# K972880: MDP2000 Medical Data Processor

- Pathway: 510(k) (Traditional)
- Applicant: Vidco, Inc.
- Decided: Oct 22, 1997 after 78 days
- Product code: MSX (System, Network And Communication, Physiological Monitors)
- Device class: 2
- Regulation: 21 CFR 870.2300
- Recalls of this device: 1 (0 Class I)
- Data: FDA decisions to Oct 25, 2026; recalls to Oct 12, 2026

## Lineage
Longest line of descent: 1 generation back to K926162 (1993). 5 ancestors in all.
No later device cites it as a predicate yet.
0 of 6 devices in its family (counting itself) have a Class I recall.

## Predicates on the family tree (from the FDA summary)
- K926162: 580CD Display Driver, Vidco, Inc. (1993) Summary page 1.
- K932894: 620CSR Display Driver, Vidco, Inc. (1994) Summary page 1.
- K934093: 580CDR Display Driver, Vidco, Inc. (1994) Summary page 1.
- K926219: 516YT Display Driver, Vidco, Inc. (1993) Summary page 1.
- K926298: 520 Display Driver- Status I, Vidco, Inc. (1994) Summary page 1.

## Recalls of this device
- Nov 8, 2015 · Class II · Z-0582-2016: Testing at customer site showed unit Remote Patient Monitoring System MDP2040-0100 in a continuous trap condition, not allowing system to reset and reboot. Two customers recently complained of the sys

## Indications for use
The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected. It accepts analog and/or digital physiological information from patient monitoring equipment and processes these signals to provide outputs to: Color monitors for local and remote display of information Chart recorders or printers for hardcopy records Media for storage Other devices to share information Examples of areas where the MDP2000 is used include ICU, CCU, PACU, Emergency, Telemetry, Step-down and other areas where patient’s physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values such as heart rate and systolic, mean, and diastolic pressure.

Source page: https://510kreport.com/device/K972880
Methodology: https://510kreport.com/methodology