510(k) · Cleared
ClearedShelhigh NO-React Pericardial Patches
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Shelhigh, Inc. · DXZ Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene
Cleared on Jun 3, 1998 as substantially equivalent to K964467 (Shelhigh, Inc., 1997) after a 177-day review. Its family tree reaches back 1 generation to 1997. 15 later devices descend from it.
- Decided
- Jun 3, 1998
- Review
- 177 days
- Class
- Class II
- Product code
- DXZ
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 1 generation to K964467 (1997). 15 later devices descend from it, 3 citing it directly.
- Root · 1 generation backK964467 Shelhigh Pericardial PatchShelhigh, Inc. · 1997
- This deviceK974914 Shelhigh NO-React Pericardial PatchesShelhigh, Inc. · 1998
- Descendant · cites it directly · 12 of its ownK982810 Shelhigh NO-React VascupatchShelhigh, Inc. · 1998
- Descendant · 2 generations on · 11 of its ownK063349 Cormatrix Patch for Cardiac Tissue RepairCormatrix Cardiovascular, Inc. · 2007
- Descendant · 3 generations on · 4 of its ownK140306 Cormatrix Protect Ecm EnvelopeCormatrix Cardiovascular, Inc. · 2014 · Recalled
- Descendant · 4 generations on · 3 of its ownK173242 CanGaroo Envelope (small), CanGaroo Envelope (medium), CanGaroo…Aziyo Biologics, Inc. · 2017 · Recalled
- Descendant · 5 generations on · 2 of its ownK182255 CanGaroo Neuro Envelope (small) ; CanGaroo Neuro Envelope (medium)…Aziyo Biologics, Inc. · 2018 · Recalled
Showing 5 of 15 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 15 · See all- K982810Shelhigh NO-React VascupatchShelhigh, Inc. · 1998 · cites it directly12descendants
- K063349Cormatrix Patch for Cardiac Tissue RepairCormatrix Cardiovascular, Inc. · 2007 · 2 generations on11descendants
- K140306Cormatrix Protect Ecm EnvelopeCormatrix Cardiovascular, Inc. · 2014 · 3 generations on4descendants
- K173242CanGaroo Envelope (small), CanGaroo Envelope (medium), CanGaroo Envelope (large)…Aziyo Biologics, Inc. · 2017 · 4 generations on3descendants
- K182255CanGaroo Neuro Envelope (small) ; CanGaroo Neuro Envelope (medium) ; CanGaroo Neuro…Aziyo Biologics, Inc. · 2018 · 5 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K964467 | Shelhigh Pericardial Patch | Shelhigh, Inc. | 1997 | — |
| 0 (this device) | K974914 | Shelhigh NO-React Pericardial Patches | Shelhigh, Inc. | 1998 | — |
| +1 | K982810 | Shelhigh NO-React Vascupatch | Shelhigh, Inc. | 1998 | — |
| +2 | K063349 | Cormatrix Patch for Cardiac Tissue Repair | Cormatrix Cardiovascular, Inc. | 2007 | — |
| +3 | K140306 | Cormatrix Protect Ecm Envelope | Cormatrix Cardiovascular, Inc. | 2014 | 2 |
| +4 | K173242 | CanGaroo Envelope (small), CanGaroo Envelope (medium), CanGaroo Envelope… | Aziyo Biologics, Inc. | 2017 | 1 |
| +5 | K182255 | CanGaroo Neuro Envelope (small) ; CanGaroo Neuro Envelope (medium) ; CanGaroo… | Aziyo Biologics, Inc. | 2018 | 1 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1998 | 2 | 0 |
| 1999 | 1 | 0 |
| 2007 | 1 | 0 |
| 2008 | 1 | 0 |
| 2011 | 1 | 0 |
| 2014 | 2 | 0 |
| 2016 | 1 | 0 |
| 2017 | 1 | 0 |
| 2018 | 2 | 0 |
| 2019 | 1 | 0 |
| 2020 | 1 | 0 |
| 2021 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DXZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DXZ device
- #1 K240775 PeriBeam® Pericardial Membranescore 9.2 · 1 generation · no Class I recall
- #2 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K964467 | Shelhigh Pericardial PatchSummary page 2…ong history of success as a permanently implanted material. V. Substantial Equivalence It is substantially equivalent to the Shelhigh pericardial patch #K964467 manufactured by Shelhigh Inc.. VI. Comparison with Predicate Device The Shelhigh No-React® Pericardial patch like the Shelhigh… | 1997 |
What descends from it? (15)
See all 153 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All DXZ devicesDXZ Patch, Pledget And Intracardiac, Petp, Ptfe, Polypropylene: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DXZ
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DXZ
106in 2014
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DXZ
2in 2025
Class I (2 since 2007) Class II/III not classified
Adverse event reports per year · DXZ
119in 2025
MAUDE reports naming product code DXZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 1 | — | 0 | 50 |
| 2008 | 2 | — | 0 | 6 |
| 2009 | 1 | — | 0 | 109 |
| 2010 | 3 | 91 | 2 | 81 |
| 2011 | 2 | — | 2 | 47 |
| 2012 | 1 | — | 0 | 21 |
| 2013 | 1 | — | 2 | 69 |
| 2014 | 3 | 106 | 0 | 93 |
| 2015 | 2 | — | 0 | 171 |
| 2016 | 2 | — | 0 | 66 |
| 2017 | 1 | — | 0 | 30 |
| 2018 | 0 | — | 1 | 44 |
| 2019 | 0 | — | 0 | 55 |
| 2020 | 1 | — | 1 | 121 |
| 2021 | 0 | — | 0 | 80 |
| 2022 | 0 | — | 0 | 77 |
| 2023 | 0 | — | 2 | 40 |
| 2024 | 0 | — | 0 | 49 |
| 2025 | 1 | — | 2 (2) | 119 |
| 2026 | 0 | — | 1 | — |
What is it indicated for?
The Shelhigh No-React® Pericardial patch is intended for use during cardiac surgical procedure as an intracardiac patch for open heart surgery, as well as to close intercavitary defect, and enlarge the aortic root . The patch may be tailored during surgery to meet the surgeon's needs.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K974914 · printed Nov 5, 2026
Cite this page
510(k) Report. “K974914: Shelhigh NO-React Pericardial Patches.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K974914
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026