510(k) Report

510(k) · Cleared

Cleared

Medtronic Model 6416 Transvenous Bipolar Temporary Pacing Lead

K973360510(k) TraditionalCleared May 26, 1998260-day reviewClass II21 CFR 870.3680

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Vascular · LDF Electrode, Pacemaker, Temporary

Cleared on May 26, 1998 as substantially equivalent to K772103 (Medtronic Vascular, 1977) and 2 other predicates after a 260-day review. Its family tree reaches back 1 generation to 1977.

Decided
May 26, 1998
Review
260 days
Class
Class II
Product code
LDF
Recalls
None
Predicates
3

Its longest line of descent runs back 1 generation to K772103 (1977). 3 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 1995: 1 other ancestor1979: 1 other ancestorK772103 · Bipolar Leads, Disp. · Medtronic Vascular · 1977197519801985199019952000
Other ancestors, by generation 1 generation back: 2 other ancestors K772103 · Bipolar Leads, Disp. · Medtronic Vascular · 19770
  1. Root · 1 generation backK772103 Bipolar Leads, Disp.Medtronic Vascular · 1977
  2. This deviceK973360 Medtronic Model 6416 Transvenous Bipolar Temporary Pacing LeadMedtronic Vascular · 1998
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K772103Bipolar Leads, Disp.Medtronic Vascular1977—
0 (this device)K973360Medtronic Model 6416 Transvenous Bipolar Temporary Pacing LeadMedtronic Vascular1998—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
199510
FIG. 1 — Lineage of K973360, 1977–1998Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LDF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LDF device

  1. #1 K241334 Semi-Floating Temporary Pacing Electrode…score 12.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K772103Bipolar Leads, Disp.
Summary page 1
…ry Systems Device Panel into Class I, (21 CFR 870.3680(a)). 3. SUBSTANTIALLY EQUIVALENT DEVICE(S) Temporary Transvenous Pacing Lead, Medtronic Model 6704, marketed via K772103, K790261 Medtronic Vector X Guiding Catheter, marketed via K950179 4. DEVICE DESCRIPTION The Model 6416 lead consists of a sma…
Medtronic Vascular1977—
*K790261Medtronic Temptron
Summary page 1
…s Device Panel into Class I, (21 CFR 870.3680(a)). 3. SUBSTANTIALLY EQUIVALENT DEVICE(S) Temporary Transvenous Pacing Lead, Medtronic Model 6704, marketed via K772103, K790261 Medtronic Vector X Guiding Catheter, marketed via K950179 4. DEVICE DESCRIPTION The Model 6416 lead consists of a small distal…
Medtronic Vascular1979—
*K950179Medtronic (R) GC Iii Coronary Guiding Catheter
Summary page 1
…TANTIALLY EQUIVALENT DEVICE(S) Temporary Transvenous Pacing Lead, Medtronic Model 6704, marketed via K772103, K790261 Medtronic Vector X Guiding Catheter, marketed via K950179 4. DEVICE DESCRIPTION The Model 6416 lead consists of a small distal active fixation helix, tip and ring electrodes, an insulat…
Medtronics Interventional Vascular1995—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 3 ancestors has a recall.

How does its category compare?

All LDF devices

LDF Electrode, Pacemaker, Temporary: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LDF

5in 2025

2008: 12014: 12015: 22016: 32017: 12018: 12019: 12024: 12025: 52026 (partial year): 1
2007high 5 (2025) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LDF

266in 2025

2016: 138 days median2025: 266 days median
2007high 266 (2025) · low 138 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LDF

1in 2018

2008: 22009: 22011: 12012: 12018: 1, 1 Class I2026 (partial year): 1, 1 Class I
2007high 2 (2008) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LDF

139in 2025

2007: 252008: 182009: 442010: 482011: 352012: 682013: 742014: 682015: 812016: 902017: 792018: 912019: 1302020: 832021: 1252022: 1172023: 2122024: 1902025: 139
2007high 212 (2023) · low 18 (2008)2025

MAUDE reports naming product code LDF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—025
20081—218
20090—244
20100—048
20110—135
20120—168
20130—074
20141—068
20152—081
20163138090
20171—079
20181—1 (1)91
20191—0130
20200—083
20210—0125
20220—0117
20230—0212
20241—0190
202552660139
20261—1 (1)—

What is it indicated for?

The Medtronic Model 6416 Temporary, Transvenous Pacing Lead System features an active fixation, bipolar lead and a soft-tipped lubricated guide catheter. The system is designed for temporary intracardiac pacing and/or EGM recording.

The system is disposable, for temporary single patient use with a contemplated implant duration of 7 days or less. The lead and accessories are supplied sterile.

The lead is introduced transvenously using the guide catheter. Once within the appropriate chamber, the helical tip electrode of the lead is actively fixed into the endocardium. After lead placement, the guide catheter is removed by sliding it over the lead’s bifurcated connector.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K973360: Medtronic Model 6416 Transvenous Bipolar Temporary Pacing Lead.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K973360

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026