) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DXH

322in 2025

2007: 202 days median2008: 41 days median2013: 206 days median2014: 186 days median2015: 142 days median2017: 150 days median2018: 294 days median2019: 177 days median2020: 141 days median2021: 107 days median2022: 150 days median2025: 322 days median
2007high 322 (2025) · low 41 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DXH

1in 2015

2008: 42011: 32015: 12026 (partial year): 1
2007high 4 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DXH

27in 2025

2009: 12010: 42011: 12012: 12013: 12014: 12016: 52017: 42020: 52021: 22022: 62023: 42024: 22025: 27
2007high 27 (2025) · low 0 (2007)2025

MAUDE reports naming product code DXH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007320200
200874140
20092—01
20102—04
20110—31
20121—01
2013320601
2014618601
2015414210
20161—05
2017515004
2018429400
2019617700
2020514105
2021310702
2022415006
20231—04
20240—02
20254322027
20261—1—

What is it indicated for?

The event recorder, called the “CardioCall,” is intended to be worn by the patient and to record a short period of electrocardiogram when the patient depresses a button upon sensing symptoms described to him by his physician. The CardioCall can either operate in the so-called “Looping” mode, in which it is always active, and in which it records a minute of one channel of cardiac activity before and a minute after the depression of the button; or in the…

Read the full statement (177 words)

The event recorder, called the “CardioCall,” is intended to be worn by the patient and to record a short period of electrocardiogram when the patient depresses a button upon sensing symptoms described to him by his physician. The CardioCall can either operate in the so-called “Looping” mode, in which it is always active, and in which it records a minute of one channel of cardiac activity before and a minute after the depression of the button; or in the “Post-Activation” mode, in which the device records two channels of cardiac activity for one minute after the depression of the record button. The PC computer program, called the “cardioConnect,” is indicated when it is desired to transfer the memory of the event recorder into a computer by direct connection, instead of using transtelephonic means. The CardioCall recorder is indicated if it is desired to record the electrocardiogram at the time when a patient experiences intermittent or transient symptoms such as dizziness, palpitations, or shortness of breath, and the patient is capable of activating the recorder at these times.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K972649: Cardiocall Cardiac Event Recorder.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K972649

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026