510(k) · Cleared
Recalled · 2003Avl Omni Analyzer
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Avl Scientific Corp. · CGA Glucose Oxidase, Glucose
Cleared on Sep 9, 1997 as substantially equivalent to K954612 (Radiometer America, Inc., 1995) and 3 other predicates after a 50-day review. Its family tree reaches back 1 generation to 1995. It has been recalled 6 times.
- Decided
- Sep 9, 1997
- Review
- 50 days
- Class
- Class II
- Product code
- CGA
- Recalls
- 6
- Predicates
- 4
Its longest line of descent runs back 1 generation to K946206 (1995). 4 ancestors in all. No later device cites it as a predicate yet.
- Root · 1 generation backK946206 Ciba Corning 8-- Series CO-OximeterCiba Corning Diagnostics Corp. · 1995
- This deviceK972733 Avl Omni AnalyzerAvl Scientific Corp. · 1997
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K946206 | Ciba Corning 8-- Series CO-Oximeter | Ciba Corning Diagnostics Corp. | 1995 | — |
| 0 (this device) | K972733 | Avl Omni Analyzer | Avl Scientific Corp. | 1997 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1995 | 1 | 0 |
| 1996 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in CGA? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new CGA device
- #1 K252488 ABL90 FLEX PLUS Systemscore 12.8 · 6 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K954612 | Abl System 625, 615, 605Summary page 1…Analyzers.' Specifically: Radiometer 625 CHEMISTRY CHEMISTRY CHEMISTRY CHEMISTRY CHEMISTRY CHEMISTRY HEMATOLOGY HEMATOLOGY HEMATOLOGY HEMATOLOGY CHEMISTRY CHEMISTRY [K954612], NOVA Biomedical ULTRA [K960120], Chiron 865 [K946206] and Instrumentation Laboratories BGGE [K955793]. 510(k)… | 1995 |
| * | K960120 | Stat Profile Ultra F,G,H,I,J, and K AnalyzersSummary page 1…eter 625 CHEMISTRY CHEMISTRY CHEMISTRY CHEMISTRY CHEMISTRY CHEMISTRY HEMATOLOGY HEMATOLOGY HEMATOLOGY HEMATOLOGY CHEMISTRY CHEMISTRY [K954612], NOVA Biomedical ULTRA [K960120], Chiron 865 [K946206] and Instrumentation Laboratories BGGE [K955793]. 510(k) Summary AVL OMNI Combi Analyzer… | 1996 |
| * | K946206 | Ciba Corning 8-- Series CO-OximeterSummary page 1…EMISTRY CHEMISTRY CHEMISTRY CHEMISTRY CHEMISTRY HEMATOLOGY HEMATOLOGY HEMATOLOGY HEMATOLOGY CHEMISTRY CHEMISTRY [K954612], NOVA Biomedical ULTRA [K960120], Chiron 865 [K946206] and Instrumentation Laboratories BGGE [K955793]. 510(k) Summary AVL OMNI Combi Analyzer (4) Description of th… | 1995 |
| * | K955793 | IL Bgge Blood Gas with Glucose SystemSummary page 1…EMATOLOGY HEMATOLOGY HEMATOLOGY HEMATOLOGY CHEMISTRY CHEMISTRY [K954612], NOVA Biomedical ULTRA [K960120], Chiron 865 [K946206] and Instrumentation Laboratories BGGE [K955793]. 510(k) Summary AVL OMNI Combi Analyzer (4) Description of the new device The AVL OMNI™ Analyzer is a fully-a… | 1996 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Dec 29, 2003 | Not classified | Z-0336-04 | Other | The firm is withdrawing claims that pleural, pericardial, ascitic and cerebrospinal fluids are appropriate sample types for use on these instruments. |
| Dec 29, 2003 | Not classified | Z-0335-04 | Other | The firm is withdrawing claims that pleural, pericardial, ascitic and cerebrospinal fluids are appropriate sample types for use on these instruments. |
| Dec 29, 2003 | Not classified | Z-0337-04 | Other | The firm is withdrawing claims that pleural, pericardial, ascitic and cerebrospinal fluids are appropriate sample types for use on these instruments. |
| Dec 15, 2003 | Not classified | Z-0326-04 | Other | If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume. |
| Dec 15, 2003 | Not classified | Z-0328-04 | Other | If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume. |
| Dec 15, 2003 | Not classified | Z-0327-04 | Other | If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume. |
The 5 latest of 6 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- 1 of the 4 ancestors has a recall of any class.
- 1 other device also cites its predicate K954612.
How does its category compare?
All CGA devicesCGA Glucose Oxidase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · CGA
2in 2024
2026 to Oct 25, outlined.
Median 510(k) review, days · CGA
249in 2023
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · CGA
3in 2023
Class I (2 since 2007) Class II/III not classified
Adverse event reports per year · CGA
59in 2025
MAUDE reports naming product code CGA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 7 | 122 | 2 | 29 |
| 2008 | 6 | 56 | 2 | 37 |
| 2009 | 10 | 155 | 4 | 75 |
| 2010 | 8 | 122 | 6 | 172 |
| 2011 | 9 | 217 | 3 | 423 |
| 2012 | 10 | 123 | 0 | 233 |
| 2013 | 3 | 441 | 3 | 172 |
| 2014 | 5 | 265 | 2 | 122 |
| 2015 | 4 | 150 | 6 (2) | 155 |
| 2016 | 4 | 157 | 2 | 36 |
| 2017 | 2 | — | 0 | 92 |
| 2018 | 2 | — | 0 | 69 |
| 2019 | 2 | — | 1 | 116 |
| 2020 | 2 | — | 0 | 68 |
| 2021 | 1 | — | 0 | 48 |
| 2022 | 1 | — | 0 | 272 |
| 2023 | 4 | 249 | 3 | 1,141 |
| 2024 | 2 | — | 0 | 266 |
| 2025 | 0 | — | 0 | 59 |
| 2026 | 1 | — | 0 | — |
What is it indicated for?
The AVL OMNI™ is a modular Critical Care Analyzer intended to be used for the measurement of pH, PCO₂, PO₂, ctHb, HHb, O₂Hb, COHb, MetHb, SulfHb, Na⁺, K⁺, Cl, ionized Ca™, Hematocrit, glucose and lactate in whole blood, serum, plasma, dialysate and QC materials as appropriate by minimally trained personnel qualified to perform and to report these values in a clinical laboratory setting.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K972733 · printed Nov 5, 2026
Cite this page
510(k) Report. “K972733: Avl Omni Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K972733
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026