510(k) Report

510(k) · Cleared

Recalled · 2004

Roche Amplicor CT/NG Test for Chlamydia Trachomatis

K973707510(k) TraditionalCleared Aug 4, 1999674-day reviewClass I21 CFR 866.3120Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Molecular Systems, Inc. · MKZ Dna Probe, Nucleic Acid Amplification, Chlamydia

Cleared on Aug 4, 1999 after a 674-day review. The FDA summary names no predicate by number. 1 later device descends from it. It has been recalled once.

Decided
Aug 4, 1999
Review
674 days
Class
Class I
Product code
MKZ
Recalls
1
Predicates
0

1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2007: 1 descendant cleared1995200020052010201520202025
  1. This deviceK973707 Roche Amplicor CT/NG Test for Chlamydia TrachomatisRoche Molecular Systems, Inc. · 1999
  2. Descendant · cites it directly · 0 of its ownK070174 Amplicor CT/NG Test for Chlamydia Trachomatis; Roche Scripts for…Roche Diagnostics Corporation · 2007

Most influential descendants

Showing 1 of 1 · See all
  • K070174Amplicor CT/NG Test for Chlamydia Trachomatis; Roche Scripts for CT/NG Test AccessoryRoche Diagnostics Corporation · 2007 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K973707Roche Amplicor CT/NG Test for Chlamydia TrachomatisRoche Molecular Systems, Inc.1999—
+1K070174Amplicor CT/NG Test for Chlamydia Trachomatis; Roche Scripts for CT/NG Test…Roche Diagnostics Corporation2007—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
FIG. 1 — Lineage of K973707, 1999–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MKZ device

  1. #1 K230451 Aptima® Chlamydia trachomatis Assayscore 6.3 · 1 generation · no Class I recall
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K070174Amplicor CT/NG Test for Chlamydia Trachomatis; Roche Scripts for CT/NG Test AccessoryRoche Diagnostics Corporation20070

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 7, 2004Not classifiedZ-1360-04OtherAn increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent during PCR detection.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K070174 (Roche Diagnostics Corporation, 2007) has gone furthest, with 0 descendants of its own.

How does its category compare?

All MKZ devices

MKZ Dna Probe, Nucleic Acid Amplification, Chlamydia: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKZ

1in 2023

2007: 22008: 22009: 22012: 12014: 12023: 1
2007high 2 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKZ

Too few MKZ 510(k)s in any year since 2007 to give a median.

Recalls per year · MKZ

1in 2010

2009: 12010: 1
2007high 1 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MKZ

1in 2025

2010: 12011: 92012: 12013: 62014: 22015: 12016: 12017: 62018: 32019: 32020: 72021: 92022: 22023: 42024: 72025: 1
2007high 9 (2011) · low 0 (2007)2025

MAUDE reports naming product code MKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—00
20092—10
20100—11
20110—09
20121—01
20130—06
20141—02
20150—01
20160—01
20170—06
20180—03
20190—03
20200—07
20210—09
20220—02
20231—04
20240—07
20250—01
20260—0—

What is it indicated for?

The AMPLICOR CT/NG Test for Chlamydia trachomatis is a qualitative in vitro test for the detection of Chlamydia trachomatis plasmid DNA in urine from males and females, in endocervical swab specimens from symptomatic or asymptomatic females, and in urethral swab specimens as evidence of symptomatic or asymptomatic infection with Chlamydia trachomatis. Chlamydia trachomatis DNA is detected by Polymerase Chain Reaction (PCR) amplification of target DNA and by hybridization capture of amplified target and by hybridization capture of the amplified target.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K973707: Roche Amplicor CT/NG Test for Chlamydia Trachomatis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K973707

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026